CC BY-NC-ND 4.0 · Endoscopy 2022; 54(S 02): E767-E768
DOI: 10.1055/a-1792-2955
E-Videos

Endoscopic tapered sheath-assisted removal of a proximally migrated pancreatic stent

Akihisa Kato
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
,
Makoto Natsume
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
,
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
,
Katsuyuki Miyabe
2   Department of Gastroenterology, Japanese Red Cross Aichi Medical Center Nagoya Daini Hospital, Nagoya, Japan
,
Yasuki Hori
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
,
Itaru Naitoh
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
,
Kazuki Hayashi
1   Department of Gastroenterology and Metabolism, Nagoya City University Graduate School of Medical Sciences, Nagoya, Japan
› Author Affiliations
 

Endoscopic pancreatic stenting is beneficial for patients with a variety of pancreatic conditions; however, proximal migration of the stent occurs occasionally and removal is technically demanding [1] [2]. It is sometimes quite challenging to pass stent-removal devices such as forceps, baskets, and snares across a stricture, particularly in cases in which the migrated stent is located further away from the pancreatic duct stricture [3]. We present an impressive case in which a new endoscopic tapered sheath (EndoSheather; Piolax, Inc., Kanagawa, Japan) ([Fig. 1]) [4] contributed successfully to the removal of a proximally migrated stent in a patient with a distal stricture. This novel device allowed the stricture to be passed and devices to be inserted easily through an indwelling outer sheath, providing a bridge to the target space even in the deep pancreatic tail.

Zoom Image
Fig. 1 The endoscopic tapered sheath (EndoSheather; Piolax, Inc., Kanagawa, Japan). a Tip calibers of the inner catheter and the outer sheath are similar. A 0.025-inch guidewire is used with the inner catheter. Outer-sheath external diameter: 7.2 Fr (2.44 mm); interior diameter: 6.2 Fr (2.06 mm). The outer-sheath tip has a radiopaque marker (blue arrowhead). b A retrieval basket can be inserted through the outer sheath along the 0.025-inch guidewire.

An 84-year-old woman with a pancreatic head intraductal tubulopapillary neoplasm developed abdominal pain caused by a pancreatic duct stricture. Placement of a 7 Fr plastic stent (7 cm) across the stricture improved her symptoms. However, the pancreatic stent migrated beyond the stricture ([Fig. 2]). A conventional balloon catheter failed to remove the stent. None of the devices used to grab the migrated stent could access the proximal duct due to the severe stricture. Therefore, the new endoscopic sheath, with a tapered inner catheter tip to allow easy passage across the stricture, was inserted into the pancreatic duct over a 0.025-inch guidewire. Only the inner catheter was removed, leaving the outer sheath near the target position. Subsequently, the retrieval basket was inserted smoothly through the outer sheath to reach the migrated stent. The migrated stent was grasped with the basket and successfully dragged out to the duodenum without adverse events ([Fig. 3], [Video 1]).

Zoom Image
Fig. 2 The proximally migrated stent was located away from the stricture (red arrowheads). Yellow arrows indicate the proximal head and the distal end of the migrated stent. The new endoscopic tapered sheath was inserted deep into the pancreatic duct over a 0.025-inch guidewire. Blue arrowhead shows the radiopaque marker of the outer sheath.
Zoom Image
Fig. 3 The retrieval basket reached the head of the migrated stent smoothly through the outer sheath. The migrated stent was successfully removed using the Lariat technique by grabbing the distal end of the stent. a Fluoroscopic image. b Endoscopic image.

Video 1 A new endoscopic tapered sheath contributed to the successful removal of a proximally migrated stent that was located away from the stricture.


Quality:

This technique may be a useful option for removing a proximally migrated pancreatic stent that is beyond the stricture or deep in the pancreatic duct.

Endoscopy_UCTN_Code_TTT_1AR_2AB

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.

This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos


#

Competing interests

The authors declare that they have no conflict of interest.

  • References

  • 1 Gabbrielli A, Pandolfi M, Mutignani M. et al. Efficacy of main pancreatic-duct endoscopic drainage in patients with chronic pancreatitis, continuous pain, and dilated duct. Gastrointest Endosc 2005; 61: 576-581
  • 2 Price LH, Brandabur JJ, Kozarek RA. et al. Good stents gone bad: endoscopic treatment of proximally migrated pancreatic duct stents. Gastrointest Endosc 2009; 70: 174-179
  • 3 Matsumoto K, Katanuma A, Maguchi H. Endoscopic removal technique of migrated pancreatic plastic stents. J Hepatobiliary Pancreat Sci 2014; 21: E34-E40
  • 4 Matsumori T, Uza N, Shiokawa M. et al. Mapping biopsy for bile duct cancer using a novel device delivery system. Endoscopy 2021; DOI: 10.1055/a-1479-1969.

Corresponding author

Michihiro Yoshida, MD, PhD
Department of Gastroenterology and Metabolism
Nagoya City University Graduate School of Medical Sciences
1 Kawasumi
Mizuho-cho, Mizuho-ku
Nagoya 467-8601
Japan   

Publication History

Article published online:
31 March 2022

© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

  • References

  • 1 Gabbrielli A, Pandolfi M, Mutignani M. et al. Efficacy of main pancreatic-duct endoscopic drainage in patients with chronic pancreatitis, continuous pain, and dilated duct. Gastrointest Endosc 2005; 61: 576-581
  • 2 Price LH, Brandabur JJ, Kozarek RA. et al. Good stents gone bad: endoscopic treatment of proximally migrated pancreatic duct stents. Gastrointest Endosc 2009; 70: 174-179
  • 3 Matsumoto K, Katanuma A, Maguchi H. Endoscopic removal technique of migrated pancreatic plastic stents. J Hepatobiliary Pancreat Sci 2014; 21: E34-E40
  • 4 Matsumori T, Uza N, Shiokawa M. et al. Mapping biopsy for bile duct cancer using a novel device delivery system. Endoscopy 2021; DOI: 10.1055/a-1479-1969.

Zoom Image
Fig. 1 The endoscopic tapered sheath (EndoSheather; Piolax, Inc., Kanagawa, Japan). a Tip calibers of the inner catheter and the outer sheath are similar. A 0.025-inch guidewire is used with the inner catheter. Outer-sheath external diameter: 7.2 Fr (2.44 mm); interior diameter: 6.2 Fr (2.06 mm). The outer-sheath tip has a radiopaque marker (blue arrowhead). b A retrieval basket can be inserted through the outer sheath along the 0.025-inch guidewire.
Zoom Image
Fig. 2 The proximally migrated stent was located away from the stricture (red arrowheads). Yellow arrows indicate the proximal head and the distal end of the migrated stent. The new endoscopic tapered sheath was inserted deep into the pancreatic duct over a 0.025-inch guidewire. Blue arrowhead shows the radiopaque marker of the outer sheath.
Zoom Image
Fig. 3 The retrieval basket reached the head of the migrated stent smoothly through the outer sheath. The migrated stent was successfully removed using the Lariat technique by grabbing the distal end of the stent. a Fluoroscopic image. b Endoscopic image.