Rofo 2004; 176(11): 1560-1565
DOI: 10.1055/s-2004-813629
Rapid Communication

© Georg Thieme Verlag KG Stuttgart · New York

Evaluation of Balanced Steady-State Free Precession (TrueFISP) and K-space Segmented Gradient Echo Sequences for 3D Coronary MR Angiography with Navigator Gating at 3 Tesla

Evaluation von Balanced Steady-State Free Precession (TrueFISP) und k-Raum segmentierter Gradientenechosequenzen für die 3D-MR-Koronarangiographie mit Navigator-Technik bei 3 TeslaM. G. Kaul1 , A. Stork1 , P. M. Bansmann1 , C. Nolte-Ernsting1 , G. K. Lund2 , C. Weber1 , G. Adam1
  • 1Radiologisches Zentrum, Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie, Universitätsklinikum Hamburg-Eppendorf
  • 2Herzzentrum, Klinik und Poliklinik für Kardiologie/Angiologie, Universitätsklinikum Hamburg-Eppendorf
The authors wish to thank Dr. Kay Nehrke from the Philips Research Laboratories, Hamburg, for his technical support. This study was supported by the EU Project MRI-MARCB.
Further Information

Publication History

Publication Date:
20 October 2004 (online)

Zusammenfassung

Ziel: Erprobung von k-Raum segmentierten Gradientenecho-Pulssequenzen für die frei geatmete Magnetresonanz-Koronarangiographie (cMRA) auf einem klinischen 3-Tesla (T)-System. Methoden: Es wurden T2-präparierte, fettsupprimierte Turbo-Field-Echo (TFE, turboFLASH, SFPGR) und balanced-TFE (b-TFE, TrueFISP, FIESTA, segmentierte SSFP) Sequenzen mit Navigator-Gating zur prospektiven Bewegungskorrektur auf einem 3T System, ausgestattet mit einer sechselementigen Phased-array-Spule, durchgeführt. Bei 15 gesunden Probanden wurde die rechte Koronararterie (RCA) mit der TFE und der b-TFE-Sequenz und aufgrund von Untersuchungszeitbeschränkungen bei 10 Probanden die linke Koronararterie (LM/LAD) ausschließlich mit der TFE Sequenz untersucht. Die Bildqualität wurde auf einer fünfstufigen Skala eingestuft (0 = nicht sichtbar bis 4 = exzellent). Die Längen, die Durchmesser und die Schärfe der Gefäße und das Kontrast-zu-Rausch-Verhältnis (CNR) wurden bestimmt. Ergebnisse: 98 % aller Hauptsegmente (proximales, mittleres, distales) der RCA konnten mit der TFE und 82 % mit der b-TFE gesehen werden. Die Bildqualität der drei Segmente war höher für die TFE (2,7/2,7/1,5) als für die b-TFE (1,9/1,6/0,9) mit p: (≤ 0,001/≤ 0,004/≤ 0,056). Die gemessene Gefäßlänge war länger für die TFE (95 ± 22 mm) als für die b-TFE (80 ± 40 mm; p ≤ 0,115). Es wurden keine signifikanten Änderungen (p ≤ 0,074, p ≤ 0,145) im Durchmesser und der Schärfe der RCAs zwischen der TFE (2,4 ± 0,3 mm, 60 % ± 5) und der b-TFE Sequenz (2,4 ± 0,3 mm, 62 % ± 6) beobachtet. Das CNR war höher für TFE (10,1 ± 3,4) als für b-TFE (6,6 ± 2,1; p ≤ 0,014). Alle zehn Hauptstämme und proximalen Segmente der LM/LAD, die nur mit der TFE-Sequenz untersucht wurden, waren sichtbar mit einer Bildqualität von 2,5 und 2,1. Das mittlere Segment war sichtbar in 7 Fällen mit einer Bildqualität von 1,3. In 3 Fällen war das distale Segment sichtbar mit einer Bildqualität von 0,5. Die Gefäßlänge betrug 78 ± 27 mm und das CNR 11,9 ± 2,4. Schlussfolgerungen: Die konventionelle TFE-Technik hat die Machbarkeit der cMRA bei 3T gezeigt. Die b-TFE ist in ihrer jetzigen Verfügbarkeit bei 3T der TFE-Sequenz unterlegen.

Abstract

Purpose: To test the feasibility of k-space segmented gradient-echo pulse sequences for free-breathing coronary magnetic resonance angiography (cMRA) on a clinical 3T system. Materials and Methods: T2-prepared, fat-suppressed turbo field echo (TFE, turboFLASH, SFPGR) as well as balanced TFE (b-TFE, trueFISP, FIESTA, segmented SSFP) sequences with navigator gating for prospective motion correction were applied on a 3T system equipped with a six-element phased-array cardiac coil. In 15 healthy volunteers, the right coronary artery (RCA) was examined with TFE and b-TFE sequences. Due to examination time limitations, the left coronary artery (LM/LAD) was examined exclusively with the TFE sequence in ten volunteers. Image quality was graded on a five point scale (0 = not visualized to 4 = excellent). The length, diameter and sharpness of the vessels and the contrast-to-noise ratios (CNR) were measured. Results: 98 % of all major segments (proximal/middle/distal) of the RCA could be seen with the TFE sequence and 82 % with the b-TFE sequence. The image quality for the three segments was graded higher for the TFE sequence (2.7/2.7/1.5) than for the b-TFE sequence (1.9/1.6/0.9) with P: (≤ 0.001/≤ 0.004/≤ 0.056). The kappa of the interobserver variability was 0.75 for the TFE sequence and 0.8 for the b-TFE sequence. The measured vessel lengths were longer for the TFE sequence (95 ± 22 mm) than for the b-TFE sequence (80 ± 40 mm; P ≤ 0.115). No significant changes (P ≤ 0.074, P ≤ 0.145) in diameter and vessel sharpness of the RCAs were observed between the TFE (2.4 ± 0.3 mm, 60 % ± 5) and b-TFE sequences (2.4 ± 0.3 mm, 62 % ± 6). The CNR was higher for the TFE sequence (10.1 ± 3.4) than for the b-TFE sequence (6.6 ± 2.1; P ≤ 0.014). All ten main and proximal segments of the LM/LAD, which were examined exclusively with the TFE sequence, were visible with grade 2.5 and 2.1. The middle segment was visible in seven cases with grade 1.3. In three cases, the distal segment was visible with grade 0.5. The vessel length was 78 ± 27 mm and the CNR 11.9 ± 2.4. Conclusion: The conventional TFE technique has demonstrated good feasibility for cMRA at 3T. In its operational availability at 3T, the b-TFE sequence is inferior to the TFE sequence.

References

  • 1 Wang Y, Riederer S, Ehman R. Respiratory motion of the heart: kinematics and the implications for the spatial resolution in coronary imaging.  Magn Reson Med. 1995;  33 713-719
  • 2 Stuber M, Botnar R, Fischer S. et al . Preliminary report on in vivo coronary MRA at 3 Tesla in humans.  Magn Reson Med. 2002;  48 425-429
  • 3 Spuentrup E, Bornert P, Botnar R. et al . Navigator-gated free-breathing three-dimensional balanced fast field echo (TrueFISP) coronary magnetic resonance angiography.  Invest Radiol. 2002;  37 637-642
  • 4 Barkhausen J, Hunold P, Jochims M. et al . Comparison of gradient-echo and steady state free precession sequences for 3D-navigator MR angiography of coronary arteries.  Fortschr Röntgenstr. 2002;  174 725-730
  • 5 Spuentrup E, Katoh M, Buecker A. et al . Free-breathing 3D Steady-State Free Precession Coronary MR Angiography with Radial k-Space Sampling: Comparison with Cartesian k-Space Sampling and Cartesian Gradient-Echo Coronary MR Angiography - Pilot Study.  Online Radiology. 2004; 
  • 6 Fischer S, Wickline S, Lorenz C. Novel real-time R-wave detection algorithm based on the vectorcardiogram for accurate gated magnetic resonance acquisitions.  Magn Reson Med. 1999;  42 361-370
  • 7 Wang Y, Vidan E, Bergman G. Cardiac motion of coronary arteries: variability in the rest period and implications for coronary MR angiography.  Radiology. 1999;  213 751-758
  • 8 Nehrke K, Bornert P, Groen J. et al . On the performance and accuracy of 2D navigator pulses.  Magn Reson Imaging. 1999;  17 1173-1181
  • 9 Stuber M, Botnar R, Danias P. et al . Double-oblique free-breathing high resolution three-dimensional coronary magnetic resonance angiography.  J Am Coll Cardiol. 1999;  34 524-531
  • 10 Brittain J, Hu B, Wright G. et al . Coronary angiography with magnetization-prepared T2 contrast.  Magn Reson Med. 1995;  33 689-696
  • 11 Deimling M, Heid O. Magnetization prepared truef FISP imaging. San Francisco; Proceedings of the 2nd Annual Meeting of SMR 1994: 495
  • 12 Scanlon P, Faxon D, Audet A. et al . ACC/AHA guidelines for coronary angiography. A report of the American College of Cardiology/American Heart Association Task Force on practice guidelines (Committee on Coronary Angiography). Developed in collaboration with the Society for Cardiac Angiography and Interventions.  J Am Coll Cardiol. 1999;  33 1756-1824
  • 13 Etienne A, Botnar R, Van M A. et al . “Soap-Bubble” visualization and quantitative analysis of 3D coronary magnetic resonance angiograms.  Magn Reson Med. 2002;  48 658-666
  • 14 Noeske R, Seifert F, Rhein K. et al . Human cardiac imaging at 3 T using phased array coils.  Magn Reson Med. 2000;  44 978-982
  • 15 Schar M, Kozerke S, Fischer S. et al . Cardiac SSFP imaging at 3 Tesla.  Magn Reson Med. 2004;  51 799-806
  • 16 Hargreaves B, Vasanawala S, Pauly J. et al . Characterization and reduction of the transient response in steady-state MR imaging.  Magn Reson Med. 2001;  46 149-158
  • 17 Hennig J, Speck O, Scheffler K. Optimization of signal behavior in the transition to driven equilibrium in steady-state free precession sequences.  Magn Reson Med. 2002;  48 801-809
  • 18 Sommer T, Hofer U, Hackenbroch M. et al . Submillimeter 3D coronary MR angiography with real-time navigator correction in 107 patients with suspected coronary artery disease.  Fortschr Röntgenstr. 2002;  174 459-466
  • 19 Kim W Y, Danias P G, Stuber M. et al . Coronary Magnetic Resonance Angiography for the Detection of Coronary Stenoses.  N Engl J Med. 2001;  345 1863-1869
  • 20 Manke D, Nehrke K, Bornert P. Novel prospective respiratory motion correction approach for free-breathing coronary MR angiography using a patient-adapted affine motion model.  Magn Reson Med. 2003;  50 122-131

Michael Kaul

Universitätsklinikum Hamburg-Eppendorf, Radiologisches Zentrum, Klinik und Poliklinik für Diagnostische und Interventionelle Radiologie

Martinistraße 52

20246 Hamburg

Phone: ++ 49/40/4 28 03 40 10

Email: mkaul@uke.uni-hamburg.de