Cent Eur Neurosurg 2011; 72(1): 22-27
DOI: 10.1055/s-0030-1253350
Original Article

© Georg Thieme Verlag KG Stuttgart · New York

Does Electrode Placement Influence Quality of Intraoperative Monitoring in Vestibular Schwannoma Surgery?

S. Rampp1 , J. Prell1 , J. C. Rachinger1 , C. Scheller1 , C. Strauss1
  • 1University of Halle-Wittenberg, Department of Neurosurgery, Halle (Saale), Germany
Further Information

Publication History

Publication Date:
11 June 2010 (online)

Abstract

Object: Continuous recording of electromyographic signals (EMG) is a standard method for intraoperative monitoring of facial nerve function in cerebello-pontine angle surgery. Subcutaneous needle electrodes in the facial muscles are used in different setups. The goal of this study was to compare two commonly used electrode setups concering sensitivity for pathological EMG activity.

Patients and methods: A group of 10 patients undergoing vestibular schwannoma surgery were examined. Continuous EMG from facial muscles was recorded using needle electrodes in setups according to Kartush or Møller, with narrow or wide interelectrode distances, respectively. Quantity of pathological A-train activity and signal-to-noise ratios were compared between setups.

Results: A-train activity was seen in all patients. On average, 37% of A-train activity was seen in the Kartush setup alone, 4% in Møller setups alone and 59% in both setups synchronously (p<0.05; ratio of median train time – Kartush:Møller 3:2). The wide interelectrode distance of the Møller setup was found to be significantly more susceptible to artefacts, especially to low frequency and power line noise. Artefacts were the main reason for the Møller setup to fail detecting A-train activity.

Conclusions: For continuous intraoperative monitoring of facial nerve function, narrow interelectrode distance should be used.

References

  • 1 Delgado TE, Bucheit WA, Rosenholtz HR. et al . Intraoperative monitoring of facial muscle evoked responses obtained by intracranial stimulation of the facial nerve: a more accurate technique for facial nerve dissection.  Neurosurgery. 1979;  4 418-421
  • 2 Kartush JM. Intra-operative monitoring in acoustic neuroma surgery.  Neurol Res. 1998;  20 593-596
  • 3 Møller AR, Janetta PJ. Preservation of facial function during removal of acoustic neuromas. Use of monopolar constant-voltage stimulation and EMG.  J Neurosurg. 1984;  61 757-760
  • 4 Prass RL, Kinney SE, Hardy Jr RW. et al . Acoustic (loudspeaker) facial EMG monitoring: II. Use of evoked EMG activity during acoustic neuroma resection.  Otolaryngol Head Neck Surg. 1987;  97 541-551
  • 5 Prass RL, Luders H. Acoustic (loudspeaker) facial electromyographic monitoring: part 1. Evoked electromyographic activity during acoustic neuroma resection.  Neurosurgery. 1986;  19 392-400
  • 6 Romstőck J, Strauss C, Fahlbusch R. Continuous electromyography monitoring of motor cranial nerves during cerebellopontine angle surgery.  J Neurosurg. 2000;  93 586-593
  • 7 Prell J, Rampp S, Romstock J. et al . Train time as a quantitative electromyographic parameter for facial nerve function in patients undergoing surgery for vestibular schwannoma.  J Neurosurg. 2007;  106 826-832
  • 8 Rampp S, Prell J, Thielemann H. et al . Baseline correction of intraoperative electromyography using discrete wavelet transform.  J Clin Monit Comput. 2007;  21 219-226
  • 9 Scheller C, Richter HP, Engelhardt M. et al . The influence of prophylactic vasoactive treatment on cochlear and facial nerve functions after vestibular schwannoma surgery: a prospective and open-label randomized pilot study.  Neurosurgery. 2007;  61 92-97 ; discussion 97–98
  • 10 Strauss C, Romstock J, Fahlbusch R. et al . Preservation of facial nerve function after postoperative vasoactive treatment in vestibular schwannoma surgery.  Neurosurgery. 2006;  59 577-584
  • 11 Schipper J, Ridder GJ, Arapakis I. et al . [Neurophysiologic intraoperative monitoring to preserve cranial nerve function in base of skull surgery].  HNO. 2004;  52 897-907
  • 12 Grayeli AB, Guindi S, Kalamarides M. et al . Four-channel electromyography of the facial nerve in vestibular schwannoma surgery: sensitivity and prognostic value for short-term facial function outcome.  Otol Neurotol. 2005;  26 114-120
  • 13 Schlake HP, Goldbrunner R, Milewski C. et al . Technical developments in intra-operative monitoring for the preservation of cranial motor nerves and hearing in skull base surgery.  Neurol Res. 1999;  21 11-24

Correspondence

Dr. S. Rampp

University Halle-Wittenberg

Neurosurgery Ernst-Grube-Straße 40

06097 Halle (Saale)

Germany

Phone: +49-345-5571407

Fax: +49-345-5571412

Email: stefan.rampp@medizin.uni-halle.de