Thorac Cardiovasc Surg 2014; 62(07): 575-577
DOI: 10.1055/s-0034-1371698
Short Communication
Georg Thieme Verlag KG Stuttgart · New York

72 Hours Standby Time of Wet-Primed Cardiopulmonary Bypass Circuits: A Microbiological Quality Assurance Study

Sebastian Schulz-Stübner
1   Deutsches Beratungszentrum für Hygiene (BZH GmbH), Freiburg, Germany
,
Cornelia Schorer
2   Mediclin Herzzentrum Lahr/Baden, Schwarzwald, Germany
,
Jürgen Ennker
2   Mediclin Herzzentrum Lahr/Baden, Schwarzwald, Germany
,
Stefan Bauer
2   Mediclin Herzzentrum Lahr/Baden, Schwarzwald, Germany
,
Reiner Schaumann
1   Deutsches Beratungszentrum für Hygiene (BZH GmbH), Freiburg, Germany
› Institutsangaben
Weitere Informationen

Publikationsverlauf

10. November 2013

09. Dezember 2013

Publikationsdatum:
01. Mai 2014 (online)

Abstract

In a microbiological sample study of 15 wet-primed cardiopulmonary bypass circuits in standby mode for 72 hours under regular clinical conditions, no contamination of the priming fluid or the connectors could be detected. Hand contact surfaces of the machines demonstrated environmental microorganisms. These findings indicate the safe use of primed cardiopulmonary bypass circuits in standby mode for 72 hours. A surface disinfection of hand contact surfaces immediately before use is recommended.

 
  • References

  • 1 Yavuz SS, Tarçın Ö, Ada S , et al. Incidence, aetiology, and control of sternal surgical site infections. J Hosp Infect 2013; 85 (3) 206-212
  • 2 Loftus RW, Muffly MK, Brown JR , et al. Hand contamination of anesthesia providers is an important risk factor for intraoperative bacterial transmission. Anesth Analg 2011; 112 (1) 98-105
  • 3 Karimova A, Robertson A, Cross N , et al. A wet-primed extracorporeal membrane oxygenation circuit with hollow-fiber membrane oxygenator maintains adequate function for use during cardiopulmonary resuscitation after 2 weeks on standby. Crit Care Med 2005; 33 (7) 1572-1576
  • 4 Witschi A, Widmer AF, Schnider BF, Wehrle J, Zerkowski HR, Bernet F. Safety of heart-lung machines during prolonged standby. Infection 2003; 31 (5) 350-352
  • 5 Hamers LA, Linssen CF, Lancé MD , et al. Positive cultures from cardiopulmonary bypass: prevalence and relevance regarding postoperative infection. Eur J Cardiothorac Surg 2011; 40 (2) 372-378