Thromb Haemost 1983; 50(03): 676-677
DOI: 10.1055/s-0038-1665285
DOI: 10.1055/s-0038-1665285
Original Article
The Use of Evacuated Tubes for Blood Collection in Oral Anticoagulant Control
Weitere Informationen
Publikationsverlauf
Received 24. Mai 1983
Accepted 15. Juli 1983
Publikationsdatum:
18. Juli 2018 (online)
Summary
The prothrombin time (PT) test is used to monitor oral anticoagulant therapy. Contact activation of patient blood samples in vitro may shorten the PT and thus lead to underestimation of the degree of anticoagulation. Commercial evacuated tubes for blood collection have been reported to induce shortening of the PT. The Thrombotest modification of the PT determination was used to investigate the properties of different brands of evacuated siliconized tubes. Two recently manufactured batches of such tubes did not induce appreciable shortening of the PT and can therefore be recommended for use in monitoring oral anticoagulant treatment.
-
References
- 1 Blackburn EK, Shinton NK. Blood specimen collection tubes for coagulation tests. Lancet 1979; 1: 680
- 2 Palmer RN, Kessler CM, Gralnick HR. Warfarin Anticoagulation: Difficulties in interpretation of the prothrombin time. Thromb Res 1982; 25: 125-130
- 3 Van den Besselaar AM H P, Loeliger EA. The effect of contact activation on the prothrombin time with special reference to the use of evacuated tubes. In: Triplett DA. (Ed) Standardization of Coagulation Assays: An Overview. College of American Pathologists; Skokie, Illinois: 1982. pp 95-104
- 4 Owren PA. Thrombotest. A new method for controlling anticoagulant therapy Lancet 1959; 2: 754-758
- 5 Loeliger EA, Van Halem-Visser LP. Biological properties of the thromboplastins and plasmas included in the ICTH/ICSH collaborative study on prothrombin time standardization. Thromb Haemostas 1979; 42: 1115-1127
- 6 Loeliger EA, Van der Hoeff-Van Halem R, Van Halem-Visser LP. Thromboplastin Calibration. Experience of the Dutch Reference Laboratory for Anticoagulant Control Thromb Haemostas 1978; 40: 272-287
- 7 Berendsen GE. Preparation and characterization of well-defined chemically bonded stationary phases for high pressure liquid chromotagraphy. Ph. D. thesis, Technische School Delft, The Netherlands: Delft University Press; 1980. p 8
- 8 Norén I, Blombäck M, Fridell E, Wallinder U. Vacutainer sampling for blood coagulation assays. Scand J Clin Lab Invest 1978; 38: 63-68