Endoscopy 2020; 52(06): 521-522
DOI: 10.1055/a-1073-7445
E-Videos
© Georg Thieme Verlag KG Stuttgart · New York

Complete colonic closure with an over-the-scope clip placed for a colorectal anastomotic fistula: a big mistake

Benedetto Mangiavillano
1   Gastrointestinal Endoscopy Unit, Humanitas Mater Domini, Castellanza, Varese, Italy
2   Humanitas University, Milan, Italy
,
Francesco Auriemma
1   Gastrointestinal Endoscopy Unit, Humanitas Mater Domini, Castellanza, Varese, Italy
,
Mario Bianchetti
1   Gastrointestinal Endoscopy Unit, Humanitas Mater Domini, Castellanza, Varese, Italy
,
Alessandro Repici
2   Humanitas University, Milan, Italy
3   Digestive Endoscopy Unit, Istituto Clinico Humanitas Research Hospital, Milan, Italy
› Author Affiliations
Further Information

Publication History

Publication Date:
20 December 2019 (online)

Colorectal postsurgical leaks and fistulas are severe complications that increase morbidity and mortality. The recent development of the over-the-scope clip (OTSC) has dramatically decreased the number of surgical reinterventions required [1].

A 73-year-old man with a history of anterior rectal resection for a T3N1 adenocarcinoma and subsequent neoadjuvant chemotherapy was referred to our unit because of a suspicion of a colorectal fistula, which had developed 30 days after the surgical intervention. A lower gastrointestinal endoscopy showed the colorectal anastomosis with a fistula (9 mm in diameter) at 5 cm from the anal verge.

We decided to close the defect with an 11/6 t OTSC (Ovesco, Tübingen, Germany). Using a gastroscope, the suction technique, and with the aid of the anchor device, we placed an OTSC over the fistula but, after the clip was released, we observed a complete closure of the colonic lumen, also evidence by an absence of air coming from the bowel. Because of the memory loss of the nitinol in the clip at temperatures under 4° C, we decided to irrigate the colonic lumen with cold water (< 4° C) for 10 minutes, until the color of the mucosa changed to white. Following this, we removed the OTSC using a rat-tooth forceps, without any complications. In the same session, following clip removal, we placed another 11/6 t OTSC with the aid of the anchor device, successfully sealing the fistula ([Video 1]).

Video 1 Removal and replacement of a previously placed over-the-scope clip that caused complete closure of the colonic lumen.


Quality:

In conclusion, endoscopic OTSC closure of colorectal postsurgical leaks and fistulas is a safe technique, with a high success rate in both acute and chronic cases. However complete bowel closure is a rare adverse event that can happen accidentally, especially in non-expert hands. The nitinol memory loss in water at < 4° C allowed OTSC deformation, which helped us in its removal.

Endoscopy_UCTN_Code_CPL_1AJ_2AG

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high
quality video and all contributions are
freely accessible online.

This section has its own submission
website at
https://mc.manuscriptcentral.com/e-videos