Contrast-enhanced low mechanical index endoscopic ultrasound (CELMI-EUS)

C.  F.  Dietrich1
  • 1Medical and Imaging Department, Caritas-Krankenhaus, Bad Mergentheim, Germany
Further Information

Publication History

Publication Date:
13 March 2009 (online)

This report describes contrast-enhanced low mechanical index endoscopic ultrasound (CELMI-EUS), a new real-time technique with high resolution.

A 73-year-old man presented with unspecific abdominal pain and suspected liver cirrhosis ([Fig. 1]) and a pancreatic mass. CELMI-EUS was performed using the Hitachi 900 HV device combined with an electronic echo endoscope (Hitachi/Pentax EG3870UTK; Hitachi, Tokyo, Japan) and adapted dynamic contrast harmonic wideband pulsed inversion software, with low mechanical index (0.09 – 0.30), before and up to 180 seconds after injection of 4.8 ml SonoVue (Bracco, Milan, Italy). To improve the contrast in microvessels, microbubble tracing imaging (MTI) was applied.

Enhancement of the liver parenchyma persisted for up to 5 minutes under continuous scanning conditions. No significant destruction of bubbles was visualized. CELMI-EUS revealed arterial rim enhancement and delineated central necrosis, typical signs of neuroendocrine metastases ([Fig. 2] and [Video 1]). Histological examination revealed neuroendocrine tumor metastasis.

While EUS has become established over the past 20 years, real-time contrast applications using low mechanical index were restricted to the transabdominal approach [2]. Recent advances in technology have supported the development of new endoscopic systems, making it possible to use CELMI-EUS as with a recently described prototype [3]. To distinguish between contrast medium signals and tissue signals, two ultrasound pulses with a 180° inverted phase are sent out at the same vector line. The received echo information from these two pulses is added, eliminating the linear information of the fundamental (tissue) signals. The nonlinear echo information of the contrast media can be displayed ([Fig. 3]) [3]. CELMI-EUS may improve our understanding of morphologic imaging methods through additional analysis of functional criteria such as perfusion, with better characterization of, for example, tumors, lymph node staging, and organ infiltration. This improved image documentation will enhance confidence in the method.

Fig. 1 Conventional B-mode imaging showed diffuse liver disease and ascites.

Fig. 2 Contrast-enhanced endoscopic ultrasound demonstrated peripheral rim enhancement of a liver lesion, which was shown histologically to be a metastasis of a neuroendocrine tumor of the pancreas. (See also [Video 1].)

Fig. 3 To distinguish between contrast medium signals and tissue signals, two ultrasound pulses with a 180° inverted phase are sent out at the same vector line. The received echo information from these two pulses is summed and the nonlinear echo information from the contrast medium is displayed.


Quality:

Video 1 Contrast-enhanced endoscopic ultrasound demonstrated peripheral rim enhancement of a liver lesion, which was shown histologically to be a metastasis of a neuroendocrine tumor of the pancreas. (See also Fig. 2.)

Endoscopy_UCTN_Code_CCL_1AF_2AG_3AD

References

  • 1 Mörk H, Ignee A, Schuessler G. et al . Analysis of neuroendocrine tumour metastases in the liver using contrast enhanced ultrasonography.  Scand J Gastroenterol. 2007;  42 652-662
  • 2 Claudon M, Cosgrove D, Albrecht T. et al . Guidelines and good clinical practice recommendations for contrast enhanced ultrasound (CEUS) – Update 2008.  Ultraschall Med. 2008;  29 28-44
  • 3 Dietrich C F, Ignee A, Frey H. Contrast enhanced endoscopic ultrasound with low mechanical index, a new technique.  Z Gastroenterol. 2005;  43 1219-1223

C. F. DietrichMD 

Innere Medizin 2
Caritaskrankenhaus Bad Mergentheim

Uhlandstr. 7
D-97980 Bad Mergentheim
Germany

Fax: +49-7931-58-2290

Email: christoph.dietrich@ckbm.de