Semin Musculoskelet Radiol 2015; 19(01): 012-020
DOI: 10.1055/s-0034-1396763
Thieme Medical Publishers 333 Seventh Avenue, New York, NY 10001, USA.

Radiographic Evaluation of Hip Implants

Connie Y. Chang
1   Division of Musculoskeletal Imaging and Intervention, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
,
Ambrose J. Huang
1   Division of Musculoskeletal Imaging and Intervention, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
,
William E. Palmer
1   Division of Musculoskeletal Imaging and Intervention, Massachusetts General Hospital, Harvard Medical School, Boston, Massachusetts
› Author Affiliations
Further Information

Publication History

Publication Date:
29 January 2015 (online)

Preview

Abstract

Serial radiographs are the mainstay in the longitudinal assessment of hip implants. The prosthesis, periprosthetic bone, and juxta-articular soft tissues are inspected for fracture, periosteal reaction, stress shielding, calcar resorption, osteolysis, bony remodeling, metallic debris, and heterotopic ossification. Comparison radiographs best confirm implant migration, subsidence, and aseptic loosening. Infection, particle disease, reaction to metal, and mechanical impingement are important causes of postsurgical pain, but in their earliest stages they may be difficult to diagnose using radiographs. This article addresses the role of radiography following hip arthroplasty.