Background and Study Aims: The accuracy of endoscopic ultrasound-guided fine-needle aspiration (EUS-FNA) depends on immediate specimen review by a cytopathologist. Stromal tumors, lymphoma, and well-differentiated pancreatic cancer are difficult to diagnose on the basis of cytology alone. To overcome these limitations, a 19-gauge Trucut needle has been developed to obtain histological samples at EUS. This pilot study compares the specimen adequacy and diagnostic accuracy of EUS-guided Trucut needle biopsy (EUS-TNB) with EUS-FNA.
Patients and Methods: A total of 18 patients underwent EUS-TNB and EUS-FNA. The specimen adequacy and diagnostic accuracy of the two techniques was compared. The technical performance and safety profile of the Trucut needle were also evaluated.
Results: The EUS-TNB specimen was adequate for evaluation in 15/18 patients compared with 18/18 with EUS-FNA (83 % vs. 100 %, not significant). The diagnostic accuracy of EUS-TNB was not significantly different from EUS-FNA (78 % vs. 89 %). Two complications were encountered: one patient developed mediastinitis and required surgery; another had immediate bleeding that was managed conservatively. One technical problem was encountered: the Trucut needle failed to deploy after two passes when a gastric stromal cell tumor was being biopsied.
Conclusion: The diagnostic accuracy of the new EUS-TNB is comparable to that of EUS-FNA. In our experience, the overall efficacy and safety profile of the Trucut needle appears modest.
References
-
1
Gress F G, Hawes R H, Savides T J. et al .
Endoscopic ultrasound-guided fine-needle aspiration biopsy using linear array and radial scanning endosonography.
Gastrointest Endosc.
1997;
45
243-250
-
2
Brandwein S L, Farrell J J, Centeno B A, Brugge W R.
Detection and tumor staging of malignancy in cystic, intraductal, and solid tumors of the pancreas by EUS.
Gastrointest Endosc.
2001;
53
722-727
-
3
Reed C E, Mishra G, Sahai A. et al .
Esophageal cancer staging: improved accuracy by endoscopic ultrasound of celiac lymph nodes.
Ann Thorac Surg.
1999;
67
319-321
-
4
Wiersema M J, Vazquez-Sequeiros E, Wiersema L M.
Evaluation of mediastinal lymphadenopathy with endoscopic US-guided fine-needle aspiration biopsy.
Radiology.
2001;
219
252-257
-
5
Ribeiro A, Vazquez-Sequeiros E, Wiersema L M. et al .
EUS-guided fine-needle aspiration combined with flow cytometry and immunocytochemistry in the diagnosis of lymphoma.
Gastrointest Endosc.
2001;
53
485-491
-
6
Erickson R A, Sayage-Rabie L, Beissner R S.
Factors predicting the number of EUS-guided fine-needle passes for diagnosis of pancreatic malignancies.
Gastrointest Endosc.
2000;
51
184-190
-
7
Brandt K R, Charboneau J W, Stephens D H. et al .
CT- and US-guided biopsy of the pancreas.
Radiology.
1993;
187
99-104
-
8
Welch T J, Sheedy P F, Johnson C D. et al .
CT-guided biopsy: prospective analysis of 1 000 procedures.
Radiology.
1989;
171
493-496
-
9
Harrison B D, Thorpe R S, Kitchener P G. et al .
Percutaneous Trucut lung biopsy in the diagnosis of localised pulmonary lesions.
Thorax.
1984;
39
493-499
-
10
Hatada T, Ishii H, Ichii S. et al .
Diagnostic value of ultrasound-guided fine-needle aspiration biopsy, core-needle biopsy, and evaluation of combined use in the diagnosis of breast lesions.
J Am Coll Surg.
2000;
190
299-303
-
11
Durup S cheel-Hincke, Mortensen M B, Pless T, Hovendal C P.
Laparoscopic four-way ultrasound probe with histologic biopsy facility using a flexible Tru-cut needle.
Surg Endosc.
2000;
14
867-869
-
12
Chau T N, Tong S W, Li T M. et al .
Transjugular liver biopsy with an automated Trucut-type needle: comparative study with percutaneous liver biopsy.
Eur J Gastroenterol Hepatol.
2002;
14
19-24
-
13
Indinnimeo M, Cicchini C, Stazi A. et al .
Trans anal full thickness Tru-cut needle biopsies in anal canal tumors after conservative treatment.
Oncol Rep.
1998;
5
325-327
-
14
Wiersema M J, Vilmann P, Giovannini M. et al .
Endosonography-guided fine-needle aspiration biopsy: diagnostic accuracy and complication assessment.
Gastroenterology.
1997;
112
1087-1095
-
15
Ribeiro A, Vasquez-Sequeiros E, Wiersema L M. et al .
EUS-guided fine-needle aspiration combined with flow-cytometry and immunohistochemistry in the diagnosis of lymphoma.
Gastrointest Endosc.
2001;
53
485-491
-
16
Levy M J, Jondal M L, Clain J, Wiersema M J.
Preliminary experience with an EUS-guided trucut biopsy needle compared with EUS-guided FNA.
Gastrointest Endosc.
2003;
57
101-106
-
17
Chang K J.
Maximizing the yield of EUS-guided fine-needle aspiration.
Gastrointest Endosc.
2002;
56
S28-S34
-
18
Binmoeller K F, Thul R, Rathod V. et al .
Endosonographic ultrasound-guided, 18 gauge, fine needle aspiration biopsy of the pancreas using a 2.8 mm channel convex array echoendoscope.
Gastrointest Endosc.
1998;
47
121-127
S. Varadarajulu, M. D.
Division of Gastroenterology-Hepatology · University of Alabama at Birmingham
Lyons-Harrison Research Bldg Room 410 · 1530, 3rd Ave S · Birmingham, AL · 35294-0007 · USA ·
Fax: +1-205-975-6381
eMail: svaradarajulu@yahoo.com