Endoscopy 2018; 50(12): E344-E345
DOI: 10.1055/a-0725-7718
E-Videos
© Georg Thieme Verlag KG Stuttgart · New York

Stent migration requiring surgical removal: a serious adverse event after bariatric megastent placement

Antonio C. Madruga Neto
1   Gastrointestinal Endoscopy Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
,
Vitor O. Brunaldi
1   Gastrointestinal Endoscopy Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
,
Ossamu Okazaki
1   Gastrointestinal Endoscopy Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
,
Marco A. Santo Filho
2   Bariatric and Metabolic Surgery Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
,
Antônio A. Miranda Neto
2   Bariatric and Metabolic Surgery Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
,
Vera L. Anapaz
3   Gastroenterology Department, Hospital Professor Doutor Fernando Fonseca, Lisboa, Portugal
,
Eduardo G. H. de Moura
1   Gastrointestinal Endoscopy Unit, Hospital das Clínicas, University of São Paulo Medical School, São Paulo, Brazil
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Publikationsdatum:
17. Oktober 2018 (online)

Leaks are serious complications after sleeve gastrectomy, with an incidence rate up to 7 % [1]. Currently, stent placement and other endoscopic techniques are recommended for the treatment of post-bariatric leaks (PBLs) [2]. Bariatric stents are a feasible, effective and life-saving method [3]. They achieve leak closure rates between 72.8 % and 87.8 % but are associated with significant migration rates (16.9 % – 28.2 %) [1] [4]. The large bariatric-specific stent (LBSS) has been developed as an important device to fit bariatric anatomy and avoid migration because of its long length with the distal edge being placed in the duodenum [5].

We report the case of 34-year-old woman who underwent sleeve gastrectomy. On the 5th post-operative day (POD), she developed abdominal pain and purulent output from the drain. Computed tomography (CT) scanning showed a leak at the angle of His without any collections.

We opted to place an LBSS (24 cm × 28 mm; Hanarostent, MITECH) to bypass the whole stomach ([Fig. 1]). The LBSS was removed 4 weeks later (33th POD), but the leak persisted. We then placed two single-pigtail stents ([Fig. 2]) and kept them in place for 10 days, but this was also unsuccessful. On the 43 rd POD, we removed the pigtail stents, performed a septotomy, and placed a second LBSS ([Fig. 3]).

Zoom Image
Fig. 1 Radiographic image showing the final position of the first megastent placed to treat a leak at the angle of His following sleeve gastrectomy.
Zoom Image
Fig. 2 Endoscopic image showing the two pigtail stents placed at the site of the leak.
Zoom Image
Fig. 3 Endoscopic image showing the second megastent in position.

After 3 weeks (64th POD), the patient again presented with abdominal pain. On this occasion, esophagogastroduodenoscopy (EGD) and upper gastrointestinal series showed complete closure of the leak, but the LBSS could not be seen. A CT scan identified the stent in the proximal jejunum ([Fig. 4]). An enteroscopic attempt to retrieve the stent and 7 days of conservative treatment (laxative, diet, and antispasmodic drug) were unsuccessful, and a laparoscopic enterectomy and primary anastomosis were eventually needed ([Fig. 5]). Finally, 15 days later (86th POD), she was asymptomatic and an EGD showed only a shallow ulcer at the angle of His ([Fig. 6]; [Video 1]).

Zoom Image
Fig. 4 Computed tomography scan showing the megastent located in the proximal jejunum.
Zoom Image
Fig. 5 Macroscopic appearance of the 27-cm enterectomy specimen.
Zoom Image
Fig. 6 Endoscopic image showing a shallow ulcer at the angle of His after closure of the leak.

Video 1 Insertion of the first bariatric megastent to treat a leak; removal of the megastent and insertion of two single-pigtail stents; removal of pigtail stents and insertion of a second megastent; the bariatric megastent is found to have migrated into the small bowel and eventually required surgical removal.


Qualität:

An LBSS seems to be a useful device for PBL management, but it may cause serious adverse events. Further controlled studies should assess the precise success and complication rates of such stent.

Endoscopy_UCTN_Code_CPL_1AH_2AK

Endoscopy E-Videos
https://eref.thieme.de/e-videos

Endoscopy E-Videos is a free access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos

 
  • References

  • 1 Okazaki O, Bernardo WM, Brunaldi VO. et al. Efficacy and safety of stents in the treatment of fistula after bariatric surgery: a systematic review and meta-analysis. Obes Surg 2018; 28: 1788-1796
  • 2 Kim J, Azagury D, Eisenberg D. et al. ASMBS position statement on prevention, detection, and treatment of gastrointestinal leak after gastric bypass and sleeve gastrectomy, including the roles of imaging, surgical exploration, and nonoperative management. Surg Obes Relat Dis 2015; 11: 739-748
  • 3 de Moura EG, Galvão-Neto MP, Ramos AC. et al. Extreme bariatric endoscopy: stenting to reconnect the pouch to the gastrojejunostomy after a Roux-en-Y gastric bypass. Surg Endosc 2012; 26: 1481-1484
  • 4 Puli SR, Spofford IS, Thompson CC. Use of self-expandable stents in the treatment of bariatric surgery leaks: a systematic review and meta-analysis. Gastrointest Endosc 2012; 75: 287-293
  • 5 Shehab H, Abdallah E, Gawdat K. et al. Large bariatric-specific stents and over-the-scope clips in the management of post-bariatric surgery leaks. Obes Surg 2018; 28: 15-24