Endoscopy 2020; 52(09): 754-760
DOI: 10.1055/a-1145-3562
Original article

A newly designed duodenoscope with detachable distal cap significantly reduces organic residue contamination after reprocessing

Wiriyaporn Ridtitid
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
2   Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
,
Piyapoom Pakvisal
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
,
Tanittha Chatsuwan
3   Department of Microbiology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
,
Stephen J. Kerr
4   Biostatistics Excellence Center, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
,
Kasenee Tiankanon
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
,
Panida Piyachaturawat
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
,
Parit Mekaroonkamol
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
,
Pradermchai Kongkam
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
2   Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
,
Rungsun Rerknimitr
1   Division of Gastroenterology, Faculty of Medicine, Chulalongkorn University and King Chulalongkorn Memorial Hospital, Thai Red Cross Society, Bangkok, Thailand
2   Center of Excellence for Innovation and Endoscopy in Gastrointestinal Oncology, Faculty of Medicine, Chulalongkorn University, Bangkok, Thailand
› Author Affiliations


Abstract

Background A newly designed duodenoscope with detachable distal cap may reduce bacterial contamination by allowing better access to the elevator. We compared bacterial contamination and organic residue evaluated by rapid adenosine triphosphate (ATP) test and culture from duodenoscopes with detachable vs. fixed distal caps after high-level disinfection (HLD).

Methods During December 2018–April 2019, 108 used newly designed duodenoscopes were enrolled. In group A (n = 54), the distal cap of the duodenoscope was detached before manual cleaning. In group B (n = 54), the distal cap was not detached. After HLD, samples were collected from the elevator, submitted for culture, and evaluated using the ATP test, using the cutoff value of 40 relative light units (RLUs).

Results After HLD, the proportion of potential bacterial contamination and organic residue in group A was significantly lower than in group B (37.0 % vs. 75.9 %; P  < 0.001; relative risk 0.49, 95 % confidence interval 0.33–0.71), and also confirmed by median ATP values (45.2 vs. 141.0 RLU; P  < 0.001). In group B, one sample culture was positive for nonpathogenic bacteria. Pathogenic bacteria were not found in any culture from either group.

Conclusions The detachable distal cap was more effective at eliminating bacterial contamination and reducing organic residue than a fixed cap. Nonpathogenic bacteria were detected in the fixed cap group after reprocessing. The ATP test with 40 RLU cutoff is a practical method to ensure the cleanliness of duodenoscope reprocessing without the need to wait for bacterial culture results.



Publication History

Received: 23 December 2020

Accepted: 10 March 2020

Article published online:
16 April 2020

© Georg Thieme Verlag KG
Stuttgart · New York

 
  • References

  • 1 Epstein L, Hunter JC, Arwady MA. et al. New Delhi metallo-beta-lactamase-producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes. JAMA 2014; 312: 1447-1455
  • 2 Wendorf KA, Kay M, Baliga C. et al. Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak. Infect Control Hosp Epidemiol 2015; 36: 634-642
  • 3 US Food and Drug Administration. FDA recommends health care facilities and manufacturers begin transitioning to duodenoscopes with disposable components to reduce risk of patient infection. Available from (Accessed 9.11.2019): https://www.fda.gov/news-events/press-announcements/fda-recommends-health-care-facilities-and-manufacturers-begin-transitioning-duodenoscopes-disposable
  • 4 Centers for Disease Control and Prevention. Healthcare-associated infections. Available from (Accessed 19.10.2018): https://www.cdc.gov/hai/index.html
  • 5 US Food and Drug Administration (FDA), the Centers for Disease Control and Prevention (CDC), and American Society for Microbiology (ASM) Working Group on Duodenoscope Culturing. Duodenoscope surveillance sampling and culturing protocols: reducing the risks of infection. Available from (Accessed 19.10.2018): https://www.fda.gov/media/111081/download
  • 6 Sethi S, Huang RJ, Barakat MT. et al. Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes. Gastrointest Endosc 2017; 85: 1180-1187
  • 7 Alfa MJ, Fatima I, Olson N. The adenosine triphosphate test is a rapid and reliable audit tool to assess manual cleaning adequacy of flexible endoscope channels. Am J Infect Control 2013; 41: 249-253
  • 8 Alfa MJ, Fatima I, Olson N. Validation of adenosine triphosphate to audit manual cleaning of flexible endoscope channels. Am J Infect Control 2013; 41: 245-248
  • 9 Quan E, Mahmood R, Naik A. et al. Use of adenosine triphosphate to audit reprocessing of flexible endoscopes with an elevator mechanism. Am J Infect Control 2018; 46: 1272-1277
  • 10 Hansen D, Benner D, Hilgenhoner M. et al. ATP measurement as method to monitor the quality of reprocessing flexible endoscopes. Ger Med Sci 2004; 2: Doc04
  • 11 Fernando G, Collignon P, Beckingham W. ATP bioluminescence to validate the decontamination process of gastrointestinal endoscopes. Healthc Infect 2014; 19: 59-64
  • 12 Olafsdottir LB, Wright SB, Smithey A. et al. Adenosine triphosphate quantification correlates poorly with microbial contamination of duodenoscopes. Infect Control Hosp Epidemiol 2017; 38: 678-684
  • 13 Batailler P, Saviuc P, Picot-Gueraud R. et al. Usefulness of adenosinetriphosphate bioluminescence assay (ATPmetry) for monitoring the reprocessing of endoscopes. Infect Control Hosp Epidemiol 2015; 36: 1437-1443
  • 14 Visrodia K, Hanada Y, Pennington KM. et al. Duodenoscope reprocessing surveillance with adenosine triphosphate testing and terminal cultures: a clinical pilot study. Gastrointest Endosc 2017; 86: 180-186
  • 15 Pakvisal P, Ridtitid W, Phathong C. et al. Sa1041 Performance characteristics and optimal cut-off value of adenosine triphosphate testing for rapid evaluation of bacterial contamination after reprocessing duodenoscopes. Gastrointest Endosc 2018; 87: AB155-AB156
  • 16 Pakvisal P, Ridtitid W, Kongkam P. et al. Sa1046 Performance characteristics of triple adenylate nucleotides (3A) versus adenosine triphosphate (ATP) alone as a quick surveillance for bacterial contamination after duodenoscope reprocessing. Gastrointest Endosc 2019; 89: AB153
  • 17 Beilenhoff U, Neumann CS, Rey JF. et al. ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy 2007; 39: 175-181
  • 18 Centers for Disease Control and Prevention. Interim protocol for healthcare facilities regarding surveillance for bacterial contamination of duodenoscopes after reprocessing 2015. Available from (Accessed 04.01.2018) https://www.cdc.gov/hai/pdfs/cre/interim-duodenoscope-surveillance-Protocol.pdf
  • 19 Snyder GM, Wright SB, Smithey A. et al. Randomized comparison of 3 high-level disinfection and sterilization procedures for duodenoscopes. Gastroenterology 2017; 153: 1018-1025
  • 20 Kim S, Russell D, Mohamadnejad M. et al. Risk factors associated with the transmission of carbapenem-resistant Enterobacteriaceae via contaminated duodenoscopes. Gastrointest Endosc 2016; 83: 1121-1129
  • 21 Kola A, Piening B, Pape UF. et al. An outbreak of carbapenem-resistant OXA-48-producing Klebsiella pneumonia associated to duodenoscopy. Antimicrob Resist Infect Control 2015; 4: 8
  • 22 Gastmeier P, Vonberg RP. Klebsiella spp. in endoscopy-associated infections: we may only be seeing the tip of the iceberg. Infection 2014; 42: 15-21
  • 23 Verfaillie CJ, Bruno MJ, Voor in t Holt AF. et al. Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa . Endoscopy 2015; 47: 502
  • 24 Beilenhoff U, Biering H, Blum R. et al. Prevention of multidrug-resistant infections from contaminated duodenoscopes: position statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA). Endoscopy 2017; 49: 1098-1106
  • 25 McCarthy M. FDA issues warnings to duodenoscope makers linked to infection outbreaks. BMJ 2015; 351: h4544
  • 26 Rauwers AW, Voor in t Holt AF, Buijs JG. et al. High prevalence rate of digestive tract bacteria in duodenoscopes: a nationwide study. Gut 2018; 67: 1637-1645
  • 27 Olafsdottir LB, Whelan J, Snyder GM. A systematic review of adenosine triphosphate as a surrogate for bacterial contamination of duodenoscopes used for endoscopic retrograde cholangiopancreatography. Am J Infect Control 2018; 46: 697-705
  • 28 Rex DK, Sieber M, Lehman GA. et al. A double-reprocessing high-level disinfection protocol does not eliminate positive cultures from the elevators of duodenoscopes. Endoscopy 2018; 50: 588-596
  • 29 Vickery K, Ngo QD, Zou J. et al. The effect of multiple cycles of contamination, detergent washing, and disinfection on the development of biofilm in endoscope tubing. Am J Infect Control 2009; 37: 470-475
  • 30 Pajkos A, Vickery K, Cossart Y. Is biofilm accumulation on endoscope tubing a contributor to the failure of cleaning and decontamination?. J Hosp Infect 2004; 58: 224-229