Klin Monbl Augenheilkd 2021; 238(09): 1004-1009
DOI: 10.1055/a-1179-0131
Klinische Studie

The Impact of Artificial Tear Viscosity on the Results of Optical Coherence Tomography

Article in several languages: English | deutsch
Elisa Hannemann
1   Optometrie, Fachbereich VII, Beuth Hochschule für Technik, Berlin
,
Carl Erb
2   Augenklinik am Wittenbergplatz, Berlin
› Author Affiliations

Abstract

Background In earlier studies, an irregular tear film could be responsible for artifacts in imaging procedures in ophthalmology. As a result, this study will investigate the influence of hyaluronic acid-containing tear substitutes of different viscosities on the measurement results of optical coherence tomography and on the non-invasive tear film break-up time.

Methods The study included three sessions in which three hyaluronic acid tear substitutes of different viscosities (Hylo-Vision 0.1%, 0.2%, 0.3%, OmniVision GmbH, Puchheim, Germany) were applied to each subject. In 20 healthy volunteers, the non-invasive tear film break-up time was measured before and after the application of the tear substitutes using Keratograph 5M (Oculus GmbH, Wetzlar, Germany) and optical coherence tomography using 3D OCT-2000 (Topcon, Hamburg, Germany).

Results The median age of the volunteers was 28.5 years. 11 women (55%) and 9 men (45%) were examined. The non-invasive tear film break-up time improved significantly (p = 0.027) after the application of 0.3% hyaluronic acid. In the study, no significant difference was found in all OCT parameters used before and after application of the different viscous hyaluronic acid-containing tear substitutes.

Conclusion In this study, no influence of tear substitutes containing different viscosities of hyaluronic acid on the measurement results of optical coherence tomography could be determined. Therefore, 0.1%, 0.2% and 0.3% hyaluronic acid can be applied to the patient to improve the corneal surface before the examination with optical coherence tomography, without influencing the measurement results of optical coherence tomography.



Publication History

Received: 24 September 2019

Accepted: 16 April 2020

Article published online:
09 October 2020

© 2020. Thieme. All rights reserved.

Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany

 
  • References/Literatur

  • 1 Erb C, Gast U, Schremmer D. German register for glaucoma patients with dry eye. I. Basic outcome with respect to dry eye. Graefes Arch Clin Exp Ophthalmol 2008; 246: 1593-1601
  • 2 Leung EW, Medeiros FA, Weinreb RN. Prevalence of ocular surface disease in glaucoma patients. J Glaucoma 2008; 17: 350-355
  • 3 Kaercher T. Epidemiologie des Trockenen Auges. In: Kunert KS, Sickenberger W, Brewitt H. Hrsg. Trockenes Auge. Heidelberg: KADEN; 2016: 43-44 46–50
  • 4 Brewitt H, Polzer H, Zippel F. Medikamentöse Substitutionstherapie des Trockenen Auges – Klinische und pharmakologische Aspekte. In: Kunert KS, Sickenberger W, Brewitt H. Hrsg. Trockenes Auge. Heidelberg: KADEN; 2016: 179-192
  • 5 Rüfer F, Erb C. Einflüsse des trockenen Auges auf die Glaukomdiagnostik. Ophthalmologe 2012; 109: 1082-1086
  • 6 Asrani S, Edghill B, Gupta Y. et al. Optical coherence tomography errors in glaucoma. J Glaucoma 2010; 19: 237-242
  • 7 Sickenberger W, Oehring D. Diagnostik des Trockenen Auges. In: Kunert KS, Sickenberger W, Brewitt H. Hrsg. Trockenes Auge. Heidelberg: KADEN; 2016: 137-172
  • 8 Tian L, Qu JH, Zhang XY. et al. Repeatability and reproducibility of noninvasive Keratograph 5M measurements in patients with dry eye disease. J Ophthalmol 2016; DOI: 10.1155/2016/8013621.
  • 9 Sweeney DF, Millar TJ, Raju SR. Tear film stability: a review. Exp Eye Res 2013; 117: 28-38
  • 10 Mengher LS, Bron AJ, Tonge SR. et al. A non-invasive instrument for clinical assessment of the pre-corneal tear film stability. Curr Eye Res 1985; 4: 1-7
  • 11 Mengher LS, Pandher KS, Bron AJ. et al. Effect of sodium hyaluronate (0,1 %) on break-up time (NIBUT) in patients with dry eyes. Br J Ophthalmol 1986; 70: 442-447
  • 12 Johnson ME, Murphy PJ, Boulton M. Effectiveness of sodium hyaluronate eyedrops in the treatment of dry eye. Graefes Arch Clin Exp Ophthalmol 2006; 244: 109-112
  • 13 Ehnes A. Entwicklung eines Schichtsegmentierungsalgorithmus zur automatischen Analyse von individuellen Netzhautschichten in optischen Kohärenztomographie-B-Scans [Inauguraldissertation]. Gießen: Justus-Liebig-Universität Gießen; 2013: 10-27
  • 14 Ghazi NG, Much JW. The effect of lubricating eye drops on optical coherence tomography imaging of the retina. Digit J Ophthalmol 2009; 15: 24-26 doi:10.5693/djo.01.2009.006
  • 15 Stein DM, Wollstein G, Ishikawa H. et al. Effect of corneal drying on optical coherence tomography. Ophthalmology 2006; 113: 985-991