Endoscopic submucosal dissection of an ileocecal valve laterally spreading tumor using an adaptable magnetic traction system
Traction-assistance is very effective in facilitating endoscopic submucosal dissection (ESD) of superficial colorectal lesions [1], notably in the case of difficult locations such as the ileocecal valve. We have recently developed a multitraction strategy to increase the efficacy of traction for lesions in such locations, using a rubber band to connect the different traction points [2]. A new magnetic traction device (ProdiGI Traction Magnet, Medtronic) has been developed that allows adjustment of the traction strategy during the procedure ([Fig. 1 a, b]; [Video 1]).
Video 1 Endoscopic submucosal dissection (ESD) of a laterally spreading tumor at the ileocecal valve, using an adaptable magnetic traction system.
Quality:
We report here the case of a 64-year-old patient with a 2-cm nongranular laterally spreading tumor of the cecum, extending to the lower lip of the ileocecal valve. In order to make the incision with satisfactory margins, the first step of the ESD procedure consisted in exposing the inner part of the valve using a first clip as previously described [3], by connecting the magnet attached to the first clip to the magnet of a second clip placed on the opposite colonic wall ([Fig. 2]). Once the incision had been made, we had to expose the submucosa for the dissection itself, with a third clip placed on the edge of the lesion, that was magnetically connected to the clip on the opposite colonic wall. At this point, the first clip used to pull the inner part of the valve to allow incision on the ileal side had become troublesome because the traction angle was not optimal for submucosal exposure. Therefore, we separated the magnet of the first clip from the two others using the cap of the endoscope, in order to obtain optimal traction for the dissection itself ([Fig. 3]). This technique allowed a curative R0 resection of the lesion ([Fig. 4]).
To our knowledge, this is the first case in humans of the use of a fully internalized magnetic traction system that can pass through the operating channel of the endoscope [4]. This technique seems attractive in difficult locations where a change in traction configuration is necessary during the procedure. Further studies are needed to confirm its effectiveness.
Endoscopy_UCTN_Code_TTT_1AQ_2AD
Endoscopy E-Videos is an open access online section, reporting on interesting cases and new techniques in gastroenterological endoscopy. All papers include a high quality video and all contributions are freely accessible online. Processing charges apply (currently EUR 375), discounts and wavers acc. to HINARI are available.
This section has its own submission website at https://mc.manuscriptcentral.com/e-videos
Publication History
Article published online:
25 August 2022
© 2022. The Author(s). This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Bordillon P, Pioche M, Wallenhorst T. et al. Double-clip traction for colonic endoscopic submucosal dissection: a multicenter study of 599 consecutive cases (with video). Gastrointest Endosc 2021; 94: 333-343
- 2 Lafeuille P, Rivory J, Jacques J. et al. Diagnostic endoscopic submucosal dissection for invasive cancer with the four cardinal points traction strategy. Endoscopy 2021; 54: E281-E282
- 3 Wallenhorst T, Pioche M, Bouguen G. et al. Ileocecal valve opening with double clip and rubber band for countertraction facilitates R0 en bloc resection of laterally spreading tumors. Endoscopy 2020; 52: E390-E391
- 4 Matsuzaki I, Yamauchi H, Goto N. et al. Magnetic anchor-guided endoscopic submucosal dissection using a stainless steel anchor. Endoscopy 2020; 52: E80-E81