RSS-Feed abonnieren
DOI: 10.1055/a-2053-9238
Spontaneous detachment of the spiral overtube from the endoscope in the upper esophagus: a case report and literature review of a rare complication during motorized spiral enteroscopy
Spontanes Ablösen der Spirale vom Endoskop im oberen Ösophagus: Ein Fallbericht und Literaturüberblick einer seltenen Komplikation während der motorisierten SpiralenteroskopieAbstract
Background Motorized spiral enteroscopy (MSE) is a recent modality for diagnostic and therapeutic examination of the small bowel using a spiral-shaped overtube with an integrated electric motor attached to an endoscope. With growing evidence of its efficacy, MSE is increasingly being used by endoscopic facilities. We herein present an uncommon case of the detachment of the spiral overtube from the endoscope during MSE.
Case Report We performed antegrade MSE in a 75-year-old female under general anesthesia for coagulation of multiple small-bowel arteriovenous malformations and resection of a polyp detected on a previous capsule endoscopy. The management was successful, and we had no passage difficulties. However, during the withdrawal phase of the enteroscope, the entire spiral overtube disconnected from the endoscopic shaft and became stuck in the esophagus and pharynx. Attempts to remove the lodged spiral using various endoscopic forceps, graspers, snares, or a dilatation balloon failed. Also, the laryngoscopic use of Magill forceps did not allow grabbing the spiral properly. Eventually, the spiral could be retrieved laryngoscopically with arthroscopic grasping forceps. This incident caused laryngeal swelling and bleeding lacerations of the upper esophagus, which were clipped on a subsequent gastroscopy. No further long-lasting complications occurred in the patient.
Conclusion Detachment of the spiral from the enteroscope is a potentially life-threatening and challenging complication of MSE. Artificial ventilation under observation by an anesthesiologist allowed for calm management of this dangerous situation and increased the overall safety of the procedure. Hence, we believe that in the context of antegrade MSE, general anesthesia with endotracheal intubation should be mandatory.
Zusammenfassung
Hintergrund Die motorisierte Spiralenteroskopie (MSE) ist eine neue Technik zur diagnostischen und therapeutischen Untersuchung des Dünndarms. Hierbei verwendet man einen spiralförmigen Übertubus mit integriertem Elektromotor, welcher an ein Endoskop gekoppelt ist. Aufgrund zahlreicher evidenzbasierter Belege ihrer Praktikabilität wird die MSE zunehmend in endoskopischen Einrichtungen eingesetzt. Vor diesem Hintergrund präsentieren wir hier einen ungewöhnlichen Fall, bei dem sich der oben genannte Spiralübertubus während einer MSE gänzlich vom Endoskop löste.
Kasuistik Wir führten eine antegrade MSE bei einer 75-jährigen Patientin unter Vollnarkose zur Koagulation mehrerer Angioektasien und zur Resektion eines Polypen durch. Diese waren bei einer vorherigen Kapselendoskopie aufgefallen. Das Management verlief erfolgreich und wir hatten keine Schwierigkeiten bei der Passage mit dem Endoskop. Während der Rückzugsphase des Enteroskops löste sich jedoch der gesamte Spiralübertubus vom Schaft des Endoskops und blieb im Ösophagus und Pharynx stecken. Versuche, die festsitzende Spirale mit verschiedenen endoskopischen Zangen, Greifern, Schlingen oder einem Dilatationsballon zu bergen, schlugen fehl. Auch laryngoskopisch konnte eine Magillzange die Spirale nicht suffizient greifen. Schließlich konnte sie dann aber mit einer Arthroskopiezange geborgen werden. Der Zwischenfall verursachte ein Larynxödem und blutende Lazerationen im Ösophagus, die bei einer anschließenden Gastroskopie geklippt wurden. Weitere langanhaltende Komplikationen traten bei der Patientin nicht auf.
Schlussfolgerung Das Ablösen der Spirale vom Endoskop ist eine potenziell lebensbedrohliche und schwierig zu bewältigende Komplikation der MSE. Die Vollnarkose unter anästhesistischer Supervision ermöglichte einen ruhigen Umgang und erleichterte das Management dieser gefährlichen Situation. Daher sind wir der Auffassung, dass bei antegraden MSEs die Vollnarkose mit endotrachealer Intubation obligatorisch sein sollte.
Keywords
Antegrade motorized spiral enteroscopy - Enteroscopy - Spiral detachment - Endoscopy complication - General anesthesia - Endotracheal intubation - PolypSchlüsselwörter
Antegrade Motorisierte Spiralenteroskopie - Enteroskopie - Ablösen der Spirale - Endoskopische Komplikation - Vollnarkose - Endotracheale Intubation - PolypPublikationsverlauf
Eingereicht: 01. Februar 2023
Angenommen: 06. März 2023
Artikel online veröffentlicht:
12. April 2023
© 2023. Thieme. All rights reserved.
Georg Thieme Verlag KG
Rüdigerstraße 14, 70469 Stuttgart, Germany
-
References
- 1 Beyna T, Arvanitakis M, Schneider M. Total motorized spiral enteroscopy: first prospective clinical feasibility trial. Gastrointest Endosc 2021; 93: 1362-1370 DOI: 10.1016/j.gie.2020.10.028. (PMID: 33144239)
- 2 Neuhaus H, Beyna T, Schneider M. et al. Novel motorized spiral enteroscopy: first clinical case. VideoGIE 2016; 1: 32-33 DOI: 10.1016/j.vgie.2016.08.005. (PMID: 29905207)
- 3 Schneider M, Höllerich J, Gerges C. et al. Motorized spiral enteroscopy-assisted ERCP in surgically altered anatomy: early experience from a retrospective cohort study. Endoscopy 2023; DOI: 10.1055/a-1964-2100.
- 4 Steinbrück I, Allgaier HP. [Erstbeschreibung einer kompletten Enteroskopie von der Ileozökalklappe bis zum Pylorus durch eine retrograde motorisierte Spiralenteroskopie]. Z Gastroenterol 2022; DOI: 10.1055/a-1949-8148. (PMID: 36343640)
- 5 Pennazio M. Safety of motorized spiral enteroscopy: is what we know today enough?. Endoscopy 2022; 54: 1156-1157 DOI: 10.1055/a-1881-4691. (PMID: 35858637)
- 6 Despott E, Murino A. The return of the helix: an evaluation of motorized spiral enteroscopy in actual daily clinical practice; a few points to “torque” about and ponder upon…: Referring to Al-Toma A et al. Endoscopy 2022; 54 DOI: 10.1055/a-1866-3917. (PMID: 35803274)
- 7 Beyna T, Arvanitakis M, Schneider M. Motorised spiral enteroscopy: first prospective clinical feasibility study. Gut 2021; 70: 261-267 DOI: 10.1136/gutjnl-2019-319908. (PMID: 32332141)
- 8 Ramchandani M, Rughwani H, Inavolu P. Diagnostic yield and therapeutic impact of novel motorized spiral enteroscopy in small-bowel disorders: a single-center, real-world experience from a tertiary care hospital (with video). Gastrointest Endosc 2021; 93: 616-626 DOI: 10.1016/j.gie.2020.07.001. (PMID: 32663489)
- 9 Al-Toma A, Beaumont H, Koornstra JJ. The performance and safety of motorized spiral enteroscopy, including in patients with surgically altered gastrointestinal anatomy: a multicenter prospective study. Endoscopy 2022; DOI: 10.1055/a-1783-4802.
- 10 Prasad M, Prasad VG, Sangameswaran A. A spiraling journey into the small bowel: a case series of novel motorized power spiral enteroscopies. VideoGIE 2020; 5: 591-596 DOI: 10.1016/j.vgie.2020.08.001. (PMID: 33204928)
- 11 Buchholz H, Mende M, Hornoff S. Ergebnisse der motorisierten Spiralenteroskopie des Dünndarms an 83 konsekutiven Patienten [Results of motorized spiral enteroscopy in 83 consecutive patients]. Z Gastroenterol 2022; DOI: 10.1055/a-1744-6738.
- 12 Beyna T, Moreels T, Arvanitakis M. Motorized spiral enteroscopy: results of an international multicenter prospective observational clinical study in patients with normal and altered gastrointestinal anatomy. Endoscopy 2022; 54: 1147-1155 DOI: 10.1055/a-1831-6215. (PMID: 35451040)
- 13 Singh P, Singla V, Bopanna S. et al. Safety and efficacy of the novel motorized power spiral enteroscopy: A single-center experience. DEN Open 2023; 3: e148 DOI: 10.1002/deo2.148. (PMID: 35898825)
- 14 Pal P, Vishwakarma P, Singh AP. et al. Diagnostic yield and technical performance of novel motorized spiral enteroscopy compared to single-balloon enteroscopy in suspected Crohn’s disease: a prospective study (with videos). Gastrointest Endosc 2022; DOI: 10.1016/j.gie.2022.10.017.
- 15 Giordano A, Casanova G, Escapa M. et al. Motorized Spiral Enteroscopy Is Effective in Patients with Prior Abdominal Surgery. Dig Dis Sci 2022; DOI: 10.1007/s10620-022-07688-1.
- 16 Chan W, Wei LK, Tan T. et al. Motorized spiral enteroscopy versus double-balloon enteroscopy: a case-matched study. Gastrointest Endosc 2023; 97: 314-324 DOI: 10.1016/j.gie.2022.09.030. (PMID: 36228697)
- 17 Falt P, Tichý T, Urban O. Motorized spiral enteroscopy – Our first experience. Gastroenterologie a Hepatologie 2020; 74: 212-216 DOI: 10.14735/amgh2020212.
- 18 Liang R, Kalaiyarasi K, Ng WK. et al. Motorised spiral enteroscopy: pilot experience from a tertiary care centre in Singapore. Singapore Med J 2022; DOI: 10.11622/smedj.2022096.
- 19 Steiner E, Mayer A, Maieron A. Motorized spiral enteroscopy – when things happen: troubleshooting spiral disconnection at 90 minutes into the enteroscopic procedure. Endoscopy 2021; DOI: 10.1055/a-1532-1639.
- 20 Muñoz Pérez R, Compañy Catalá L, Ruiz Gómez F. Disconnection of spiral overtube: an infrequent complication of motorized spiral enteroscopy. Endoscopy 2022; 54: S69 DOI: 10.1055/s-0042-1744719.
- 21 Barthet L, Benech N, Lupu A. et al. Detachment of an enteroscope spiral fragment during anterograde motorized spiral enteroscopy. Endoscopy 2023; 55: E31-e32 DOI: 10.1055/a-1930-5846.
- 22 Rajput S, Bhatt P, Bachkaniwala V. et al. Spontaneous Migration of Spiral Fin Overtube During Motorized Spiral Enteroscopy. ACG Case Rep J 2022; 9: e00808 DOI: 10.14309/crj.0000000000000808.