Antibodies for the radioimmunoassay of arginine vasopressin (AVP) described here were produced in rabbits using synthetic AVP coupled to rabbit γ-globulin with carbodiimide. In three out of six rabbits, significant antibody titres were obtained. Using the best antisera produced, 40% of labeled AVP was bound at a final dilution of 1:50.000. After iodination of synthetic AVP with 125I using the chloramin-T method, a gel filtration on Sephadex G-25 was performed to purify the iodinated AVP. For separation of antibody bound and free hormone, a second antibody precipitation was used. There was no crossreactivity with oxytocin. AVP was extracted from plasma after ammoniumsulfate precipitation of the proteins by adsorption to Florisil. The recovery of AVP added to plasma in amounts between 5-25 pg/ml was 60 ± 15% (n=6). The minimum amount of AVP detectable was 1 pg per ml plasma. The plasma level in normal adults under standard conditions was 3.4 ± 2.2 pg/ml. This is in agreement with data recently published by other researchers. The applicability and reproducibility was further tested in measurements of samples taken hourly during the entire day under water diuresis and after hormonal stimulation of AVP.
Key words
Vasopressin - Radioimmunoassay - Secretion Pattern
1 A preliminary report of part of this work was published in Acta endocr. (Kbh.) Suppl. 184 (1974) 52.
1 A preliminary report of part of this work was published in Acta endocr. (Kbh.) Suppl. 184 (1974) 52.