Endoscopes, including duodenoscopes, are medical devices that are frequently associated with outbreaks of nosocomial infections. We investigated an outbreak of multidrug-resistant Pseudomonas aeruginosa sepsis affecting three patients after endoscopic retrograde cholangiopancreaticography (ERCP). Epidemiologic investigation supplemented by molecular typing revealed that one ERCP scope was the source of infection with P. aeruginosa. No contamination with this microorganism was found after screening of washer-disinfectors, connecting tubes, and environmental surfaces in the endoscopy center. Pseudomonas isolates from blood and endoscope channels before gas sterilization with ethylene oxide (ETO) were characterized by molecular typing as ”linked isolates”. Though the current surveillance system did not prevent the infections in three patients, our microbiological surveillance protocol with routine culturing of endoscopes was helpful in detecting the source of contamination and probably avoided numerous cross-contaminations in other patients who underwent ERCP procedures with endoscopes.
References
- 1 Kimmey M B, Burnett D A, Carr-Locke D L. et al . Transmission of infection by gastrointestinal endoscopy. Gastrointest Endosc. 1993; 36 885-888
- 2 Srinivasan A. Epidemiology and prevention of infections related to endoscopy. Curr Inf Dis Rep. 2003; 5 467-472
- 3 Gastroenterological Nurses College of Australia, Gastroenterological Society of Australia .Infection control in endoscopy. 2nd edition. 2003. Digestive Health Foundation. Available at: http://www.gesa.org.au/pdf/booklets/I_Control_2nd_Edition.pdf; Publication date: March 2006
- 4 Banerjee S, Shen B, Baron T H. et al . ASGE Standards of Practice Committee. Antibiotic prophylaxis for gastrointestinal endoscopy. Gastrointest Endosc. 2008; 67 791-798
- 5 Buss A J, Been M H, Borgers R P. et al . Endoscope disinfection and its pitfalls – requirement for retrograde surveillance cultures. Endoscopy. 2008; 40 327-332
- 6 Pajkos A, Vickery K, Cossart Y. Is biofilm accumulation on endoscope tubing a contributor to the failure of cleaning and decontamination?. J Hosp Infect. 2004; 58 224-229
-
7 Guideline of the Dutch Infection Prevention Working Party (WIP) for cleaning and disinfection of endoscopes. Revision: April 2009. Available from: http://www.wip.nl/UK/disclaim.asp?url=http://www.wip.nl/UK/free_content/Richtlijnen/Cleaning%20and%20disinfection%20endoscopes070731.pdf; Publication date: April 2004
- 8 Beilenhoff U, Neumann C S, Rey J F. et al . ESGE-ESGENA guideline: cleaning and disinfection in gastrointestinal endoscopy. Endoscopy. 2008; 40 939-957
- 9 Doléans-Jordheim A, Cournoyer B, Bergeron E. et al . Reliability of Pseudomonas aeruginosa semi-automated rep-PCR genotyping in various epidemiological situations. Eur J Clin Microbiol Infect Dis. DOI: DOI 10.1007/s10096-009-0755-z Available from: http://www.springerlink.com/content/r5p87673273j4021/fulltext.pdf; Publication date: 16 May 2009
- 10 Nelson D B, Muscarella L F. Current issues in endoscope reprocessing and infection control during gastrointestinal endoscopy. World J Gastroenterol. 2006; 12 3953-3964
- 11 Srinivasan A, Wolfenden L L, Song X. et al . An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes. N Engl J Med. 2003; 348 221-227
- 12 Kirschke D L, Jones T F, Craig A S. et al . Pseudomonas aeruginosa and Serratia marcescens contamination associated with a manufacturing defect in bronchoscopes. N Engl J Med. 2003; 348 214-220
- 13 Corne P, Godreuil S, Jean-Pierre H. et al . Unusual implication of biopsy forceps in outbreaks of Pseudomonas aeruginosa infections and pseudo-infections related to bronchoscopy. J Hosp Infect. 2005; 61 20-26
- 14 Alfa M J, DeGagne P, Olson N. et al . Comparison of ion plasma, vaporized hydrogen peroxide, and 100 % ethylene oxide sterilizers to the 12 / 88 ethylene oxide gas sterilizer. Infect Cont Hosp Epidemiol. 1996; 17 92-100
- 15 Marion K, Freney J, Bergeron E. et al . Using an efficient biofilm detaching agent: an essential step for the improvement of endoscope reprocessing protocols. J Hosp Infect. 2006; 64 136-142
- 16 Rutala W A, Weber D J. New disinfection and sterilization methods. Emerg Infect Dis. 2001; 7 348-353
- 17 Colt H G, Beamis J J, Harrell J H. et al . Novel flexible bronchoscope and single-use disposable-sheath endoscope system. A preliminary technology evaluation. Chest. 2000; 1181 183-187
- 18 Systchenko R, Marchetti B, Canard J N. et al . Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy. Endoscopy. 2000; 32 807-818
- 19 Beilenhoff U, Neumann C S, Rey J F. et al . ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy. 2007; 39 175-181
- 20 Gillespie E E, Kotsanas D, Stuart R L. Microbiological monitoring of endoscopes: 5-year review. J Gastroenterol Hepatol. 2008; 23 1069-1074
J. KovalevaMD
Department of Medical Microbiology
University Medical Centre Groningen
PO Box 300019700 RB Groningen
The Netherlands
Fax: +31-50-3619105