Potential Risk for Intra-Aortic Balloon-Induced Obstruction to the Celiac Axis or the Renal Artery in the Asian Population
Publication History
received July 16, 2010
Publication Date:
07 March 2011 (online)
Abstract
To evaluate the potential risk for intra-aortic balloon (IAB)-induced obstruction to the celiac axis (CA) or the renal artery (RA) when IAB size is chosen according to patient height and the tip is placed 2 cm distal to the origin of the left subclavian artery (LSCA), the computed tomography images of 150 Asian adults were reviewed to identify the distance from the LSCA to the CA (LSCA‐CA) and to the RA (LSCA‐RA). The diameter of the aorta at the level of the T9 vertebra, CA, and RA was also measured. The length and caliber of the IAB was selected according to two manufacturer's recommendations (Datascope Corp and Tokai Medical Products). The Datascope IAB potentially blocked the CA in 84 % and the RA in 66 %, while the Tokai IAB obstructed the CA in 61 % and the RA in 10 % of cases. The caliber of the IAB overlying the orifice of the RA was longer than the diameter of the aorta in 5 cases (3.3 %) using the Datascope IAB, and in 1 case (0.7 %) using the Tokai IAB. When placing an IAB selected based on patient height, the IAB could obstruct the orifice of the CA and RA in the majority of Asian patients, especially a Datascope IAB.
Key words
aortic disease - heart disease - myocardial protection
References
- 1 Trost J C, Hillis L D. Intra-aortic balloon counterpulsation. Am J Cardiol. 2006; 97 1391-1398
- 2 Sasaki H. The right gastroepiploic artery in coronary artery bypass grafting. J Card Surg. 2008; 23 398-407
- 3 Roquette J. Intraaortic balloon position. Ann Thorac Surg. 1993; 55 199
- 4 Shin H, Yozu R, Sumida T et al. Acute ischemic hepatic failure resulting from intraaortic balloon pump malposition. Eur J Cardiothorac Surg. 2000; 17 492-494
- 5 Papaioannou T G, Stefanadis C. Basic principles of the intraaortic balloon pump and mechanisms affecting its performance. ASAIO J. 2005; 51 296-300
- 6 Swartz M T, Sakamoto T, Arai H et al. Effects of intraaortic balloon position on renal artery blood flow. Ann Thorac Surg. 1992; 53 604-610
- 7 Boehmer J P, Popjes E. Cardiac failure: mechanical support strategies. Crit Care Med. 2006; 34 S268-S277
- 8 Cho Y S, Lim C, Han M J et al. Should we consider the ethnic difference in selecting size of intraaortic balloon by commercial guideline?. ASAIO J. 2009; 55 519-522
- 9 Karlson K B, Martin E C, Bregman D et al. Superior mesenteric artery obstruction by intraaortic counterpulsation balloon simulating embolism: a case report. Cardiovasc Intervent Radiol. 1981; 4 236-238
- 10 Venkateswaran R V, Charman S C, Goddard M et al. Lethal mesenteric ischaemia after cardiopulmonary bypass: a common complication?. Eur J Cardiothorac Surg. 2002; 22 534-538
- 11 Jarmolowski C R, Poirier R L. Small bowel infarction complicating intra-aortic balloon counterpulsation via the ascending aorta. J Thorac Cardiovasc Surg. 1980; 79 735-737
- 12 O'Rourke M F, Shepherd K M. Protection of the aortic arch and subclavian artery during intra-aortic balloon pumping. J Thorac Cardiovasc Surg. 1973; 65 543-546
- 13 Yahel J, Arensburg B. The topographic relationships of the unpaired visceral branches of the aorta. Clin Anat. 1998; 11 304-309
- 14 Pennington N, Soames R W. The anterior visceral branches of the abdominal aorta and their relationship to the renal arteries. Surg Radiol Anat. 2005; 27 395-403
- 15 Pearce W H, Slaughter M S, LeMaire S et al. Aortic diameter as a function of age, gender, and body surface area. Surgery. 1993; 114 691-697
Prof. Jin-Tae Kim
Department of Anesthesiology and Pain Medicine
Seoul National University Hospital
101 DaeHakRo, JongRoGu
110-744 Seoul
Republic of Korea
Phone: +82 22 01 12 32 95
Fax: +82 27 45 55 87
Email: jinoben@naver.com