Endoscopy 2013; 45(12): 997-1005
DOI: 10.1055/s-0033-1344985
Original article
© Georg Thieme Verlag KG Stuttgart · New York

In vitro evaluation of the radial and axial force of self-expanding esophageal stents

Meike M. C. Hirdes
1   Department of Gastroenterology and Hepatology, University Medical Center Utrecht, Utrecht, The Netherlands
,
Frank P. Vleggaar
1   Department of Gastroenterology and Hepatology, University Medical Center Utrecht, Utrecht, The Netherlands
,
Matthieu de Beule
2   FEops bvba, Ghent, Belgium
3   IBiTech-bioMMeda, Faculty of Engineering and Architecture, Ghent University, Ghent, Belgium
,
Peter D. Siersema
1   Department of Gastroenterology and Hepatology, University Medical Center Utrecht, Utrecht, The Netherlands
› Author Affiliations
Further Information

Publication History

submitted 10 March 2013

accepted after revision 15 July 2013

Publication Date:
28 November 2013 (online)

Background and study aims: Technological innovation in esophageal stent design has progressed over the past decades, but the association between the mechanical properties of stent design and clinical outcome is still poorly understood. In this study the radial force and axial force of currently available stent designs were evaluated using an in vitro testing model.

Methods: A total of 10 partially and fully covered self-expanding metal stents (SEMSs), a self-expanding plastic stent (SEPS), and an uncovered biodegradable stent were evaluated. Radial force and axial force were measured using a radial force measurement machine (RX500) and a force gauge in an oven at 37°C.

Results: A wide range of radial force measurements were observed between the different stent designs, ranging from 4 to 83 N at 15 mm expansion. All braided nitinol stents displayed comparable mechanical characteristics with a relatively low radial force (< 150 N) that gradually decreased to 0 N during expansion, whereas plastic and metal stents that were constructed in a nonbraided manner displayed an initially high radial force (> 300 N) followed by a steep decline to 0 N during expansion. Conversely, peak axial force was relatively high for braided nitinol SEMSs (> 1.5 N), whereas nonbraided SEMSs showed a much lower peak axial force (< 1.5 N). Based on radial and axial force data, five groups of stents with comparable mechanical properties could be distinguished.

Conclusions: All currently available stents have a characteristic radial and axial force pattern, which may aid in the understanding of the occurrence of specific symptoms and complications after stent placement. Nonetheless, the overall clinical behavior of a stent is probably more complex and cannot be explained by these factors alone.

 
  • References

  • 1 Sharma P, Kozarek R. Role of esophageal stents in benign and malignant diseases. Am J Gastroenterol 2010; 105: 258-273
  • 2 Hirdes MM, Vleggaar FP, Siersema PD. Stent placement for esophageal strictures: an update. Expert Rev Med Devices 2011; 8: 733-755
  • 3 Schembre DB. Recent advances in the use of stents for esophageal disease. Gastrointest Endosc Clin N Am 2010; 20: 103-121, vii
  • 4 Verschuur EM, Repici A, Kuipers EJ et al. New design esophageal stents for the palliation of dysphagia from esophageal or gastric cardia cancer: a randomized trial. Am J Gastroenterol 2008; 103: 304-312
  • 5 Conio M, Repici A, Battaglia G et al. A randomized prospective comparison of self-expandable plastic stents and partially covered self-expandable metal stents in the palliation of malignant esophageal dysphagia. Am J Gastroenterol 2007; 102: 2667-2677
  • 6 van Boeckel PG, Siersema PD, Sturgess R et al. A new partially covered metal stent for palliation of malignant dysphagia: a prospective follow-up study. Gastrointest Endosc 2010; 72: 1269-1273
  • 7 van Boeckel PG, Repici A, Vleggaar FP et al. A new metal stent with a controlled-release system for palliation of malignant dysphagia: a prospective, multicenter study. Gastrointest Endosc 2010; 71: 455-460
  • 8 Repici A, Vleggaar FP, Hassan C et al. Efficacy and safety of biodegradable stents for refractory benign esophageal strictures: the BEST (Biodegradable Esophageal Stent) study. Gastrointest Endosc 2010; 72: 927-934
  • 9 Hirdes MM, Siersema PD, Vleggaar FP. A new fully covered metal stent for the treatment of benign and malignant dysphagia: a prospective follow-up study. Gastrointest Endosc 2012; 75: 712-718
  • 10 Uitdehaag MJ, van Hooft JE, Verschuur EM et al. A fully-covered stent (Alimaxx-E) for the palliation of malignant dysphagia: a prospective follow-up study. Gastrointest Endosc 2009; 70: 1082-1089
  • 11 Uitdehaag MJ, Siersema PD, Spaander MC et al. A new fully covered stent with antimigration properties for the palliation of malignant dysphagia: a prospective cohort study. Gastrointest Endosc 2010; 71: 600-605
  • 12 Chan AC, Shin FG, Lam YH et al. A comparison study on physical properties of self-expandable esophageal metal stents. Gastrointest Endosc 1999; 49: 462-465
  • 13 Moon T, Hong D, Chun HJ et al. New approach to radial expansive force measurement of self expandable esophageal metal stents. ASAIO J 2001; 47: 646-650
  • 14 Dyet JF, Watts WG, Ettles DF et al. Mechanical properties of metallic stents: how do these properties influence the choice of stent for specific lesions?. Cardiovasc Intervent Radiol 2000; 23: 47-54
  • 15 Isayama H, Nakai Y, Toyokawa Y et al. Measurement of radial and axial forces of biliary self-expandable metallic stents. Gastrointest Endosc 2009; 70: 37-44
  • 16 Gyongyosi M, Yang P, Khorsand A et al. Longitudinal straightening effect of stents is an additional predictor for major adverse cardiac events. Austrian Wiktor Stent Study Group and European Paragon Stent Investigators. J Am Coll Cardiol 2000; 35: 1580-1589
  • 17 De Beule M, Van Cauter S, Mortier P et al. Virtual optimization of self-expandable braided wire stents. Med Eng Phys 2009; 31: 448-453
  • 18 Capelli C, Bosi GM, Cerri E et al. Patient-specific simulations of transcatheter aortic valve stent implantation. Med Biol Eng Comput 2012; 50: 183-192
  • 19 Gundert TJ, Shadden SC, Williams AR et al. A rapid and computationally inexpensive method to virtually implant current and next-generation stents into subject-specific computational fluid dynamics models. Ann Biomed Eng 2011; 39: 1423-1437