RSS-Feed abonnieren
DOI: 10.1055/s-0034-1385333
Comparison of Methods for Acceptance and Constancy Testing in Dental Cone-beam Computed Tomography
Methodenvergleich zur Abnahme- und Konstanzprüfung in der dentalen digitalen VolumentomografiePublikationsverlauf
30. Juni 2014
08. September 2014
Publikationsdatum:
12. November 2014 (online)
Abstract
Purpose: The aim of this work was to implement, validate, and compare two procedures for routine image quality (IQ) assurance in dental cone-beam computed tomography (CBCT): 1. the German standard DIN 6868 – 161 introduced in 2013 and 2. the established standard IEC 61 223 – 3-5 for clinical CT x-ray equipment referenced as “DIN” and “IEC” below.
Materials and Methods: The approximated in-plane modulation transfer function (MTF), the contrast-to-noise indicator (CNI), and the uniformity indicator (UI*) were determined in accordance with DIN. Image noise, the uniformity index (UI), the contrast-to-noise ratio (CNR), and the 3 D MTF were measured according to IEC 61 223 – 3-5 using a previously proposed quality assurance (QA) framework. For this, a modular phantom was used. All experiments were performed on a clinical dental CBCT unit. The severity of image artefacts was measured at different z-positions. A dedicated computer program was implemented to allow for automated QA procedure.
Results: The position and orientation of the phantoms were detected automatically in all of the measurements providing a reproducible placement of the evaluation regions and volumes. 50 % and 10 % in-plane MTF values of the approximated and the exact MTF calculation procedure were in agreement to within 5 %. With increasing axial distance from the isocentre, UI* and CNI dropped by 30 % and 19 %, respectively. Conventional IQ parameters showed higher sensitivity to image artefacts; i. e., UI and CNR were reduced by about 197 % and 37 %.
Conclusion: The implemented automated QA routines are compatible with both the DIN and the IEC approach and offer reliable and quantitative tracking of imaging performance in dental CBCT for clinical practice. However, there is no equivalence between the DIN and the IEC metrics. In addition, direct measurements of physical IQ parameters such as image contrast and noise, uniformity, and axial resolution are not supported by the new concept according to DIN.
Key points:
• The new DIN 6868 – 161 is not equivalent to the established IEC 61 223 – 3-5.
• Noise, uniformity, and contrast are well-suited to assess image artefacts.
• The implemented automated quality assurance program fits clinical routine.
Citation Format:
• Steiding C, Kolditz D, Kalender W. Comparison of Methods for Acceptance and Constancy Testing in Dental Cone-beam Computed Tomography. Fortschr Röntgenstr 2015; 187: 283 – 290
Zusammenfassung
Ziel: Das Ziel dieser Arbeit war die Anwendung, Validierung und der Vergleich zweier Konzepte zur Beurteilung der Bildqualität (BQ) in der dentalen digitalen Volumentomografie (DVT): 1. der 2013 neu eingeführte, deutsche Standard DIN 6868 – 161 und 2. der fest etablierte Standard IEC 61 223 – 3-5 von Röntgeneinrichtungen für klinische CT (nachfolgend als „DIN“ und „IEC“ bezeichnet).
Material und Methoden: Die approximierte, transversale Modulationsübertragungsfunktion (MÜF), der Homogenitätsindikator (H*) und Kontrast-Rausch-Indikator (KRI) wurden nach DIN bestimmt. Unter Hinzunahme eines kürzlich vorgestellten, IEC-entsprechenden Qualitätssicherungskonzeptes wurden Bildrauschen, der Homogenitätsindex (H), das Kontrast-Rausch-Verhältnis (KRV) und die 3 D-MÜF mit einem modularen Prüfkörper gemessen. Alle Messungen wurden an einem klinischen DVT-Gerät durchgeführt. Beide Phantome wurden an variierenden z-Positionen platziert, um das Ausmaß der Bildartefakte zu untersuchen. Ein spezielles Computerprogramm wurde für die automatisierte Bildqualitätssicherung implementiert.
Ergebnisse: Die Detektion beider Prüfkörper erfolgte in den Messungen automatisiert und gewährleistete das reproduzierbare Platzieren der Auswerteregionen und -volumen. Die 50 %- und 10 %-MÜF-Werte nach approximierter und exakter Berechnungsmethode stimmten bis auf 5 % überein. Mit zunehmendem axialen Abstand vom Messfeldzentrum fielen H* und KRI um 30 % beziehungsweise 19 %. Die konventionellen BQ-Parameter zeigten eine höhere Sensitivität gegenüber den Bildartefakten; H und KRV reduzierten sich um 197 % und 37 %.
Schlussfolgerung: Die automatisierten Qualitätssicherungsprozeduren offerieren eine zuverlässige und quantitative Beurteilung der Leistungsmerkmale zur Bildgebung in der DVT für die klinische Praxis. Es gibt keine Äquivalenz zwischen den BQ-Maßen nach DIN und IEC. Zudem wird die direkte Charakterisierung physikalischer Bildgebungseigenschaften in Form von Bildkontrast und -rauschen, Homogenität und axiale Ortsauflösung durch die neue DIN-Norm nicht unterstützt.
Kernaussagen:
• Gleichwertigkeit zwischen der neuen DIN 6868 – 161 und IEC 61 223 – 3-5 ist nicht gegeben.
• Rauschen, Homogenität und Kontrast sind gut zur Untersuchung von Bildartefakten geeignet.
• Das implementierte, automatisierte Qualitätssicherungsprogramm kann in der klinischen Routine eingesetzt werden.
-
References
- 1 Mozzo P, Procacci C, Tacconi A et al. A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results. Eur Radiol 1998; 8: 1558-1564
- 2 Arai Y, Tammisalo E, Iwai K et al. Development of a compact computed tomographic apparatus for dental use. Dentomaxillofac Radiol 1999; 28: 245-248
- 3 Vannier MW. Craniofacial computed tomography scanning: technology, applications and future trends. Orthod Craniofac Res 2003; 6: 23-30
- 4 Holberg C, Steinhäuser S, Geis P et al. Cone-beam computed tomography in orthodontics: benefits and limitations. J Orofac Orthop 2005; 66: 434-444
- 5 Hatcher DC. Operational principles for cone-beam computed tomography. J Am Dent Assoc 2010; 141: 3S-6S
- 6 Kalender WA. Computed Tomography: Fundamentals, System Technology, Image Quality, Applications. 3rd ed. Erlangen: Publicis Publishing; 2011
- 7 Suomalainen A, Kiljunen T, Käser Y et al. Dosimetry and image quality of four dental cone beam computed tomography scanners compared with multislice computed tomography scanners. Dentomaxillofac Radiol 2009; 38: 367-378
- 8 Ludlow JB, Ivanovic M. Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008; 106: 106-114
- 9 Eggers G, Klein J, Welzel T et al. Geometric accuracy of digital volume tomography and conventional computed tomography. Br J Oral Maxillofac Surg 2008; 46: 639-644
- 10 Dalchow CV, Weber AL, Yanagihara N et al. Digital volume tomography: radiologic examinations of the temporal bone. Am J Roentgenol 2006; 186: 416-423
- 11 Hatcher DC, Aboudara CL. Diagnosis goes digital. Am J Orthod Dentofacial Orthop 2004; 125: 512-515
- 12 Heiland M, Schulze D, Rother U et al. Postoperative imaging of zygomaticomaxillary complex fractures using digital volume tomography. J Oral Maxillofac Surg 2004; 62: 1387-1391
- 13 Schulze R, Hassfeld S, Schulze D et al. S1-Empfehlung: Dentale Volumentomographie (DVT). Dtsch Zahnärztl Z 2009; 64: 490-496
- 14 Health Protection Agency. Recommendations for the design of X-ray facilities and quality assurance of dental Cone Beam CT (Computed Tomography) systems HPA-RPD-065. Chilton: Health Protection Agency; 2010
- 15 Carter L, Farman AG, Geist J et al. American Academy of Oral and Maxillofacial Radiology executive opinion statement on performing and interpreting diagnostic cone beam computed tomography. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2008; 106: 561-562
- 16 Horner K, Islam M, Flygare L et al. Basic principles for use of dental cone beam computed tomography: consensus guidelines of the European Academy of Dental and Maxillofacial Radiology. Dentomaxillofac Radiol 2009; 38: 187-195
- 17 Vassileva J, Stoyanov D. Quality control and patient dosimetry in dental cone beam CT. Radiat Prot Dosimetry 2010; 139: 310-312
- 18 Pauwels R, Stamatakis H, Manousaridis G et al. Development and applicability of a quality control phantom for dental cone-beam CT. J Appl Clin Med Phys 2011; 12: 245-260
- 19 Steiding C, Kolditz D, Kalender WA. A quality assurance framework for the fully automated and objective evaluation of image quality in cone-beam computed tomography. Med Phys 2014; 41: 031901
- 20 SEDENTEXCT Guideline Development Panel. Radiation Protection: Cone Beam CT for Dental and Maxillofacial Radiology. Evidence based guidelines. A report prepared by the SEDENTEXCT project. 2011 Available on: www.sedentexct.eu
- 21 Deutsches Institut für Normung (DIN). “Sicherung der Bildqualität in röntgendiagnostischen Betrieben – Teil 161: Abnahmeprüfung nach RöV an zahnmedizinischen Röntgeneinrichtungen zur digitalen Volumentomographie,” DIN Report Nr. 6868-161 (DIN, Berlin, 2013).
- 22 International Electrotechnical Commission (IEC). “Evaluation and routine testing in medical imaging departments – Part 3–5: Acceptance tests – Imaging performance of computed tomography X-ray equipment,” IEC Report No. 61223-3-5. Geneva: IEC; 2004
- 23 Kyriakou Y, Kolditz D, Langner O et al. Digital volume tomography (DVT) and multislice spiral CT (MSCT): an objective examination of dose and image quality. Fortschr Röntgenstr 2011; 183: 144-153
- 24 Blendl C, Fiebich M, Voigt JM et al. Investigation on the 3 D geometric accuracy and on the image quality (MTF, SNR and NPS) of volume tomography units (CT, CBCT and DVT). Fortschr Röntgenstr 2012; 184: 24-31
- 25 Bundesministerium für Umwelt, Naturschutz, Bau und Reaktorsicherheit (BMU). “Richtlinie zur Durchführung der Qualitätssicherung bei Röntgeneinrichtungen zur Untersuchung oder Behandlung von Menschen nach den §§ 16 und 17 der Röntgenverordnung – Qualitätssicherungs-Richtlinie (QS-RL),”. Berlin: BMU; 2014