RSS-Feed abonnieren
DOI: 10.1055/s-0034-1391886
Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa
Publikationsverlauf
submitted17. Dezember 2014
accepted after revision02. März 2015
Publikationsdatum:
31. März 2015 (online)
Background and study aims: Infections are a recognized risk of endoscopic retrograde cholangiopancreatography (ERCP). This paper reports on a large outbreak of VIM-2-producing Pseudomonas aeruginosa that was linked to the use of a recently introduced duodenoscope with a specific modified design (Olympus TJF-Q180V).
Methods: Epidemiological investigations and molecular typing were executed in order to identify the source of the outbreak. Audits on implementation of infection control measures were performed. Additional infection control strategies were implemented to prevent further transmission. The design and the ability to clean and disinfect the duodenoscope were evaluated, and the distal tip was dismantled.
Results: From January to April 2012, 30 patients with a VIM-2-positive P. aeruginosa were identified, of whom 22 had undergone an ERCP using a specific duodenoscope, the TJF-Q180V. This was a significant increase compared with the hospital-wide baseline level of 2 – 3 cases per month. Clonal relatedness of the VIM-2 P. aeruginosa was confirmed for all 22 cases and for the VIM-2 strain isolated from the recess under the forceps elevator of the duodenoscope. An investigational study of the new modified design, including the dismantling of the duodenoscope tip, revealed that the fixed distal cap hampered cleaning and disinfection, and that the O-ring might not seal the forceps elevator axis sufficiently. The high monthly number of cases decreased below the pre-existing baseline level following withdrawal of the TJF-Q180V device from clinical use.
Conclusions: Duodenoscope design modifications may compromise microbiological safety as illustrated by this outbreak. Extensive pre-marketing validation of the reprocessability of any new endoscope design and stringent post-marketing surveillance are therefore mandatory.
-
References
- 1 Rutala WA, Weber DJ. New developments in reprocessing semicritical items. Am J Infection Control 2013; 41: 60-66
- 2 Kovaleva J, Peters FT, van der Mei HC et al. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Reviews 2013; 26: 231-254
- 3 Rutala WA, Weber DJ. Healthcare Infection Control Practices Advisory Committee. Guideline for Disinfection and Sterilization in Healthcare Facilities. 2008 Centers for Disease Control and Prevention. Atlanta, GA. Available from: http://www.cdc.gov/hicpac/Disinfection_Sterilization/acknowledg.html
- 4 Baron TH. Endoscopic management of biliary disorders: diagnostic and therapeutic. Surg Clin North Am 2014; 94: 395-411
- 5 Aumeran C, Poincloux L, Souweine B et al. Multidrug-resistant Klebsiella pneumoniae outbreak after endoscopic retrograde cholangiopancreatography. Endoscopy 2010; 42: 895-899
- 6 Holodniy M, Oda G, Schirmer PL et al. Results from a large-scale epidemiologic look-back investigation of improperly reprocessed endoscopy equipment. Infect Control Hosp Epidemiol 2012; 33: 649-656
- 7 Muscarella LF. Investigation and prevention of infectious outbreaks during endoscopic retrograde cholangiopancreatography. Endoscopy 2010; 42: 957-959
- 8 Schwaber MJ, Carmeli Y. Carbapenem-resistant Enterobacteriaceae: a potential threat. JAMA 2008; 300: 2911-2913
- 9 Andremont A, Lucet JC. Extensively resistant VIM-2-positive Pseudomonas aeruginosa . Lancet Infect Dis 2013; 13: 828-829
- 10 Livermore DM. Of Pseudomonas, porins, pumps and carbapenems. J Antimicrob Chemother 2001; 47: 247-250
- 11 Walsh TR. Clinically significant carbapenemases: an update. Curr Opin Infect Dis 2008; 21: 367-371
- 12 Van der Bij AK, Van der Zwan D, Peirano G et al. Metallo-beta-lactamase-producing Pseudomonas aeruginosa in The Netherlands: the nationwide emergence of a single sequence type. Clin Microbiol Infect 2012; 18: E369-372
- 13 Advisory Board Cleaning and Disinfection Flexible Endoscopes (SFERD). Professional standard handbook. Flexible endoscopes – Cleaning and disinfection. 2011 Available from: http://www.google.be/url?sa=t&rct=j&q=&esrc=s&frm=1&source=web&cd=2&cad=rja&uact=8&ved=0CCwQFjAB&url=http%3A%2F%2Fwww.sterilisatievereniging.nl%2Fpublicaties%2Fdoc_download%2F207-sferd-professional-standard-handbook-flexibele-endoscopes-cleaning-and-disinfection-2011-&ei=59z5VI-nJ4njUY-ng_AE&usg=AFQjCNGQIcu220u2WlZkQMgmSzD-bPsQjw
- 14 Pitout JD, Chow BL, Gregson DB et al. Molecular epidemiology of metallo-beta-lactamase-producing Pseudomonas aeruginosa in the Calgary Health Region: emergence of VIM-2-producing isolates. J Clin Microbiol 2007; 45: 294-298
- 15 Loeve AJ. Investigational report on a TJF-Q180V duodenoscope following contamination after cleaning and disinfection. 2012 Available on request from the Dutch Healthcare Inspectorate/National Institute for Public Health and the Environment (Bilthoven, The Netherlands)
- 16 Gastmeier P, Vonberg RP. Klebsiella spp. in endoscopy-associated infections: we may only be seeing the tip of the iceberg. Infection 2014; 42: 15-21
- 17 Centers for Disease Control and Prevention. Notes from the Field: New Delhi metallo-β-lactamase-producing Escherichia coli associated with endoscopic retrograde cholangiopancreatography – Illinois, 2013. Morb mortal Wkly Rep 2014; 62: 1051
- 18 Otter JA, Vickery K, Walker JT et al. Surface-attached cells, biofilms and biocide susceptibility: implications for hospital cleaning and disinfection. J Hosp Infect 2015; 89: 16-27
- 19 Bajolet O, Ciocan D, Vallet C et al. Gastroscopy-associated transmission of extended-spectrum beta-lactamase-producing Pseudomonas aeruginosa . J Hosp Infect 2013; 83: 341-343
- 20 Fraser TG, Reiner S, Malczynski M et al. Multidrug-resistant Pseudomonas aeruginosa cholangitis after endoscopic retrograde cholangiopancreatography: failure of routine endoscope cultures to prevent an outbreak. Infect Control Hosp Epidemiol 2004; 25: 856-859
- 21 Silva CV, Magalhaes VD, Pereira CR et al. Pseudo-outbreak of Pseudomonas aeruginosa and Serratia marcescens related to bronchoscopes. Infect Control Hosp Epidemiol 2003; 24: 195-197
- 22 Kirschke DL, Jones TF, Craig AS et al. Pseudomonas aeruginosa and Serratia marcescens contamination associated with a manufacturing defect in bronchoscopes. New Engl J Med 2003; 348: 214-220
- 23 Srinivasan A, Wolfenden LL, Song X et al. An outbreak of Pseudomonas aeruginosa infections associated with flexible bronchoscopes. New Engl J Med 2003; 348: 221-227
- 24 Morris J, Duckworth GJ, Ridgway GL. Gastrointestinal endoscopy decontamination failure and the risk of transmission of blood-borne viruses: a review. J Hospital Infection 2006; 63: 1-13
- 25 Buss AJ, Been MH, Borgers RP et al. Endoscope disinfection and its pitfalls – requirement for retrograde surveillance cultures. Endoscopy 2008; 40: 327-332
- 26 Olympus GmbH. Important safety advice – Safe reprocessing of TJF-Q180V. January 2013. Available from: https://www.swissmedic.ch/recalllists_dl/07207/Vk_20130123_15-e1.pdf
- 27 Olympus GmbH. Field safety corrective action – EVIS EXERA II DUODENOVIDEOSCOPE TJF-Q180V. August 2014. Available from: https://www.swissmedic.ch/recalllists_dl/10220/Vk_20140729_02-e1.pdf