Thromb Haemost 2000; 84(03): 401-409
DOI: 10.1055/s-0037-1614035
Commentary
Schattauer GmbH

von Willebrand Disease in a Pediatric-based Population - Comparison of Type 1 Diagnostic Criteria and Use of the PFA-100® and a von Willebrand Factor/Collagen-binding Assay

J.A. Dean
1   From the Division of Haematology/Oncology, Department of Paediatrics, University of Toronto, Toronto
,
V. S. Blanchette
1   From the Division of Haematology/Oncology, Department of Paediatrics, University of Toronto, Toronto
4   Department of Paediatrics, University of Toronto, Toronto
,
M. D. Carcao
1   From the Division of Haematology/Oncology, Department of Paediatrics, University of Toronto, Toronto
4   Department of Paediatrics, University of Toronto, Toronto
,
A. M. Stain
2   Department of Nursing, University of Toronto, Toronto
,
C. R. Sparling
3   Department of Paediatric Laboratory Medicine, The Hospital for Sick Children, University of Toronto, Toronto
,
J. Siekmann
7   Baxter Hyland Immuno, Vienna, Austria
,
P. L. Turecek
7   Baxter Hyland Immuno, Vienna, Austria
,
D. Lillicrap
6   Departments of Pathology and Medicine, Queen’s University, Kingston, Ontario, Canada
,
M. L. Rand
1   From the Division of Haematology/Oncology, Department of Paediatrics, University of Toronto, Toronto
5   Department of Biochemistry, University of Toronto, Toronto
› Author Affiliations
We thank Dr. Paul Foster for his insight and helpful comments regarding this work and Dr. Marian Packham for reading of the manuscript. We also acknowledge the technical assistance of Ms. Ingrid Neunteufl with the multimer analysis and the assistance of Mr. Lijing He with the correlation analyses. We thank Dade Behring Inc for providing the PFA-100 instrument and test cartridges, and International Technidyne Corp. for providing the Surgicutt and Surgicutt-Jr. bleeding time devices. D.L. is supported by a Career Investigator Award from the Heart and Stroke Foundation of Ontario.
Further Information

Publication History

Received 06 July 1999

Accepted after resubmission 15 March 2000

Publication Date:
14 December 2017 (online)

Summary

Definitive diagnosis of type 1 von Willebrand Disease (VWD) remains a problem. Provisional consensus guidelines for the diagnosis of definite and possible type 1 VWD were prepared by the Scientific Subcommittee on von Willebrand factor (VWF) of the Scientific and Standardization Committee (SSC) of the International Society on Thrombosis and Haemostasis (ISTH) during the 1996 annual meeting for the specific purpose of further evaluation in retrospective and prospective studies by a Working Party on Diagnostic Criteria (1996 Annual Report of the SSC/ISTH Subcommittee on VWF). In the first phase of this study, we compared 2 definitions of type 1 VWD, each with 3 criteria: significant bleeding history, laboratory investigations, and family history. Using the ISTH consensus guidelines for type 1 VWD definition, significantly fewer patients were diagnosed with definite type 1 disease as compared to our “in house” Hospital for Sick Children (HSC) criteria (4 vs. 31). While we recognize that the provisional ISTH consensus guidelines were not intended for clinical use, we believe that the results of our studies are of interest and will assist in any future refinements to the ISTH guidelines.

In the second phase of this study, we investigated the utility of 2 new tests, a laboratory screening test and a functional test, for VWD in our well characterized, pediatric-based population. The Platelet Function Analyzer (PFA-100®) provides an in vitro measure of primary hemostasis under conditions of high shear, using disposable cartridges containing collagen and either epinephrine or ADP. All tested subjects with types 2 or 3 VWD had prolonged PFA-100 closure times (CTs) with both cartridge types (n = 17) and prolonged bleeding times (n = 14). In subjects with definite type 1 VWD, 20/24 (83%) had prolonged CTs with the collagen/ADP cartridge (19/24 (79%) with collagen/epinephrine), compared with 7/26 (27%) with prolonged bleeding times. In subjects with definite types 1, 2, or 3 VWD, collagen/ADP CTs were abnormal in 37/41 subjects, giving an overall sensitivity of 90%. With this high sensitivity, the PFA-100 is a better screening test for VWD than the bleeding time.

We also tested a VWF collagen-binding assay (VWF:CBA) as a functional test for VWF, in comparison with the more routinely-used ristocetin cofactor assay (VWF:RCo). The VWF:CBA is based on an ELISA technique, which has the potential to be more reproducible than the VWF:RCo. We found that the VWF:CBA detected 43/49 (88%) subjects with definite types 1, 2, or 3 VWD, performing as well as the VWF:RCo, that detected 42/48 (88%). We also showed that, used in conjunction with VWF antigen levels, the VWF:CBA may be useful in classification of VWD subtypes.

Presented in part at the 39th Annual Meeting of the American Society of Hematology, San Diego, CA, USA, December 1997.


 
  • References

  • 1 von Willebrand EA. Hereditär pseudohemofili. Finska Läkarsällskapets Handl 1926; 67: 7-112.
  • 2 Meyer D, Girma JP. von Willebrand factor: Structure and function. Thromb Haemost 1993; 70: 99-104.
  • 3 Ruggeri ZM. von Willebrand factor. J Clin Invest 1997; 99: 559-64.
  • 4 Rodeghiero F, Castaman G, Dini E. Epidemiological investigation of the prevalence of von Willebrand’s disease. Blood 1987; 69: 454-9.
  • 5 Werner EJ, Broxson EH, Tucker EL, Giroux DS, Shults J, Abshire TC. Prevalence of von Willebrand disease in children: A multiethnic study. J Pediatr 1993; 123: 893-8.
  • 6 Gill JC, Endres-Brooks J, Bauer PJ, Marks Jr WJ, Montgomery RR. The effect of ABO blood group on the diagnosis of von Willebrand disease. Blood 1987; 69: 1691-5.
  • 7 Pottinger BE, Read RC, Paleolog EM, Higgins PG, Pearson JD. von Willebrand factor is an acute phase reactant in man. Thromb Res 1989; 53: 387-94.
  • 8 Abildgaard CF, Suzuki Z, Harrison J, Jefcoat K, Zimmerman TS. Serial studies in von Willebrand’s disease: Variability versus “variants”. Blood 1980; 56: 712-6.
  • 9 Miller CH, Graham JB, Goldin LR, Elston RC. Genetics of classic von Willebrand’s disease. II. Optimal assignment of the heterozygous genotype (diagnosis) by discriminant analysis. Blood 1979; 54: 137-45.
  • 10 Rick ME. Laboratory diagnosis of von Willebrand’s disease. Clin Lab Med 1994; 14: 781-94.
  • 11 Miller CH, Graham JB, Goldin LR, Elston RC. Genetics of classic von Willebrand’s disease. I. Phenotypic variation within families. Blood 1979; 54: 117-36.
  • 12 Sadler JE. A revised classification of von Willebrand disease. For the Subcommittee on von Willebrand factor of the Scientific and Standardization Committee of the International Society on Thrombosis and Haemostasis. Thromb Haemost 1994; 71: 520-5.
  • 13 Favaloro EJ, Koutts J. Laboratory assays for von Willebrand factor: Relative contribution to the diagnosis of von Willebrand’s disease. Pathology 1997; 29: 385-91.
  • 14 Montgomery RR, Coller BS. Von Willebrand disease. In: Hemostasis and Thrombosis. Basic Principles and Clinical Practice. Colman RW, Hirsh J, Marder VJ, Salzman EW. Philadelphia: J. B. Lippincott; 1994: 143-4.
  • 15 Fressinaud E, Veyradier A, Truchaud F, Martin I, Boyer-Neumann C, Trossaert M, Meyer D. Screening for von Willebrand disease with a new analyzer using high shear stress: A study of 60 cases. Blood 1998; 91: 1325-31.
  • 16 Cattaneo M, Federici AB, Lecchi A, Agati B, Lombardi R, Stabile F, Bucciarelli P. Evaluation of the PFA-100® system in the diagnosis and therapeutic monitoring of patients with von Willebrand disease. Thromb Haemost 1999; 82: 35-9.
  • 17 Brown JE, Bosak JO. An ELISA test for the binding of von Willebrand antigen to collagen. Thromb Res 1986; 43: 303-11.
  • 18 Favaloro EJ, Grispo L, Dinale A, Berndt M, Koutts J. von Willebrand’s disease: Laboratory investigation using an improved functional assay for von Willebrand factor. Pathology 1993; 25: 152-8.
  • 19 Favaloro EJ, Facey D, Grispo L. Laboratory assessment of von Willebrand factor. Am J Clin Pathol 1995; 104: 264-71.
  • 20 Landis JR, Koch GG. The measurement of observer agreement for categorical data. Biometrics 1977; 33: 159-74.
  • 21 Ruggeri Z, Zimmerman TS. The complex multimeric composition of factor VIII/von Willebrand factor. Blood 1981; 57: 1140-3.
  • 22 Aihara M, Sawada Y, Ueno K, Morimoto S, Yoshida Y, de Serres M, Cooper HA, Wagner RH. Visualisation of von Willebrand factor multimers by immunoenzymatic stain using avidin-biotin peroxidase complex. Thromb Haemost 1986; 55: 263-7.
  • 23 Carcao MD, Blanchette VS, Dean JA, He L, Kern MA, Stain AM, Sparling CR, Stephens D, Ryan G, Freedman J, Rand ML. The Platelet Function Analyzer (PFA-100®): A novel in-vitro system for evaluation of primary haemostasis in children. Br J Haematol 1998; 73: 70-3.
  • 24 Siekmann J, Turecek PL, Schwarz HP. The determination of von Willebrand factor activity by collagen binding assay. Haemophilia 1998; 04: 15-24.
  • 25 Sadler JE, Matsushita T, Dong Z, Tuley EA, Westfield LA. Molecular mechanism and classification of von Willebrand disease. Thromb Haemost 1995; 74: 161-6.
  • 26 Silwer J. von Willebrand’s disease in Sweden. Acta Paediatr Scand 1973; Suppl 238: 1-159.
  • 27 Italian Working Group. Spectrum of von Willebrand’s disease: A study of 100 cases. Br J Haematol 1977; 35: 101-12.
  • 28 Gralnick HR, Rick ME, McKeown LP, Williams SB, Parker RI, Maisonneuve P, Jenneau C, Sultan Y. Platelet von Willebrand factor: An important determinant of the bleeding time in type 1 von Willebrand’s disease. Blood 1986; 68: 58-61.
  • 29 Kundu S, Sio R, Mitu A, Ostgaard R. Evaluation of platelet function by PFA-100™. Clin Chem 1994; 40: 1827-8.
  • 30 Kundu SK, Heilmann EJ, Sio R, Garcia C, Davidson RM, Ostgaard RA. Description of an in vitro platelet function analyzer - PFA-100™. Sem Thromb Hemost 1995; 21 (Suppl. 02) 106-12.
  • 31 Kundu SK, Heilmann EJ, Sio R, Garcia C, Ostgaard RA. Characterization of an in vitro platelet function analyzer, PFA-100™. Clin Appl Thrombosis/Hemostasis 1996; 02: 241-9.
  • 32 Mammen EF, Alshameeri RS, Comp PC. Preliminary data from a field trial of the PFA-100™ system. Sem Thromb Hemost 1995; 21 (Suppl. 02) 113-21.
  • 33 Mammen EF, Comp PC, Gosselin R, Greenberg C, Hoots WK, Kessler CM, Larkin EC, Liles D, Nugent DJ. PFA-100™ system: A new method for assessment of platelet dysfunction. Sem Thromb Hemost 1998; 24: 195-202.
  • 34 Marshall PW, Williams AJ, Dixon RM, Growcott JW, Warburton S, Armstrong J, Moores J. A comparison of the effects of aspirin on bleeding time measured using Simplate™ methods and closure time measured using the PFA-100™, in healthy volunteers. Br J Clin Pharmacol 1997; 44: 151-5.
  • 35 Harrison P, Robinson MSC, Mackie IJ, Joseph J, McDonald SJ, Liesner R, Savidge GF, Pasi J, Machin SJ. Performance of the platelet function analyser PFA-100® in testing abnormalities of primary haemostasis. Blood Coag Fibrinol 1999; 10: 25-31.
  • 36 Poujol C, Nurden A, Paponneau A, Heilmann E, Nurden P. Ultrastructural analysis of the distribution of von Willebrand factor and fibrinogen in platelet aggregates formed in the PFA-100™. Platelets 1998; 09: 381-9.
  • 37 Fressinaud E, Veyradier A, Sigaud M, Boyer-Neumann C, Le Boterff C, Meyer D. Therapeutic monitoring of von Willebrand disease: Interest and limits of a platelet function analyser at high shear rates. Br J Haematol 1999; 106: 777-83.
  • 38 Rand ML, Carcao MD, Blanchette VS. Use of the PFA-100® in the assessment of primary, platelet-related hemostasis in a pediatric setting. Sem Thromb Hemost 1998; 24: 523-9.
  • 39 Furlan M. von Willebrand factor: Molecular size and functional activity. Ann Hematol 1996; 72: 341-8.
  • 40 Favaloro EJ. Laboratory assessment for von Willebrand disease: Comparison of functional assays and influence on the diagnosis. Thromb Haemost 1997; Suppl: 632.
  • 41 Caron C, Rugeri L, Reade V, Tournoys A, Goudemand J. Usefulness of a functional test based on the von Willebrand factor-collagen binding to classify von Willebrand patients. Thromb Haemost 1997; Suppl: 631.