         The influence of the radionuclidic impurities in radiopharmaceuticals on the reading
            of sample activities in a dose calibrator has been studied. As a practical example,
            commercially available 123I was used to measure the effect of radioactive impurities on “dose calibrator” readings.
            The impurities present in commercially available 123I were identified and quantified by the use of a Ge(Li) detector, and the response
            of a dose calibrator to each impurity was determined. The contribution of radioactive
            impurities to sample readings was determined for each sample of 123I studied. It was found that impurities may increase the dose calibrator readings
            up to 31 %, even though the impurities were within the warranty given by the manufacturer.
         
         
         Es wurde der Einfluß von Verunreinigungen durch Radionuklide in Radiopharmaceuticis
            auf die Ablesung der Probenaktivität mittels eines Dosiseichgerätes untersucht. Als
            praktisches Beispiel wurde kommerziell erhältliches 123J genommen, um diesen Effekt zu messen. Die in kommerziell erhältlichem 123J vorhandenen Verunreinigungen wurden mittels eines Ge(Li)-Detektors identifiziert
            und quantitativ gemessen und der Ausschlag des Eichgerätes auf jede einzelne Verunreinigung
            bestimmt. Der Beitrag der radioaktiven Verunreinigungen zu den Meßergebnissen an jeder
            123J Probe wurde untersucht. Dabei zeigte sich, daß die Verunreinigungen die Ausschläge
            des Dosiseichgerätes bis zu 3 1 % erhöhen können, obwohl die Verunreinigungen innerhalb
            der Erzeugergarantie lagen.