Nuklearmedizin 1980; 19(02): 85-90
DOI: 10.1055/s-0037-1620935
Originalarbeiten – Original Articles
Schattauer GmbH

The Influence of Smoothing Procedures on the Noise Reduction and Resolution Degradation of Scintigraphic Images

Der Einfluß von Glättungsverfahren auf die Verminderung des Untergrundes und die Auflösungsverschlechterung szintigraphischer Bilder
A. Barbara Gwiazdowska
1   From the Medical Physics Department, Institute of Oncology, Warsaw, Poland
,
Ewa T. Skrzypczak
1   From the Medical Physics Department, Institute of Oncology, Warsaw, Poland
,
J. R. Tolwinski
1   From the Medical Physics Department, Institute of Oncology, Warsaw, Poland
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Received: 03. März 1980

Publikationsdatum:
09. Januar 2018 (online)

Summary

The reduction of the statistical error and the decrease of the spatial resolution due to manyfold smoothing procedures applied to scintigraphic images were analysed. A quantitative relation was obtained between the characteristics of the error reduction factor kf, the resolution decrease factor kR and the linear size of the picture element m. The results showed that for the case of quasi-Gaussian smoothing matrices, e.g. manyfold smoothing, kR depends on the product of kf and m. Several numerical examples are given and discussed.

Zusammenfassung

Es wurden die Verminderung des statistischen Fehlers und die Abnahme des räumlichen Auflösungsvermögens, die durch Vielfach-Glättungsverfahren bei szintigraphischen Bildern bewirkt werden, untersucht. Es ergab sich eine quantitative Beziehung zwischen den Charakteristika des Fehlerverminderungsfaktors kf, des Auflösungsabnahmefaktors kR und der linearen Größe des Bildelementes m. Die Ergebnisse zeigten, daß im Falle einer quasi-Gauss’schen Glättungsmatrix, z. B. einer Vielfach-Glättung, kR vom Produkt auf kf und m abhängt. Es werden mehrere numerische Beispiele angeführt und diskutiert.

 
  • References

  • (1) Beck R. N, Zimmer L. T, Charleston D. B, Harper P. V, Hoffer P. B. Advances in fundamental aspects of imaging systems and techniques. In: Medical Radioisotope Scintigraphy 1972, Vol. I, p. 3-45, IAEA, Vienna 1973
  • (2) Cradduck T. D, MacIntyre W. J. Camera-computer systems for rapid dynamic imaging studies. Sem. nucl. Med 7: 323-336 1977;
  • (3) Gwiazdowska B, Totwinski J. Ocena pogorszenia rozdzielczosci w obrazie scyntygraficznym w wyniku procedury ušredniania danych. Postçpy Fiz. Med 14: 87-94 1979;
  • (4) Hutchinson F, Clair Neill G. D. St, Rimmer A. R. Lineprinter display of digital scintiscans. Amer. J. Roentgenol 113: 755-764 1971;
  • (5) Jordan K, Geisler S. Data display in scintigraphy by means of a high-speed electrostatic digital plotter and special computer averaging techniques. In: Medical Radioisotope Scintigraphy 1972, Vol. I, p. 635-652, IAEA, Vienna 1973
  • (6) MacIntyre W. J, Christie J. H. A comparison of data averaging of radioisotope scan data by photographic and dimensional computer techniques. In: Medical Radioisotope Scintigraphy 1968, Vol. I, p. 771-783, IAEA, Vienna 1969
  • (7) Metz C. E, Beck R. N. Quantitative effects of stationary linear image processing on noise and resolution of structure in radionuclide images. J. nucl. Med 15: 164-170 1974;
  • (8) Potschwadek B, Adomssent S. Einige Bemerkungen zur Programmstufe Mittelwertbildung bei der Computerszintigraphie. Radiobiol. Radiother 13: 329-335 1972;
  • (9) Pizer S. M, Todd-Pokropek A. E. Improvement of scintigrams by computer processing. Sem. nucl. Med 8: 125-146 1978;
  • (10) Sprau A. C, Tauxe W. N, Chaapel D. W. A computerized radioisotope-scan-data filter based on a system response to a point source. Mayo Clin. Proc 41: 585-598 1966;
  • (11) Tauxe W. N. 100-level smoothed scintiscans processed and produced by a digital computer. J. nucl. Med 9: 58-63 1968;
  • (12) Tauxe W. N. Tumour localization by scintiscanning technique: Digital computer processing. In: Radioactive Isotopes in the Localization of Tumours. Heinemann 63-68 1969
  • (13) Tauxe W. N. Digitization and data processing of scintiscan matrices by high-speed computer. In: Medical Radioisotope Scintigraphy 1968, Vol. I, p. 803-813, IAEA, Vienna 1969
  • (14) Weber D. A. Computers in nuclear medicine: Introductory concepts. Sem. nucl. Med 8: 107-112 1978;
  • (15) Zajde C. h. Interest of 256 resolution for gamma camera image processing systems. In: Information Processing in Scintigraphy 1975 (Eds.: C. Raynaud and A. Todd-Pokropek) 479-483 1976