The ability of a tomographic system to detect small lesions and to assess them quantitatively
is affected by several performance parameters, among them spatial resolution in the
transverse slice and in the axial direction. This study was done to determine reconstructed
resolution under clinical conditions for various parameters, i. e. radius of rotation,
radial position within the transverse field of view, and amount of smoothing in the
reconstruction process (various reconstruction filters and a Metz prefilter). The
axial resolution is independent of the reconstruction filter but depends on the prefilter,
the radius of rotation (ROR), and the source position. The transaxial resolution is
dependent on the reconstruction filter, the prefilter, the ROR, and the source position.
The reconstructed resolution and the system resolution were compared and differences
between both are discussed.
Zusammenfassung
Die Fähigkeit eines tomographischen Systems, kleine Läsionen nachzuweisen oder quantitativ
zu bestimmen, hängt von verschiedenen Leistungsparametern ab, unter anderem von der
räumlichen Auflösung innerhalb der transversalen Schicht und in axialer Richtung.
Unter patientenbezogenen Bedingungen wird die rekonstruierte Auflösung bestimmt in
Abhängigkeit vom Rotationsradius, der radialen Position im Gesichtsfeld und vom Grad
der Datenglättung im Rekonstruktionsprozeß (unterschiedliche Rekonstruktionsfilter,
zweidimensionale Metz-Prefilter). Die axiale und transversale Auflösung ist abhängig
von Prefilter, Rotationsradius (ROR) und Quellenposition. Die transversale Auflösung
ist zusätzlich vom Rekonstruktionsfilter abhängig. Der Vergleich der transversalen
Auflösung und der Systemauflösung – für einen dem ROR entsprechenden Meßabstand gemessen
– wird diskutiert.