A Pressure Sensitive Film Study on the Effect of Screw Omission on Bone-plate Interface Mechanics in Cadaveric Bone
Publikationsverlauf
Received for publication 21. Juli 1997
Publikationsdatum:
10. Februar 2018 (online)
Summary
Minimalising surgical exposure and the amount of hardware implanted has become the aim in promoting the biological healing of fractures. Increased spacing of fewer bone screws can increase the mechanical strength of fracture fixation without jeopardizing the stability of the construction. The omission of selective screws has also been shown to increase bone strain and has been suggested as a means of treating bone plate induced osteoporosis.
Using Fuji prescale pressure sensitive film we evaluated the effect of selective screw omission on the platebone interface mechanics. Testing was performed using 10-hole, 4.5 mm DCP applied to equine cadaveric third metacarpal bones. Film was interposed between the bone and the bone plate which were applied to a constant level of screw torque. The resultant digitized bone plate patterns enabled computation of the interface contact area (%) and the average force (Newtons-N) at the interface and a qualitative assessment of the pattern of pressure distribution.
The symmetrical omission of 40% of the total screw number resulted in significant differences between the control (10 screws) and each treatment for both interface contact area (p <0.05) and average force (p <0.02). In one pattern of screw omission there was a marginally non-significant difference between the control and treatment interface contact area (p = 0.0518).
The results suggest that a dramatic reduction in the number of screws applied to a given plate (40%) evokes significant changes in the interface mechanics. Of interest is the effect the pattern of screw omission had on the interface mechanical features; certain screw omission patterns induced a reduction in contact area and force of the same magnitude as the number of screws omitted (40%), whilst in other patterns of screw omission the effect on interface mechanical features was much less than expected (20%).
A number of implications can be postulated from these findings;
1. Subsequent to the omission of screws the level interface friction may remain at a level sufficient as to not alter the ability of the fixation, to maintain fracture reduction and stability.
2. The interface bone will benefit from a reduction in interface contact presumably resulting in less impairment to blood outflow through the cortex.
3. The use of fewer screws in a given construction may augment the inherent bone strain.
The effect on plate-bone interface mechanics of the symmetrical omission of screws was evaluated. This was accomplished using Fuji pressure sensitive film and provided an indication of interface contact area and average force. It was found that the omission of 40% of screws from a given plate significantly altered the interface contact area and the average force, however, the extent of these changes was influenced by the pattern of screw omission. This suggests that the selective omission of screws from a plate-bone construction may provide a means of limiting plate-induced osteopaenia during fracture healing, through a reduction in contact area and improved inherent bone strain.
-
REFERENCES
- 1 Beaupre GS, Carter DR, Orr TE. Stresses in plated long-bones: the role of screw tightness and interface slipping. J Orthop Res 1988; 6 (01) 39-50.
- 2 Brinker WO, Hohn RB, Prier WD. Manual of Internal Fixation in Small Animals. New York: Springer-Verlag; 1984: 107.
- 3 Cheat EJ, Mansmann KA, Di Gioia AM. et al. Role of interfragmentary strain in fracture healing: ovine model of a healing osteotomy. J Orthop Res 1991; 9 (01) 131-42.
- 4 Dennis J, Sanders R. Milne T. Minimal vs. maximal compression plating of the ulna: a biomechanical study of indirect reduction technique. J Orthop Trauma 1993; 7 (02) 152-3.
- 5 Edwards CC, Simmons SC, Browner BD. Severe open tibial fractures: results treating 202 injuries with external fixation. Clin Orth Rel Res 1988; 230: 98-115.
- 6 Field JR. McKee SL. Screw torque and bone plate fixation to equine cadaver longbones. Vet Comp Orthop Traumatol 1995; 9 (01) 1-3.
- 7 Field JR, Hearn TC, Caldwell CB. Bone plate fixation: An evaluation of interface contact area and force of the Dynamic Compression Plate and Limited Contact- Dynamic Compression Plate applied to cadaveric bone. J Orthop Trauma 1997. (in press).
- 8 Field JR, Hearn TC, Caldwell CB. The influence of screw torque, object radius of curvature, mode of bone plate application and bone plate design on bone plate interface mechanics. Injury 1997. In press.
- 9 Field JR, Tornkvist H, Hearn TC, Sumner- Smith G, Woodside TD. The influence of screw omission on construction stiffness and bone surface strain in the application of bone plates to cadaveric bone. J Orthop Res 1997. In press.
- 10 Ganz R. Mast J, Weber BG, Perren SM. Clinical aspects of 'bio-logical' plating. Injury 1991; 22 Suppll 4-8.
- 11 Gautier E, Cordey J, Mathys R. et al. Porosity and remodelling of plated bone after internal fixation: Result of stress shielding or vascular damage. Proceedings of the 4th European Conference on Biomaterials, Biomat, Biomech. Louvain, Belgium 1983: 195-200.
- 12 Gerber C, Mast JW, Ganz R. Biological internal fixation of fractures. Arch Orthop Trauma Surg 1990; 109: 295-303.
- 13 Goodship AE, Lanyon LE, McFie H. Functional adaptation of bone to increased stress. The J Bone and Joint Surg 1979; 61-A (04) 539-46.
- 14 Korvick DL, Monville JD, Pijanowski GJ. et al. The effects of screw removal on bone strain in an idealized plated bone model. Vet Surg 1988; 17 (03) 111-6.
- 15 Liggins AB, Stranart JC, E, Finlay JB. et al. Calibration and manipulation of data from Fuji Pressure-sensitive film. Experimental Mechanics. ed. Little EG. Elsevier Science Publishers; Amsterdam: 1992: 61-70.
- 16 Mast J, Jacob R, Ganz R. Planning and reduction technique in fracture surgery. Berlin: Springer-Verlag; 1989
- 17 Perren SM, Cordey J, Rahn BA. et al. Early temporary porosis of bone induced by internal fixation implants. A reaction to necrosis, not stress protection?. Clin Orthop and Rel Res 1988; 232: 139-51.
- 18 Terjesen T, Apalset K. The influence of different degrees of stiffness of fixation plates on experimental bone healing. J Orthop Res 1988; 6: 293-9.
- 19 Tornkvist H., Hearn TC, Schatzker J. The strength of plate fixation in relation to the number and spacing of bone screws. J Orthop Trauma 1996; 10 (03) 204-8.
- 20 Uhthoff HK, Dubuc FL. Bone structure changes in the dog under rigid internal fixation. Clin Orthop 1971; 81: 165-70.
- 21 Uhthoff HK, Boisvert D, Finnegan M. Cortical porosis under plates. Reaction to unloading or to necrosis. J Bone Joint Surg 1994; 76-A (10) 1507-12.
- 22 Uhthoff HK, Finnegan M. The effects of metal plates on post-traumatic remodelling and bone mass. J Bone Joint Surg 1981; 63-B (03) 427-34.
- 23 Waelchi-Suter C. Vascular changes in cortical bone following intramedullary fixation. In: Current Concepts of Internal Fixation of Fractures. Ed. Uhthoff H. Berlin: Springer Verlag; 1980: 411-5.
- 24 Wiss D, Flemming G, Matta J. et al. Comminuted and rotationally unstable fractures of the femur treated with an interlocking nail. Clin Orthop Rel Res 1986; 212: 35-47.
- 25 Woo SL, Lothringer KS, Akeson WH. et al. Less rigid internal fixation plates: Historical perspectives and new concepts. J Orthop Res 1984; 1: 431-49.