Polyethylene Wear in Retrieved Canine Acetabular Components
A Tribological Study The authors thank John Mitchell for scanning electron microscopy, Qiang Wang for statistical analysis, BioMedtrix Inc. (Allendale, NJ) for the donation of prosthetic components, and Tim Vojt and Nancy Goldshmidt for preparation of the figures.Publication History
Received
19 August 1999
Accepted
24 November 1999
Publication Date:
09 February 2018 (online)
Summary
The aim of this pilot study was to define the mechanism and mode of polyethylene wear in acetabular components retrieved from seven dogs following therapeutic total hip replacement. The articular surface of each acetabular component was examined using contact profilometry and scanning electron microscopy (SEM). Peak-tovalley distance (P-V) and arithmetic average surface roughness (Ra) were calculated for each quadrant of the surface. Deformation of the regular profile of the machining lines was the least severe form of surface alteration. Randomly orientated scratches, fine tapered filaments, ripples, and coarse surface shredding were common SEM observations. Gouging of the convex surface of the cup was seen in two cases following failure of the polyethylenecement interface. Wear of the nonarticular acetabular rim suggested neck impingement in one case. P-V and Ra were significantly lower in the craniodorsal zone, compared to the average roughness of the other three quadrants (p < 0.05). Abrasion, adhesion and fatigue were the principal mechanisms of polyethylene wear, and were implicated in the production of polyethylene particulate debris. Meticulous removal of abrasive third bodies at the time of surgery, and correct orientation of the acetabular component, should reduce early and severe wear.
Polyethylene wear is inevitable following metal-on-polyethylene total hip replacement. In this study, the nature of polyethylene wear in retrieved canine acetabular components was defined, using contact profilometry and scanning electron microscopy. Abrasion, adhesion and fatigue were the principal mechanisms of wear, and were implicated in the production of polyethylene particulate debris.
-
REFERENCES
- 1 Amstutz HC, Grigoris P. Metal on metal bearings in hip arthroplasty. Clin Orthop 1996; 329S S11-S34.
- 2 Bankston AB, Cates H, Ritter MA. et al. Polyethylene wear in total hip arthroplasty. Clin Orthop 1995; 317: 7-13.
- 3 Bankston AB, Keating EM, Ranawat C. et al. Comparison of polyethylene wear in machined versus molded polyethylene. Clin Orthop 1995; 317: 37-43.
- 4 Bhushan B. Wear. In: Principles and applications of tribology. New York: John Wiley and Sons, Inc; 1999: 479-585.
- 5 Bhushan B. Solid surface characterization. In: Principles and applications of tribology. New York: John Wiley and Sons, Inc; 1999: 91-197.
- 6 Bloebaum RD, Zou L, Bachus KN. et al. Analysis of particles in acetabular components from patients with osteolysis. Clin Orthop 1997; 338: 109-18.
- 7 Clarke IC, Black K, Rennie C, Amstutz HC. Can wear in total hip arthroplasties be assessed from radiographs?. Clin Orthop 1976; 121: 126-41.
- 8 Cook SD, Barrack RL, Baffes GC. et al. Wear and corrosion of modular interfaces in total hip replacements. Clin Orthop 1994; 298: 80-8.
- 9 Cuckler JM, Bearcroft J, Asgian CM. Femoral head technologies to reduce polyethylene wear in total hip arthroplasty. Clin Orthop 1995; 317: 57-63.
- 10 Devane PA, Robinson EJ, Bourne RB. et al. Measurement of polyethylene wear in acetabular components inserted with and without cement. A randomized trial. J Bone Joint Surg (Am) 1997; 79: 682-9.
- 11 Dyce J, Wisner ER, Wang Q, Olmstead ML. Luxation following total hip replacement in the dog. Vet Surg: submitted for publication..
- 12 Goodman SB, Lind M, Song Y, Smith RL. In vitro, in vivo, and tissue retrieval studies on particulate debris. Clin Orthop 1998; 352: 25-34.
- 13 Hailey JL, Ingham E, Stone M. et al. Quantitative comparison of acetabular cup wear rates and femoral head damage in explanted Charnley hip prostheses. Trans ORS 1997; 22: 853.
- 14 Hall RM, Siney P, Wroblewski BM, Unsworth A. Observations on the direction of wear in Charnley sockets retrieved at revision. J Bone Joint Surg (Br) 1998; 80: 1067-72.
- 15 Harris WH. The problem is osteolysis. Clin Orthop 1995; 311: 46-53.
- 16 Hernandez JR, Keating EM. Fari et al. Polyethylene wear in uncemented acetabular components. J Bone Joint Surg (Br) 1993; 76: 263-6.
- 17 Howie DW, Haynes DR, Rogers SD. et al. The particulate response to debris. Orthop Clin NA 1993; 24: 571-81.
- 18 Isaac GH, Wroblewski BM, Atkinson JR, Dowson D. A tribological study of retrieved hip prostheses. Clin Orthop 1992; 276: 115-25.
- 19 Jasty M, Bragdon C, Jiranek W. et al. Etiology of osteolysis around porous-coated cementless total hip arthroplasties. Clin Orthop 1994; 308: 111-26.
- 20 Jasty M, Goetz DD, Bragdon CR. et al. Wear of polyethylene acetabular components in total hip arthroplasty. J Bone Joint Surg (Am) 1997; 79: 349-58.
- 21 Kabo JM, Gebhard JS, Loren G, Amstutz HC. In vivo wear of polyethylene acetabular components. J Bone Joint Surg (Br) 1992; 75: 254-8.
- 22 Kobayashi A, Freeman MAR, Bonfield W. et al. Number of polyethylene particles and osteolysis in total joint replacements. A quantitative study using a tissue-digestion method. J Bone Joint Surg (Br) 1997; 78: 844-8.
- 23 Li S, Burstein AH. Ultra-high molecular weight polyethylene. The material and its use in total joint implants. J Bone Joint Surg (Am) 1994; 76: 1080-9.
- 24 Maloney WJ, Smith RL. Periprosthetic osteolysis in total hip arthroplasty: the role of particulate wear debris. J Bone Joint Surg (Am) 1997; 77: 1448-61.
- 25 Manley MT, Serekian P. Wear debris. An environmental issue in total joint replacement. Clin Orthop 1994; 298: 137-64.
- 26 Martell JM, Berdia S. Determination of polyethylene wear in total hip replacements with use of digital radiographs. J Bone Joint Surg (Am) 1997; 79: 1635-41.
- 27 McKellop HA, Campbell P, Park S-H. et al. The origin of submicron polyethylene wear in total hip arthroplasty. Clin Orthop 1995; 311: 3-20.
- 28 Minakawa H, Stone MH, Wroblewski BM. et al. Quantification of third-body damage and its effect on UHMWPE wear with different types of femoral head. J Bone Joint Surg (Br) 1998; 80: 894-9.
- 29 Olmstead ML. The canine cemented modular total hip prosthesis. J Am Anim Hosp Assoc 1995; 31: 109-24.
- 30 Rose RM, Nusbaum HJ, Schneider H. et al. On the true wear rate of ultra high-molecular-weight polyethylene in the total hip prosthesis. J Bone Joint Surg (Am) 1980; 62: 537-49.
- 31 Rostoker W, Chao EYS, Galante JO. The appearances of wear on polyethylene - a comparison of in vivo and in vitro wear surfaces. J Biomed Mater Res 1978; 12: 317-35.
- 32 Salvati EA, Betts F, Doty SB. Particulate metallic debris in cemented total hip arthroplasty. Clin Orthop 1993; 293: 160-73.
- 33 Schmalzried TP, Dorey FJ, McKellop H. The multifactorial nature of polyethylene wear in vivo. J Bone Joint Surg (Am) 1998; 80: 1234-43.
- 34 Schmalzried TP, Jasty M, Harris WH. Periprosthetic bone loss in total hip arthroplasty. Polyethylene wear debris and the concept of the effective joint space. J Bone Joint Surg (Am) 1992; 74: 849-63.
- 35 Schmalzried TP, Kwong LM, Jasty M. et al. The mechanism of loosening of cemented acetabular components in total hip arthroplasty. Analysis of specimens retrieved at autopsy. Clin Orthop 1992; 274: 60-78.
- 36 Shaver SM, Brown TD, Hillis SL, Callaghan JJ. Digital edge-detection measurement of polyethylene wear after total hip arthroplasty. J Bone Joint Surg (Am) 1997; 79: 690-700.
- 37 Sychterz CJ, Engh CA, Shah N. et al. Radiographic evaluation of penetration by the femoral head into the polyethylene liner over time. J Bone Joint Surg (Am) 1997; 79: 1040-6.
- 38 Sychterz CJ, Engh CA, Swope SW. Analysis of prosthetic femoral heads retrieved at autopsy. Clin Orthop 1999; 358: 223-34.
- 39 Sychterz CJ, Moon KH, Hashimoto Y. et al. Wear of polyethylene cups in total hip arthroplasty. A study of specimens retrieved postmortem. J Bone Joint Surg (Am) 1996; 78: 1193-200.
- 40 Wroblewski BM. Direction and rate of socket wear in Charnley low friction arthroplasty. J Bone Joint Surg (Br) 1985; 67: 757-61.
- 41 Yamaguchi M, Bauer TW, Hashimoto Y. Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups. J Bone Joint Surg (Am) 1997; 79: 1539-44.