Thromb Haemost 1993; 70(02): 286-288
DOI: 10.1055/s-0038-1649567
Original Articles
Coagulation
Schattauer GmbH Stuttgart

Effects of Hirudin on Activated Partial Thromboplastin Time Determined with Ten Different Reagents

Armando Tripodi
The Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Institute of Internal Medicine, IRCCS Maggiore Hospital and University of Milano, Milano, Italy
,
Veena Chantarangkul
The Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Institute of Internal Medicine, IRCCS Maggiore Hospital and University of Milano, Milano, Italy
,
Arnaldo A Arbini
The Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Institute of Internal Medicine, IRCCS Maggiore Hospital and University of Milano, Milano, Italy
,
Marco Moia
The Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Institute of Internal Medicine, IRCCS Maggiore Hospital and University of Milano, Milano, Italy
,
Pier Mannuccio Mannucci
The Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Institute of Internal Medicine, IRCCS Maggiore Hospital and University of Milano, Milano, Italy
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Received 21. Oktober 1992

Accepted after revision 11. Februar 1993

Publikationsdatum:
04. Juli 2018 (online)

Summary

Hirudin prolongs the APTT when added to normal plasma and the extent of prolongation depends on the type of reagent used. The aim of this study was to compare the dose-response curves of 10 widely used APTT reagents for linearity and parallelism. On each of 10 working days a normal pooled plasma was mixed with increasing amounts of recombinant hirudin (HBW023) ranging from 0 to 5 μg/ml and tested for APTT by photo optical coagulometer. Within each working day, clotting times were measured in duplicate and the order of testing with each reagent was changed every day. Results were expressed as ratios of clotting times with hirudin to clotting times without hirudin, and the values plotted against the hirudin concentration on a log-log scale. The dose-response curves for all reagents were linear over 0.3–1.2 μg/ml. The reagent-related slopes ranged from 0.225 ± 0.003 to 0.303 ± 0.003 (mean ± SE) and were significantly different. Precision studies indicated that the least sensitive reagent was also the least precise. These findings indicate that the clotting time values obtained for patients treated with hirudin will vary depending on the APTT reagent used.

 
  • References

  • 1 Markwardt F. Untersuchungen über Hirudin. Naturwissenschaften 1955; 42: 537-538
  • 2 Marki WE, Grossenbacker H, Grutter MG, Liersch MH, Meyhack B, Heim J. Recombinant hirudin: genetic engineering and structure analysis. Semin Thromb Hemostas 1991; 17: 88-93
  • 3 Markwardt F. Hirudin and derivatives as anticoagulant agents. Thromb Haemostas 1991; 66: 141-152
  • 4 Rydel TJ, Ravichandran KG, Tulinski A, Bode W, Huber R, Roitsch C, Fenton II JW. The structure of a complex of recombinant hirudin and human alpha-thrombin. Science 1990; 249: 277-280
  • 5 Brandt JT, Triplett DA. Laboratory monitoring of heparin. Effect of reagents and instruments on the activated partial thromboplastin time. Am J Clin Pathol 1981; 76 (Suppl. 00) 530-537
  • 6 BMPD Statistical software manual. University of California Press; Berkeley, CA: 1990
  • 7 Walenga JM, Hoppensteadt D, Koza M, Wallock M, Pifarre R, Fareed J. Laboratory assays for the evaluation of recombinant hirudin. Haemostasis 1991; 21 (Suppl. 01) 49-63
  • 8 Spannagl M, Bichler H, Lill H, Schramm W. A fast photometric assay for the determination of hirudin. Haemostasis 1991; 21 (Suppl. 01) 36-40
  • 9 Gray E, Watton J, Cesmeli S, Barrowcliffe TW, Thomas DP. Experimental studies on recombinant hirudin, CGP 39393. Thromb Haemostas 1991; 65: 355-359