Thromb Haemost 1983; 50(03): 712-717
DOI: 10.1055/s-0038-1665293
Original Article
Schattauer GmbH Stuttgart

A Collaborative Calibration Study of Reference Materials for Thromboplastins

J Hermans
1   The Department of Medical Statistics, Medical Faculty, University of Leiden, The Netherlands
,
A M H P van den Besselaar
2   The Division of Haemostasis and Thrombosis Research, Haematology Section, Department of Medicine, University Hospital, Leiden, The Netherlands
,
E A Loeliger
2   The Division of Haemostasis and Thrombosis Research, Haematology Section, Department of Medicine, University Hospital, Leiden, The Netherlands
,
E A van der Velde
1   The Department of Medical Statistics, Medical Faculty, University of Leiden, The Netherlands
› Institutsangaben
Weitere Informationen

Publikationsverlauf

Received 18. April 1983

Accepted 29. Juli 1983

Publikationsdatum:
18. Juli 2018 (online)

Summary

In a collaborative study of ten laboratories performed mainly within the framework of the BCR (the European Community Bureau of Reference), five thromboplastins were calibrated against the WHO (World Health Organisation) primary international reference preparation 67/40. Of these five thromboplastins, three were BCR reference materials (BCT/099, OBT/79 and RBT/79) and two were WHO secondary international reference preparations (68/434 and 70/178). Human brain tissue type is represented by 67/40 and BCT/099; bovine type by 68/434 and OBT/79, and rabbit type by 70/178 and RBT/79.

The calibration relations are expressed in terms of a linear relationship between the logarithms of the prothrombin times, measured in seconds.

 
  • References

  • 1 WHO Expert Committee on Biological Standardization. WHO Expert Committee on Biological Standardization 28th Report. WHO Technical Report Series 610. 1977 pp 14-15 and 45-51
  • 2 WHO Expert Committee on Biological Standardization. WHO Expert Committee on Biological Standardization 31st Report. WHO Technical Report Series 658 1981; pp 185-205
  • 3 Biggs R, Denson KW E. Standardization of the one-stage prothrombin time for the control of anticoagulant therapy. Br Med J 1967; 1: 84-88
  • 4 Loeliger EA, Lewis SM. Progress in laboratory control of oral anticoagulants. Lancet 1982; 2: 318-320
  • 5 Kirkwood TB L. Calibration of reference thromboplastins and standardisation of the prothrombin time ratio. Thromb Haemostas 1983; 49: 238-244
  • 6 Loeliger EA, Van den Besselaar AM H P, Hermans J, Van der Velde EA. Certification of three reference materials for thromboplastins. BCR information. Commission of the European Communities; Brussels: 1981
  • 7 Loeliger EA, Van den Besselaar AM H P, Van der Velde EA, Hermans J. Analytical protocol for the characterisation of the BCR Reference Materials for thromboplastins. 1979
  • 8 WHO Expert Committee on Biological Standardization. WHO Expert Committee on Biological Standardization 33rd Report. WHO Technical Report Series. 1983
  • 9 Combleet PJ, Gochman N. Incorrect least-squares regression coefficients in method-comparison analysis. Clin Chem 1979; 25: 432-438
  • 10 Patefield WM. On the information matrix in the linear functional relationship problem. Appl Stat 1977; 26: 69-70
  • 11 Bangham DR, Biggs R, Brozovic M, Denson KW E. Draft report of a collaborative study of two thromboplastins. In: Vascular Factors and Thrombosis. Koller F, Brinkhous KM, Biggs R, Rodman NF, Hinnom S. (eds) Thromb Haemostas 1970. 40 Suppl 341-351
  • 12 Bangham DR, Biggs R, Brozovic M, Denson KW E. Calibration of five different thromboplastins, using fresh and freeze-dried plasma. Thrombos Diathes Haemorrh 1973; 29: 228-239