Mechanical Performance of a Polyaxial Locking Plate and the Influence of Screw Angulation in a Fracture Gap Model

Jakub Kaczmarek
1   AniCura Kleintierspezialisten Augsburg, Augsburg, Germany
,
2   Poznan University of Technology, Poznan, Poland
,
Riccarda Schuenemann
1   AniCura Kleintierspezialisten Augsburg, Augsburg, Germany
,
Piotr Paczos
2   Poznan University of Technology, Poznan, Poland
,
Bartosz Gapinski
2   Poznan University of Technology, Poznan, Poland
,
Sandra Bogisch
1   AniCura Kleintierspezialisten Augsburg, Augsburg, Germany
,
Martin Unger
1   AniCura Kleintierspezialisten Augsburg, Augsburg, Germany
› Institutsangaben

Funding This study was financially supported by the AniCura GmbH—AniCura Research Found. Implants for this study were donated from Aesculap B. Braun Vet Care and DePuy Synthes.
Weitere Informationen

Publikationsverlauf

22. Dezember 2018

03. September 2019

Publikationsdatum:
18. Oktober 2019 (online)

Preview

Abstract

Objective The aim of this study was to compare the locking compression plate (LCP) with polyaxial locking system (PLS) using single cycle to failure 4-point bending test and to investigate the behaviour of PLS with screws inserted mono- and polyaxially using cyclic fatigue test in two bending directions.

Materials and Methods Tests were performed on bone surrogates in a fracture gap model. The 3.5 LCP and 3.5 PLS plates were tested in single cycle to failure. The 3.5 PLS plates with mono- and polyaxial screws were compared in a cyclic fatigue tests in two orthogonal directions. For both experiments, micro-computed tomography (CT) scans were performed pre- and post-testing to investigate the connections between the screw head and the plate hole. Means of forces and cycles needed to failure were statistically compared.

Results The PLS plates were on average 30% weaker than LCP plates. Mode of failure was plate bending in the single cycle to failure tests, and plate breakage in the cyclic fatigue tests. Neither screw breakage nor loss of the screw–plate interface occurred. Mono- and polyaxial constructs performed similarly when loaded in the same direction. Micro-CT revealed no additional internal cracks in the plates or screws after testing. It also showed for both PLS and LCP that there was only partial contact of the screw head with the plate hole.

Clinical Relevance PLS offers a durable locking system, even when the screws are placed polyaxially. The weaker bending properties of the PLS compared with LCP should be considered during preoperative planning.

Note

Results of this study were presented at the 5th WVOC Conference in Barcelona, Spain, September 2018 and was awarded as the Best Free Communication.


Authors' Contributions

Jakub Kaczmarek and Martin Unger contributed to conception of study, study design, acquisition of data and data analysis and interpretation. Riccarda Schuenemann, Sandra Bogisch contributed to conception of study, study design and data analysis and interpretation. Tomasz Bartkowiak, Bartosz Gapinski and Piotr Paczos contributed to acquisition of data and data analysis and interpretation. All authors drafted, revised and approved the submitted manuscript.