Abstract
Objective Umbilical central lines deliver life-saving medications and nutrition for neonates; however, complications associated with umbilical catheters (UCs) occur more frequently than in adults with central lines (i.e., line migration, systemic infection). We have developed a device for neonatal UC protection and stabilization to reduce catheter exposure to bacteria compared with the standard of care: “goal post” tape configuration. This study analyzes the effect of device venting and material on bacterial load of human umbilical cords in vitro .
Study Design Catheters were inserted into human umbilical cord segments in vitro , secured with plastic or silicone vented prototype versus tape, and levels of bacterial colonization were compared between groups after 7 days of incubation.
Results Nonvented plastic prototype showed increased bacterial load compared with goal post (p = 0.04). Colonization was comparable between the goal post and all vented plastic prototypes (p ≥ 0.30) and when compared with the vented silicone device (p = 1).
Conclusion A novel silicone device does not increase external bacterial colonization compared with the current standard of care for line securement, and may provide a safe, convenient alternative to standard adhesive tape for UC stabilization. Future studies are anticipated to establish safety in vivo , alongside benefits such as migration and infection reduction.
Keywords umbilical arterial catheter - umbilical venous catheter - bloodstream infection - central line migration