Int J Angiol 2025; 34(01): 060-063
DOI: 10.1055/s-0040-1715084
Case Report

A New Technique to Puncture and Remove a PTCA Balloon that Fails to Deflate

1   Department of Cardiology, Ramsay Sime Darby Medical Center, Subang Jaya, Malaysia
,
Jeyamalar Rajadurai
1   Department of Cardiology, Ramsay Sime Darby Medical Center, Subang Jaya, Malaysia
,
Gim-Hooi Choo
2   Department of Cardiology, Cardiac Vascular Sentral KL, Kuala Lumpur, Malaysia
› Author Affiliations

Abstract

We present the first clinical report and describe the technique of using a modified Heartrail II guiding catheter to successfully puncture and remove a coronary angioplasty balloon that failed to deflate after stent implantation.



Publication History

Article published online:
16 September 2020

© 2020. International College of Angiology. This article is published by Thieme.

Thieme Medical Publishers
333 Seventh Avenue, New York, NY 10001, USA.