CC BY-NC-ND 4.0 · Indian J Radiol Imaging 2021; 31(04): 933-938
DOI: 10.1055/s-0041-1741050
Review Article

Biopsy with Side-Cutting Coaxial Needle—Knowing the “Cutting Length” and “Throw Length”

Anurima Patra
1   Department of Radiodiagnosis, Division of Clinical Radiology, Christian Medical College, Vellore, Tamil Nadu, India
,
Shyamkumar N. Keshava
1   Department of Radiodiagnosis, Division of Clinical Radiology, Christian Medical College, Vellore, Tamil Nadu, India
› Author Affiliations

Abstract

Image-guided Trucut biopsy is a well-established procedure. The length of the side notch in the stylet is the “cutting length,” which entraps the tissue sample and contributes to the yield. The total distance by which the inner stylet protrudes from the outer cannula with the cutting notch open is the “throw length.” It is inevitably longer than the cutting length does not add to the yield of the sample, but potentially to the complication of the procedure. The authors highlight the importance of knowing this distinction to minimize complications during the procedure.

Financial Support and Sponsorship

Nil.




Publication History

Article published online:
10 January 2022

© 2022. Indian Radiological Association. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/)

Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India