Subscribe to RSS
DOI: 10.1055/s-0042-105559
A novel occluder for endoscopic closure of gastrotomy: an ex vivo and in vivo animal study
Publication History
submitted 26 October 2015
accepted after revision 10 March 2016
Publication Date:
21 April 2016 (online)
Background and aims: We evaluated the feasibility of a novel occluder for endoscopic closure of natural orifice transluminal endoscopic surgery (NOTES) gastrotomy.
Methods: The occluder is a self-expandable, membrane-covered, metal structure for endoscopic delivery through gastrointestinal (GI) wall defects. The procedure time and air-tightness of endoscopic closure for a 12-mm perforation were compared for a through-the-scope clip group and an occluder group in an ex vivo animal study. The feasibility and safety of the occluder for the closure of NOTES gastrotomy were also evaluated in an in vivo animal study.
Results: In the ex vivo study, the median procedure time for closure was significantly shorter and the minimum air-leakage pressure was significantly greater in the occluder group than in the endoclip group. In the in vivo study, immediate air-tight sealing was achieved by application of the occluder and there were no serious adverse events. All animals showed spontaneous detachment of the occluder into the gastric lumen and complete healing within 4 – 8 weeks.
Conclusions: The novel occluder can provide immediate and permanent sealing for NOTES gastrotomy.
-
References
- 1 Liu L, Chiu PW, Reddy N et al. Natural orifice transluminal endoscopic surgery (NOTES) for clinical management of intra-abdominal diseases. Dig Endosc 2013; 25: 565-577
- 2 Sodergren MH, Clark J, Athanasiou T et al. Natural orifice translumenal endoscopic surgery: critical appraisal of applications in clinical practice. Surg Endosc 2009; 23: 680-687
- 3 Sodergren MH, Coomber R, Clark J et al. What are the elements of safe gastrotomy closure for NOTES? A systematic review. Surg Innov 2010; 17: 318-331
- 4 Perretta S, Sereno S, Forgione A et al. A new method to close the gastrotomy by using a cardiac septal occluder: long-term survival study in a porcine model. Gastrointest Endosc 2007; 66: 809-813
- 5 Lee HJ, Jung ES, Park MS et al. Closure of a gastrotracheal fistula using a cardiac septal occluder device. Endoscopy 2011; 43: E53-E54
- 6 Kramer MR, Peled N, Shitrit D et al. Use of Amplatzer device for endobronchial closure of bronchopleural fistulas. Chest 2008; 133: 1481-1484
- 7 Malespin M, Gaspar JP, Boulay B. Palliation of a malignant gastrocolic fistula with the use of an atrial septal defect occlusion device. Endoscopy 2014; 46: E4
- 8 Melmed GY, Kar S, Geft I et al. A new method for endoscopic closure of gastrocolonic fistula: novel application of a cardiac septal defect closure device (with video). Gastrointest Endosc 2009; 70: 542-545
- 9 Iqbal A, Haider M, Stadlhuber RJ et al. A study of intragastric and intravesicular pressure changes during rest, coughing, weight lifting, retching, and vomiting. Surg Endosc 2008; 22: 2571-2575