Endoscopy 2016; 48(12): 1125-1128
DOI: 10.1055/s-0042-116595
Innovations and brief communications
© Georg Thieme Verlag KG Stuttgart · New York

Safety and effectiveness of a long, partially covered metal stent for endoscopic ultrasound-guided hepaticogastrostomy in patients with malignant biliary obstruction

Yousuke Nakai
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Hiroyuki Isayama
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Natsuyo Yamamoto
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Saburo Matsubara
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
2   Department of Gastroenterology, Tokyo Metropolitan Police Hospital, Tokyo, Japan
,
Yukiko Ito
3   Department of Gastroenterology, Japanese Red Cross Medical Center, Tokyo, Japan
,
Naoki Sasahira
4   Department of Gastroenterology, Tokyo Takanawa Hospital, Tokyo, Japan
,
Ryunosuke Hakuta
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
3   Department of Gastroenterology, Japanese Red Cross Medical Center, Tokyo, Japan
,
Gyotane Umefune
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
3   Department of Gastroenterology, Japanese Red Cross Medical Center, Tokyo, Japan
,
Naminatsu Takahara
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Tsuyoshi Hamada
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Suguru Mizuno
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Hirofumi Kogure
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Minoru Tada
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
,
Kazuhiko Koike
1   Department of Gastroenterology, Graduate School of Medicine, The University of Tokyo, Tokyo, Japan
› Institutsangaben
Weitere Informationen

Publikationsverlauf

submitted 26. Januar 2016

accepted after revision 07. Juli 2016

Publikationsdatum:
07. Oktober 2016 (online)

Background and study aims: Endoscopic ultrasound (EUS)-guided biliary drainage (EUS-BD) is potentially complicated by bile leak and stent migration. The aim of this study was to evaluate the safety and effectiveness of a long (≥ 10 cm), partially covered metal stent (LP-CMS) for EUS-guided hepaticogastrostomy (EUS-HGS) for malignant biliary obstruction. Both the stent length and the uncovered portion at the proximal end of the LP-CMS are designed to prevent stent migration.

Patients and methods: A total of 33 patients undergoing EUS-HGS using an LP-CMS in four centers were retrospectively studied. Technical and clinical success, adverse events, and recurrent biliary obstruction were evaluated.

Results: Gastric outlet obstruction (76 %) and surgically altered anatomy (15 %) were two major reasons for EUS-HGS. The technical and clinical success rates were 100 %. The median intragastric stent length was 54 mm. The adverse event rate was 9 %. No stent migration was observed. Recurrent biliary obstruction developed in 24 %, with a median cumulative time to recurrence of 8.5 months.

Conclusions: EUS-HGS using an LP-CMS for unresectable malignant biliary obstruction was safe and effective.

 
  • References

  • 1 Park do H, Jang JW, Lee SS et al. EUS-guided biliary drainage with transluminal stenting after failed ERCP: predictors of adverse events and long-term results. Gastrointest Endosc 2011; 74: 1276-1284
  • 2 Khashab MA, Valeshabad AK, Afghani E et al. A comparative evaluation of EUS-guided biliary drainage and percutaneous drainage in patients with distal malignant biliary obstruction and failed ERCP. Dig Dis Sci 2015; 60: 557-565
  • 3 Itoi T, Isayama H, Sofuni A et al. Stent selection and tips on placement technique of EUS-guided biliary drainage: transduodenal and transgastric stenting. J Hepatobiliary Pancreat Sci 2011; 18: 664-672
  • 4 Okuno N, Hara K, Mizuno N et al. Stent migration into the peritoneal cavity following endoscopic ultrasound-guided hepaticogastrostomy. Endoscopy 2015; 47 (Suppl. 01) E311
  • 5 Cotton PB, Eisen GM, Aabakken L et al. A lexicon for endoscopic adverse events: report of an ASGE workshop. Gastrointest Endosc 2010; 71: 446-454
  • 6 Shima Y, Isayama H, Ito Y et al. Crisscross anchor-stents to prevent metal stent migration during endoscopic ultrasound-guided hepaticogastrostomy. Endoscopy 2014; 46: E563
  • 7 Itoi T, Binmoeller KF. EUS-guided choledochoduodenostomy by using a biflanged lumen-apposing metal stent. Gastrointest Endosc 2014; 79: 715
  • 8 Bruckner S, Arlt A, Hampe J. Endoscopic ultrasound-guided biliary drainage using a lumen-apposing self-expanding metal stent: a case series. Endoscopy 2015; 47: 858-861
  • 9 Ogura T, Chiba Y, Masuda D et al. Comparison of the clinical impact of endoscopic ultrasound-guided choledochoduodenostomy and hepaticogastrostomy for bile duct obstruction with duodenal obstruction. Endoscopy 2016; 48: 156-163
  • 10 Ogura T, Yamamoto K, Sano T et al. Stent length is impact factor associated with stent patency in endoscopic ultrasound-guided hepaticogastrostomy. J Gastroenterol Hepatol 2015; 30: 1748-1752
  • 11 Isayama H, Nakai Y, Kawakubo K et al. Feasibility and efficacy of a 12-cm long covered metallic stent for EUS-guided hepaticogastrostomy (EUS-HGS) for unresectable malignant biliary obstruction. Gastrointest Endosc 2013; 77: AB423
  • 12 Hamada T, Nakai Y, Isayama H et al. Tandem stent placement as a rescue for stent misplacement in endoscopic ultrasonography-guided hepaticogastrostomy. Dig Endosc 2013; 25: 340-341
  • 13 Hamada T, Nakai Y, Isayama H et al. Trimming a covered metal stent during hepaticogastrostomy by using argon plasma coagulation. Gastrointest Endosc 2013; 78: 817
  • 14 Vila JJ, Perez-Miranda M, Vazquez-Sequeiros E et al. Initial experience with EUS-guided cholangiopancreatography for biliary and pancreatic duct drainage: a Spanish national survey. Gastrointest Endosc 2012; 76: 1133-1141
  • 15 Song TJ, Lee SS, Park do H et al. Preliminary report on a new hybrid metal stent for EUS-guided biliary drainage (with videos). Gastrointest Endosc 2014; 80: 707-711