RSS-Feed abonnieren
DOI: 10.1055/s-0043-120523
Prevention of multidrug-resistant infections from contaminated duodenoscopes: Position Statement of the European Society of Gastrointestinal Endoscopy (ESGE) and European Society of Gastroenterology Nurses and Associates (ESGENA)
Publikationsverlauf
Publikationsdatum:
16. Oktober 2017 (online)
RECOMMENDATIONS
Patients should be informed about the benefits and risks of endoscopic retrograde cholangiopancreatography (ERCP)
Only specially trained and competent personnel should carry out endoscope reprocessing.
Manufacturers of duodenoscopes should provide detailed instructions on how to use and reprocess their equipment.
In the case of modifications to their equipment, manufacturers should provide updated instructions for use.
Detailed reprocessing protocols based on the manufacturer’s instructions for use should clearly lay out the different reprocessing steps necessary for each endoscope model.
Appropriate cleaning equipment should be used for duodenoscopes in compliance with the manufacturer’s instructions for use.
Only purpose-designed, endoscope type-specific, single-use cleaning brushes should be used, to ensure optimal cleaning.
As soon as the endoscope is withdrawn from the patient, bedside cleaning should be performed, followed by leak testing, thorough manual cleaning steps, and automated reprocessing, in order to:
-
Remove debris from external and internal surfaces;
-
Prevent any drying of body fluids, blood, or debris;
-
Prevent any formation of biofilms.
In addition to the leak test, visual inspection of the distal end as well as regular maintenance of duodenoscopes should be performed according to the manufacturer’s instructions for use, in order to detect any damage at an early stage.
The entire reprocessing procedure in endoscope washer-disinfectors (EWDs) should be validated according to the European and International Standard, EN ISO 15883.
Routine technical tests of EWDs should be performed according to the validation reports.
Microbiological surveillance of a proportion of the department’s endoscopes should be performed every 3 months, with the requirement that all endoscopes used in the unit are tested at least once a year.
In the case of suspected endoscopy-related infection, the relevant device (e. g., endoscope, EWD) should be taken out of service until adequate corrective actions have been taken.
Outbreaks should be managed by a multidisciplinary team, including endoscopy, hygiene, and microbiology experts, manufacturers, and regulatory bodies, according to national standards and/or laws.
In the case of suspected multidrug-resistant organism (MDRO) outbreaks, close cooperation between the endoscopy unit and the clinical health provider is essential (including infection control departments and hospital hygienists).
-
References
- 1 Kovaleva J, Peters FTM, van der Mei HC. et al. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Rev 2013; 26: 231-254
- 2 Muscarella LF. Risk of transmission of carbapenem-resistant Enterobacteriaceae and related “superbugs” during gastrointestinal endoscopy. World J Gastrointest Endosc 2014; 6: 457-474
- 3 Aumeran C, Poincloux L, Souweine B. et al. Multidrug-resistant Klebsiella pneumoniae outbreak after endoscopic retrograde cholangiopancreatography. Endoscopy 2010; 42: 895-899
- 4 Epstein L, Hunter JC, Arwady MA. et al. New Delhi metallo-β-lactamase-producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes. JAMA 2014; 312: 1447-1455
- 5 Wendorf K, Kay M, Baliga C. et al. Endoscopic retrograde cholangiopancreatography-associated AmpC Escherichia coli outbreak. Infect Control Hosp Epidemiol 2015; 36: 634-642
- 6 Smith ZL, Yuong SO, Saeian K. et al. Transmission of carbapenem-resistant Enterobacteriaceae during ERCP: time to revisit the current reprocessing guidelines. Gastrointest Endosc 2015; 81: 1041-1045
- 7 Kola A, Piening B, Pape UF. et al. An outbreak of carbapenem-resistant OXA-48-producing Klebsiella pneumonia associated to duodenoscopy. Antimicrob Resist Infect Control 2015; 4: 8 [doi:10.1186/s13756-015-0049-4]
- 8 Verfaillie CJ, Bruno MJ, Voor in ʼt holt AF. et al. Withdrawal of a novel-design duodenoscope ends outbreak of a VIM-2-producing Pseudomonas aeruginosa. Endoscopy 2015; 47: 493-502
- 9 US Food and Drug Administration (FDA). Safety Communication. Available from: http://www.fda.gov/MedicalDevices/Safety/AlertsandNotices/default.htm [Accessed: April 20 2017]
- 10 Bundesinstitut für Arzneimittel und Medizinprodukte. Empfehlungen des BfArM. Hinweis des BfArM und des RKI zu komplex aufgebauten Endoskopen (Duodenoskopen), deren Aufbereitung und damit verbundenen Infektionsrisiken. German. Available from: https://www.bfarm.de/SharedDocs/Risikoinformationen/Medizinprodukte/DE/bfarm_rki_aufbereitung_endoskop.html [Accessed: April 20 2017]
- 11 Humphries RM, McDonnell G. Superbugs on duodenoscopes: the challenge of cleaning and disinfection of reusable devices. J Clin Microbiol 2015; 53: 3118-3125
- 12 Petersen BT. Duodenoscope reprocessing: risk and options coming into view. Gastrointest Endosc 2015; 82: 484-486
- 13 Gastmeier P, Vonberg RP. Klebsiella spp. in endoscopy-associated infections: we may only be seeing the tip of the iceberg. Infection 2014; 42: 15-21
- 14 Kim S, Russell D, Mohamadnejad M. et al. Risk factors associated with the transmission of carbapenem-resistant Enterobacteriaceae via contaminated duodenoscopes. Gastrointest Endosc 2016; 83: 1121-1129
- 15 Guyatt GH, Schünemann HJ, Djulbegovic B. et al. Guideline panels should not GRADE good practice statements. J Clin Epidemiol 2015; 68: 597-600
- 16 Guyatt GH, Alonso-Coello P, Schünemann HJ. Guideline panels should seldom make good practice statements: guidance from the GRADE Working Group. J Clin Epidemiol 2016; 80: 3-7
- 17 Dumonceau JM, Hassan C, Riphaus A. et al. European Society of Gastrointestinal Endoscopy (ESGE) Guideline Development Policy. Endoscopy 2012; 44: 626-629
- 18 Rutala WA, Weber DJ. Gastrointestinal endoscopes: a need to shift from disinfection to sterilization?. JAMA 2014; 312: 1405-1406 [doi:10.1001/jama.2014.12559]
- 19 Jung M, Beilenhoff U. Hygiene: The looming Achilles heel in endoscopy. Visc Med 2016; 32: 21-28
- 20 Ross AS, Baliga C, Verma P. et al. A quarantine process for the resolution of duodenoscope-associated transmission of multidrug-resistant Escherichia coli. Gastrointest Endosc 2015; 82: 477-483
- 21 Beilenhoff U, Neumann CS, Rey JF. et al. ESGE-ESGENA guideline: cleaning and disinfection in gastrointestinal endoscopy. Endoscopy 2008; 40: 939-957
- 22 Association for the Advancement of Medical Instrumentation. ANSI/AAMI ST91: 2015. Flexible and semi-rigid endoscope processing in health care facilities. Available from: http://nascecme.com.br/2014/wp-content/uploads/2015/09/ANSI-AAMI-ST91-Endoscopy-2015.pdf [Accessed: April 20 2017]
- 23 American Society for Gastrointestinal Endoscopy. Multisociety guideline on reprocessing flexible GI endoscopes: 2016 update. Gastrointest Endosc 2017; 85: 282e-294e
- 24 Association of periOperative Registered Nurses (AORN). Guideline for processing flexible endoscopes. Guidelines for Perioperative Practice 2017; DOI: 10.6015/psrp.17.01.717.
- 25 Gastroenterological Society of Australia/Gastroenterological Nurses College of Australia (GESA/GENCA). Draft document for discussion. Infection control in endoscopy – Multidrug resistant outbreaks response. Version: April 2016. Available from: http://www.genca.org/public/5/files/Infection%20Control%20Guidelines%20Draft%20April%202016.pdf [Accessed: April 20 2017]
- 26 ESGENA. European Curriculum for reprocessing flexible endoscopes and endoscopic accessories 2017. Forthcoming 3 Nov 2017, at http://www.esgena.org/statements-curricula.html
- 27 Council Directive 93/42/EEC of 14 June 1993 concerning medical devices (OJ L 169, 12.7.1993, p. 1). Available from: http://eur-lex.europa.eu/LexUriServ/LexUriServ.do?uri=CONSLEG:1993L0042:20071011:en:PDF [Accessed: September 22 2017]
- 28 Roberts CG. The role of biofilms in reprocessing medical devices. Am J Infect Control 2013; 41: S77-S80
- 29 Otter JA, Vickery K, Walker JT. et al. Surface-attached cells, biofilms and biocide susceptibility: implications for hospital cleaning and disinfection. J Hosp Infect 2015; 89: 16-27
- 30 Bertschinger P, Brenneisen M, Cathomas A. et al. Schweizerische Richtlinie zur Aufbereitung flexibler Endoskope. Gemeinsame Richtlinie der Schweizerischen Gesellschaft für Gastroenterologie (SGG), Schweizerischen Gesellschaft für Pneumologie (SGP), Schweizerischen Gesellschaft für Spitalhygiene (SGSH), Schweizerischen Vereinigung für Endoskopiepersonal (SVEP). German. Available from: https://svep-aspe.ch/fileadmin/user_upload/pdf/Schweizerische_Hygienerichtlinie2013.pdf
- 31 Ministère de la santé, de la famille et des personnes handicapées, DGS/DHOS, Comité Technique national des Infections Nosocomiales (CTIN). Bonnes pratiques de désinfection des dispositifs médicaux. Guide pour l’utilisation des laveurs-désinfecteurs d’endoscopes. French. November 2003. Available from: http://nosobase.chu-lyon.fr/recommandations/Ministere_Sante/2003_endoscopie_ministere.pdf [Accessed: September 22 2017]
- 32 Sethi S, Huang RJ, Barakat M. et al. Adenosine triphosphate bioluminescence for bacteriologic surveillance and reprocessing strategies for minimizing risk of infection transmission by duodenoscopes. Gastrointest Endosc 2017; 85: 1180-1187
- 33 Ramsey AH, Oemig TV, Davis JP. et al. An outbreak of bronchoscopy-related Mycobacterium tuberculosis infections due to lack of bronchoscope leak testing. Chest 2002; 121: 976-981
- 34 Seki M, Hashiguchi K, Tanaka A. et al. Characteristics and disease severity of healthcare-associated pneumonia among patients in a hospital in Kitakyushu, Japan. Infect Chemother 2011; 17: 363-369
- 35 Beilenhoff U, Neumann CS, Biering H. et al. ESGE Guidelines Committee. ESGE/ESGENA guideline for process validation and routine testing for reprocessing endoscopes in washer-disinfectors, according to the European Standard prEN ISO 15883 parts 1, 4 and 5. Endoscopy 2007; 39: 85-94
- 36 Beilenhoff U, Biering H, Blum R et al. European Society of Gastrointestinal Endoscopy (ESGE)-European Society of Gastroenterology and Endoscopy Nurses and Associates (ESGENA). Technical Specification for process validation and routine testing of endoscope reprocessing in washer-disinfectors, according to EN ISO 15883, parts 1, 4, and ISO/TS 15883-5. Forthcoming Nov/Dec 2017
- 37 Beilenhoff U, Neumann CS, Rey JF. et al. ESGE-ESGENA guideline for quality assurance in reprocessing: microbiological surveillance testing in endoscopy. Endoscopy 2007; 39: 175-181
- 38 Centers for Disease Control and Prevention. Interim duodenoscope surveillance protocol, updated April 3, 2015. Available from: http://www.cdc.gov/hai/organisms/cre/cre-duodenoscope-surveillance-protocol.html [Accessed: April 20 2017]