Abstract

Background The Universal 2 total wrist arthroplasty was one of the most common wrist replacements, but long-term results were disappointing, due to substantial damage to the polyethylene component and, in some cases, metallosis. The purpose of this research was to investigate the underlying reasons for this polyethylene damage.

Methods From a single clinical center, retrieval analysis was undertaken on six Universal 2 wrists and two additional polyethylene components. All components were analyzed at the macroscale and dimensional measurements of polyethylene components were undertaken. These were compared against known component orientation in vivo to identify areas of greatest material loss.

Results Of 62 Universal 2 wrists implanted, there was a tendency for smaller implants to be revised more frequently. Of the six explanted wrists, material loss was always noticeable on the ulnar side, and to a lesser extent in the dorsal direction. Five of eight polyethylene components had failed at the base of the ulnar-side blind hole. Smaller implants tended to have less thickness at the base of the blind holes, thus explaining the failure of smaller sizes. Discoloration and surface fatigue of explants indicated oxidation of components. Average time in vivo for the explants was 13.8 years which indicates the slow nature of polyethylene oxidation. All were revised due to loosening. All were implanted in rheumatoid patients with a mean age at surgery of 56.1 years. Three of the six Universal 2 wrists, all sized Extra Small, showed severe wear of the titanium alloy carpal component.

Conclusion Failure of the polyethylene components was due to a combination of inappropriate sterilization technique plus a design issue where polyethylene was thinnest on the smallest size components. Continued surveillance of patients implanted with Universal 2 wrists is recommended.

Ethics Approval

This project has been registered with departmental research committee who felt that as it is a service evaluation, formal ethical approval is not required.


There is no patient identifiable information or images within this submission and as a result informed consent has not been obtained.




Publication History

Received: 10 June 2024

Accepted: 05 September 2024

Article published online:
03 October 2024

© 2024. Thieme. All rights reserved.

Thieme Medical Publishers, Inc.
333 Seventh Avenue, 18th Floor, New York, NY 10001, USA

 
  • References

  • 1 Divelbiss BJ, Sollerman C, Adams BD. Early results of the Universal total wrist arthroplasty in rheumatoid arthritis. J Hand Surg Am 2002; 27 (02) 195-204
  • 2 Brown M, Wharton R, Begum R, Trail I, Hayton M, Talwalkar S. Total wrist arthroplasty with the Freedom® prosthesis: a short-term follow-up. J Hand Surg Eur Vol 2024; 49 (01) 40-47
  • 3 Badge R, Kailash K, Dickson DR. et al. Medium-term outcomes of the Universal-2 total wrist arthroplasty in patients with rheumatoid arthritis. Bone Joint J 2016; 98-B (12) 1642-1647
  • 4 Martínez Villén G, Rodríguez Nogué L, García González E. Postoperative assessment and management of metallosis and periprosthetic osteolysis in patients treated with metal-on-polyethylene total wrist prostheses. J Hand Surg Eur Vol 2022; 47 (09) 952-958
  • 5 Newton A, Kandemir G, Joyce T. et al. Long-term outcomes of the Universal 2 total wrist replacement: revision and loosening at 10 years and beyond. J Hand Surg Eur Vol 2023; 48 (07) 641-647
  • 6 Ferreres A, Lluch A, Del Valle M. Universal total wrist arthroplasty: midterm follow-up study. J Hand Surg Am 2011; 36 (06) 967-973
  • 7 Sagerfors M, Gupta A, Brus O, Pettersson K. Total wrist arthroplasty: a single-center study of 219 cases with 5-year follow-up. J Hand Surg Am 2015; 40 (12) 2380-2387
  • 8 Gil JA, Kamal RN, Cone E, Weiss AC. High survivorship and few complications with cementless total wrist arthroplasty at a mean followup of 9 years. Clin Orthop Relat Res 2017; 475 (12) 3082-3087
  • 9 Pfanner S, Munz G, Guidi G, Ceruso M. Universal 2 wrist arthroplasty in rheumatoid arthritis. J Wrist Surg 2017; 6 (03) 206-215
  • 10 Kennedy JW, Ross A, Wright J, Martin DJ, Bransby-Zachary M, MacDonald DJ. Universal 2 total wrist arthroplasty: high satisfaction but high complication rates. J Hand Surg Eur Vol 2018; 43 (04) 375-379
  • 11 Zijlker HJA, Ritt MJPF, IJsselstein CB. Long-term results of Universal 2 total wrist arthroplasty. J Wrist Surg 2019; 8 (04) 317-320
  • 12 Fischer P, Sagerfors M, Jakobsson H, Pettersson K. Total wrist arthroplasty: a 10-year follow-up. J Hand Surg Am 2020; 45 (08) 780.e1-780.e10
  • 13 Gendera H, Chandrasegaram-Shanmuganathan S, Walbeehm R, Samijo S. Medium term outcomes of the Universal 2 total wrist arthroplasty in patients with degenerative or posttraumatic osteoarthritis of the wrist. Acta Orthop Belg 2020; 86 (03) 549-554
  • 14 Martínez Villén G, Rodríguez Nogué L. Universal 2™ total wrist arthroplasty: a single-surgeon 6.5-year follow-up study of 22 prostheses. Hand Surg Rehabil 2021; 40 (04) 413-419
  • 15 Kandemir G, Smith S, Schmidt I, Joyce TJ. Explant analysis of a Maestro™ wrist prosthesis and calculation of its lubrication regime. J Mech Behav Biomed Mater 2020; 110: 103933
  • 16 Stringer HJ, Cash T, Kennedy JA. et al. Long-term results and explant analysis of the ReMotion total wrist arthroplasty. J Hand Surg Eur 2024; 49 (08) 995-1001
  • 17 Julian HS, Kandemir G, Redfern J, J Joyce T, Brown DJ. Impingement-related osteolysis in Motec total wrist arthroplasty: an explant analysis and review of the literature. J Hand Surg Eur Vol 2024; 49 (01) 34-39
  • 18 Kandemir G, Julian HS, Joyce TJ, Brown DJ. Investigation of the in vivo performance of metal-on-metal Motec total wrist arthroplasty – an explant analysis. J Wrist Surg. 2024; (EFirst) In press
  • 19 Harris WH. Osteolysis and particle disease in hip replacement. A review. Acta Orthop Scand 1994; 65 (01) 113-123
  • 20 Garnier R, Poupon J, Djebrani-Oussedik N, Langrand J. Titanium cutaneous metallosis after reverse total shoulder arthroplasty. JSES Rev Rep Tech 2022; 2 (02) 190-193
  • 21 Rushton PRP, Smith SL, Fender D, Bowey AJ, Gibson MJ, Joyce TJ. Metallosis is commonly associated with magnetically controlled growing rods; results from an independent multicentre explant database. Eur Spine J 2021; 30 (07) 1905-1911
  • 22 Ho J, Mahajan J, Taylor M, Byers A, Arauz P, Kwon Y-M. Metallosis in cemented titanium alloy total knee arthroplasty without apparent metal-on-metal articulation. Knee 2018; 25 (04) 728-731
  • 23 Tarpada SP, Loloi J, Schwechter EM. A case of titanium pseudotumor and systemic toxicity after total hip arthroplasty polyethylene failure. Arthroplast Today 2020; 6 (04) 710-715
  • 24 Harrington CF, McKibbin C, Rahanu M, Langton D, Taylor A. Measurement of titanium in hip-replacement patients by inductively coupled plasma optical emission spectroscopy. Ann Clin Biochem 2017; 54 (03) 362-369
  • 25 Nunez Jr FA, Wright L, Kilpatrick SE, Seitz Jr WH. Revision total wrist arthroplasty due to polyethylene wear, metallosis-induced carpal tunnel syndrome, distal ulnar impingement, and fourth carpometacarpal joint pain: case report and pitfalls to avoid. Hand (N Y) 2020; 15 (01) NP1-NP6
  • 26 Heyes GJ, Julian HS, Mawhinney I. Metallosis and carpal tunnel syndrome following total wrist arthroplasty. J Hand Surg Eur Vol 2018; 43 (04) 448-450
  • 27 Day CS, Lee AH, Ahmed I. Acute carpal tunnel secondary to metallosis after total wrist arthroplasty. J Hand Surg Eur Vol 2013; 38 (01) 80-81
  • 28 Collier JP, Sperling DK, Currier JH, Sutula LC, Saum KA, Mayor MB. Impact of gamma sterilization on clinical performance of polyethylene in the knee. J Arthroplasty 1996; 11 (04) 377-389
  • 29 Premnath V, Harris WH, Jasty M, Merrill EW. Gamma sterilization of UHMWPE articular implants: an analysis of the oxidation problem. Ultra High Molecular Weight Poly Ethylene. Biomaterials 1996; 17 (18) 1741-1753
  • 30 Ries MD, Weaver K, Rose RM, Gunther J, Sauer W, Beals N. Fatigue strength of polyethylene after sterilization by gamma irradiation or ethylene oxide. Clin Orthop Relat Res 1996; (333) 87-95
  • 31 Edidin AA, Jewett CW, Kalinowski A, Kwarteng K, Kurtz SM. Degradation of mechanical behavior in UHMWPE after natural and accelerated aging. Biomaterials 2000; 21 (14) 1451-1460
  • 32 Saikko V. Effect of shelf versus accelerated aging of UHMWPE on delamination in knee wear simulation. Tribol Int 2014; 73: 10-16
  • 33 Morco S, Bowden A. Ulnar drift in rheumatoid arthritis: a review of biomechanical etiology. J Biomech 2015; 48 (04) 725-728
  • 34 Gislason MK, Foster E, Main D. et al. Mechanical testing and modelling of the Universal 2 implant. Med Eng Phys 2016; 38 (06) 511-517