Duodenal stenting has been gradually established in recent years because it is less invasive than standard surgical procedures and produces a rapid therapeutic response. For palliation of both duodenal and biliary stenoses, double stenting may be performed. Duodenal stents offer a great advantage in allowing endoscopic retrograde cholangiopancreatography (ERCP) without the need for balloon dilation. When biliary stent dysfunction occurs, the patient undergoes diagnostic and/or therapeutic ERCP across the duodenal stent. We encountered a duodenal stent fracture in a patient who required repeated ERCPs for stent dysfunction. Duodenal stent fractures have not previously been reported. The damaged stent was successfully repaired by using a cutting wire filament and placing another duodenal stent coaxially with the first. Clinicians should be aware of the possibility of stent fracture following endoscopic procedures, such as an ERCP, that require passage through the stent. The procedure described in this report would be of significant benefit if a gastrointestinal stent is fractured and occluded by a broken part.

References

  • 1 Maetani I, Ogawa S, Hoshi H, et al. Self-expanding metal stents for palliative treatment of malignant biliary and duodenal stenoses.  Endoscopy. 1994;  26 701-704
  • 2 Feretis C, Benakis P, Dimopoulos C, et al. Duodenal obstruction caused by pancreatic head carcinoma: palliation with self-expandable endoprosthesis.  Gastrointest Endosc. 1997;  46 161-165
  • 3 Rousseau H, Dahan M, Luque D, et al. Self-expandable prostheses in the tracheobronchial tree.  Radiology. 1993;  188 199-203
  • 4 Lee K W, Im J, Han J K, et al. Tuberculous stenosis of the left main bronchus: results of treatment with balloons and metallic stents.  J Vasc Interv Radiol. 1999;  10 352-358
  • 5 Ell C, Fleig W E, Hochberger J. Broken biliary metal stent after repeated electrocoagulation for tumor ingrowth.  Gastrointest Endosc. 1992;  38 97-98
  • 6 Peck R, Wattam J. Fracture of memotherm metallic stents in the biliary tract.  Cardiovasc Intervent Radiol. 2000;  23 55-56
  • 7 Winkelbauer F W, Schöefl R, Niederle B, et al. Palliative treatment of obstructing esophageal cancer with nitinol stents.  AJR. 1996;  166 79-84
  • 8 Schöefl R, Winkelbauer F, Haefner M, et al. Two cases of fractured esophageal nitinol stents.  Endoscopy. 1996;  28 518-520
  • 9 Cwickiel W, Tranberg K, Cwickiel M, et al. Malignant dysphagia: palliation with esophageal stents - long-term results in 100 patients.  Radiology. 1998;  207 513-518
  • 10 Wengrower D, Fiorini A, Valeo J, et al. Esophacoil: long-term results in 81 patients.  Gastrointest Endosc. 1998;  48 376-382
  • 11 Wigginghaus B, Dormann A J, Grünewald T. Self-expandable metallic stents in malignant gastric outlet obstructions - an alternative approach using modified techniques.  Z Gastroenterol. 1999;  37 1093-1099
  • 12 Carré P, Rousseau H, Lambart L, et al. Balloon dilatation and self-expanding metal Wallstent insertion for management of bronchostenosis following lung transplantation.  Chest. 1994;  105 343-348

I. Maetani, M.D.

Third Dept. of Internal Medicine · Toho University · Ohashi Hospital

2-17-6 Ohashi Meguro-ku · Tokyo 153-8515 · Japan

Fax: + 81-3-34681269

Email: maet@oha.toho-u.ac.jp