Mikrobiologische und histomorphometrische Untersuchung von Knochen-Pin-Interfaces in Fixateur externes verschiedenen Materials

Experimentelle Studie mit SchafenMicrobiological and Histomorphometric Analysis of the Bone-Pin Interface with Different External Fixation PinsAn Experimental Study in SheepG.  Mora1 , E.  Alvarez1 , P.  Ripalda1 , B.  Gahr1 , F.  Forriol1
  • 1Labor für experimentelle Orthopädie, Medizinische Fakultät der Universität von Navarra
An Stryker Trauma, Genf, für die Präparation und Bereitstellung der Materialien
Further Information

Publication History

Publication Date:
26 April 2002 (online)

Zusammenfassung

Ziel: Untersuchung der Pin-Knochen-Integration und der bakteriellen Kontamination von Fixateur-externe-Pins gleicher geometrischer Form, jedoch unterschiedlichen Materials (rostfreier Stahl, Titan und rostfreier Stahl mit Silberbeschichtung). Studiendesign: Wir teilten 18 Schafe (30 - 45 kg) in 3 Gruppen auf, um die oben genannten Charakteristika der Beziehung zwischen Pin und Knochen zu untersuchen. In histomorphometrischen Untersuchungen wurden die prozentualen Anteile von kortikalem und spongiösem Knochen sowie die Größe der Kontaktfläche zwischen Pin und Knochen bestimmt. Die mikrobiologischen Proben wurden am Eintrittspunkt des Pins im Knochen und an der Pinspitze entnommen. Ergebnisse: Die höchste Anzahl von Bakterien wurde auf der Oberfläche der silberbeschichteten Pins gefunden, der Ort der stärksten bakteriellen Kontamination war der Eintrittspunkt des Pins im spongiösen Knochen. Die histomorphometrische Untersuchung zeigte, dass die Pins mit der größten Kontaktfläche mit dem Knochen diejenigen aus Titan waren, gefolgt von den Stahlpins und schließlich den silberbeschichteten Pins. Schlussfolgerungen: Die Silberbeschichtung verhindert nicht die Ausbildung von bakterieller Flora, das Ausmaß der Kontamination war statistisch signifikant größer als inden anderen beiden Gruppen. Auch bezüglich der Integrationin den Knochen waren die Ergebnisse der Titan- und Stahlpins besser, wobei die Position des Pins im Knochen keine Rolle spielte.

Abstract

Aims: To study the effects of pin material and pin position within the bone on the quality of integration and the grade of bacterial contamination. Design: 18 sheep weighing 30 - 45 kgs were divided into 3 groups, with each receiving a geometrically equal external fixator pin of different material (corrosion-free steel, titanium or silver-laminated corrosion-free steel). A computerized histomorphometry analysis program was used to determine the percentual representation of cortical as well as cancellous bone, and the contact surface between pin and bone. Microbiological specimens were taken from both the pin's insertion point in the bone and the pin top. Results: The histomorphometric studies showed the titanium pins to have the highest grade of integration, followed by the steel pins, while the least amount of contact between pin and bone was observed in the interface of the silver-laminated steel pins. Moreover, the highest numberof bacteria per field was found on the surface of the silver-laminated pins. A stronger contamination was seen in the insertion point of cancellous than the cortical bone, regardless of thepin material. Conclusions: Silver-lamination is not an efficient means to avoid contamination, as we observed less bacteria per field and better integration results in the two other groups. Inaddition, the pin's position in the bone did seem to play a minor role in both pin-bone-integration and amount of contamination.

Literatur

  • 1 Checketts R G, Otterburn M, McEachern G. Pin track infection: definition, incidence and prevention.  Int J Orthop Trauma. 1993;  3 16-18
  • 2 Collinge C A, Günter G, David S. Pin Tract Infections: Silver vs Uncoated Pins.  Orthopaedics. 1994;  17 445-448
  • 3 Cunningham J L, Evans M, Kenwright J. Measurement of fracture movement in patients treated with unilateral external sceletal fixation.  J Biomed Eng. 1989;  11 118-122
  • 4 David A, Pommer A, Eitenüller J, Muhr G. Der Einfluss der Hidroxylapatit-Beschichtung von AO/ASIF-Schrauben auf die Haftfestigkeit im Knochen.  Unfallchirurg. 1993;  96 12-17
  • 5 Evans M, Spencer M, Wang Q, White S H, Cunningham J L. Design and testing of external fixator bone screws.  J Biomed Eng. 1990;  12 457-462
  • 6 Green S A. Complications of external skeletal fixation.  Clin Orthop. 1983;  180 109-116
  • 7 Huiskes R, Chao E YS, Crippen T E. Parametric analyses of pin-bone stresses in external fracture fixation devices.  J Orthop Res. 1985;  3 341-349
  • 8 Kleeman B C, Takeuchi T, Gerhart T N, Hayes W C. Holding power and reinforcement of cancellous screws in human bone.  Clin Orthop. 1992;  284 260-266
  • 9 Lee T K, Haynes R J, Longo J A, Chu J R. Pin removal in slipped capital femoral epiphysis: the unsuitability of titanium devices.  J Pediatric Orthopaedics. 1996;  16 49-52
  • 10 Magyar G, Toksvig-Larsen S, Moroni A. Hidroxyapatite coating of threaded pins enhances fixation.  J Bone Joint Surg. 1997;  79 487-489
  • 11 Mahan J, Seligson D, Henry S L, Hynes P, Dobbins J. Factors in pin tract infections.  Orthopaedics. 1991;  14 305-308
  • 12 Matthews L S, Green C A, Goldstein S A. The termal effects of skeletal fixation-pin insertion in bone.  J Bone Joint Surg. 1984;  66 1077-1083
  • 13 Pettine K A, Chao E YS, Kelly P J. Analysis of the external fixator pin bone interface.  Clin Orthop. 1993;  293 18-27
  • 14 Wilkenheiser M A, Markel M D, Lewallen D G, Chao E YS. Termal response and torque resistance of five cortical half-pins under simulated insertion technique.  J Orthop Res. 1995;  13 615-619

F. Forriol

Dpto. COT-CUN

Avda. Pio XII sn · 31008 Pamplona

Phone: 948255400

Fax: 948425649 ·

Email: fforriol@unav.es