RSS-Feed abonnieren
DOI: 10.1055/s-2004-814498
Sunlight Interference with Wireless Capsule Endoscopy
Publikationsverlauf
Publikationsdatum:
09. Mai 2006 (online)
A 30-year-old woman underwent capsule endoscopy (Given® M2A, Given Imaging Ltd., Yoqneam, Israel) to investigate melena. Her medical history was significant for neurofibromatosis, portal and mesenteric vein thrombosis, portorenal shunt surgery, and antrectomy. On examination, her abdomen was slightly protuberant, with multiple surgical scars. A reducible, golf ball-sized, ventral hernia in the epigastric region was noted when performing a Valsalva maneuver. After swallowing the capsule, the patient left the clinic wearing a white T-shirt covering the capsule endoscopy (CE) belt pack. The next 7 h were uneventful, and no discomfort was experienced.
On reviewing of the video data, at 4 h and with the capsule in the distal small bowel, strong interference by white light was observed in numerous segments of the small bowel, making interpretation of the video image impossible. External light was shining through the bowel wall, creating a whiteout on screen (Figure [1]). When capsule transit occurred or there was an altered orientation, light was seen to transilluminate from one side of the bowel wall (Figure [2]) or was observed in the distance (Figure [3]). The capsule locator program indicated a right upper quadrant position. The total duration of intermittent interference was approximately 40 min. Capsule transit did not appear to be delayed, with normal peristalsis and no small-bowel dilation being observed [1]. In the last 2 h of recording, the CE image was not disturbed. On being questioned later, the patient recalled direct sunlight exposure corresponding to the time of the CE image whiteout.
Figure 1 Capsule-endoscopic image, showing sunlight interfering with interpretation of the capsule video image.
Figure 2 Sunlight shining through one side of the bowel wall.
Figure 3 Sunlight shining in the distance.
To the best of our knowledge, this occurrence has not previously been reported in the medical literature. During colonoscopy or percutaneous endoscopic gastrostomy procedures, it is common to observe transabdominal illumination in the right lower quadrant or epigastrium, originating from the endoscope’s light source. Wireless CE depends on its own light source for image capture and recording. At present, there are no recommendations regarding shielding from strong ambient light during CE [2] [3]. In normal circumstances, the CE belt pack provides an adequate light shield over the mid-abdominal and flank regions. However, over the upper abdomen, in patients with prior abdominal surgery, ventral hernia, or a thin abdominal wall, there is a risk of interference with CE from an extrinsic light source. We would therefore recommend that patients should avoid direct sunlight during CE or should wear light-resistant clothing for the duration of the test.
Endoscopy_UCTN_Code_TTT_1AP_2AB
References
- 1 Tang S J, Zanati S A, Dubcenco E. et al . Capsule endoscopy regional transit abnormality: a sign of underlying small bowel pathology. Gastrointest Endosc. 2003; 58 598-602
- 2 Ginsberg G G, Barkun A N, Bosco J J. et al . Wireless capsule endoscopy. Gastrointest Endosc. 2002; 56 621-624
- 3 Given® diagnostic system. User’s manual RAPID® Rev. 2. Yoqneam, Israel; Given Imaging, Ltd 2002
P. Kortan, M. D.
Center for Therapeutic Endoscopy and Endoscopic Oncology
Victoria Wing 16-048
St. Michael’s Hospital
Toronto, Ontario M5B 1W8
Canada
Fax: +1-416-864-5619
eMail: kortanp.po3.smh@smh.toronto.on.ca