Abstract
Background and Study Aims: The incidence of viral contamination in the air, water and suction/accessory channels of gastrointestinal endoscopes was examined in order to evaluate the risk of infection.
Materials and Methods: After endoscopic examinations, including biopsy procedures, in 17 patients who were positive for hepatitis B virus surface antigen and eight patients who were positive for hepatitis C virus antibody, the endoscopes were cleaned on site by suctioning and flushing the air and water channels with an enzyme detergent. First samples were then collected by flushing 5 ml of sterile water through each channel. After mechanical reprocessing, second samples were collected in the same way. Virological studies were carried out with real-time polymerase chain reactions for hepatitis B virus DNA and hepatitis C virus RNA.
Results: Hepatitis B virus DNA was detected in five of the first samples recovered from the suction/accessory channels of the endoscopes (titers of 1.3 × 104 to 2.5 × 105 copies/ml), while no contamination was detected after reprocessing (P = 0.0445). The first samples from one water channel and three air channels were also positive for hepatitis B virus DNA, but were negative after reprocessing (P > 0.5, P = 0.227, respectively). No hepatitis C virus RNA was detected in any of the samples.
Conclusions: These results indicate that all of the channels were potential sources of viral infection.
References
-
1
Lynch D A, Porter C, Murphy L. et al .
Evaluation of four commercial automatic endoscope washing machines.
Endoscopy.
1992;
24
766-770
-
2
Ido K, Ishino Y, Ota Y. et al .
Deficiencies of automatic endoscopic reprocessors: a method to achieve high-grade disinfection of endoscopes.
Gastrointest Endosc.
1996;
44
583-586
-
3
Urayama S, Kozarek R A, Sumida S. et al .
Mycobacteria and glutaraldehyde: is high-level disinfection of endoscopes possible?.
Gastrointest Endosc.
1996;
43
451-456
-
4
Tremain S C, Orientale E, Rodney W M.
Cleaning, disinfection, and sterilization of gastrointestinal endoscopes: approaches in the office.
J Fam Pract.
1991;
32
300-305
-
5
Fraser V J, Zuckerman G, Clouse R E. et al .
A prospective randomized trial comparing manual and automated endoscope disinfection methods.
Infect Control Hosp Epidemiol.
1993;
14
383-389
-
6
Ishino Y, Ido K, Koiwai H. et al .
Pitfalls in endoscope reprocessing: brushing of air and water channels is mandatory for high-level disinfection.
Gastrointest Endosc.
2001;
53
165-168
-
7
Alvarado C J, Reichelderfer M.
APIC guideline for infection prevention and control in flexible endoscopy. Association for Professionals in Infection Control.
Am J Infect Control.
2000;
28
138-155
-
8
Ogoshi K, Akamatsu T, Iishi H. et al .
JGES guideline for cleaning and disinfection in gastrointestinal endoscopy.
Dig Endosc.
2000;
12
369-382
-
9
Systchenko R, Marchetti B, Canard J N. et al .
Guidelines of the French Society of Digestive Endoscopy: recommendations for setting up cleaning and disinfection procedures in gastrointestinal endoscopy.
Endoscopy.
2000;
32
807-818
-
10
BSG E ndoscopy.
Cleaning and disinfection of equipment for gastrointestinal endoscopy: report of a working party of the British Society of Gastroenterology endoscopy committee.
Gut.
1998;
42
585-593
-
11
Rey J, Kruse A, Axon T. et al .
ESGE guidelines for the prevention of endoscopic transmission of type C hepatitis and update on Creutzfeldt-Jakob disease.
Endoscopy.
1997;
29
203-204
-
12
Kovacs B J, Chen Y K, Kettering J D. et al .
High-level disinfection of gastrointestinal endoscopes: are current guidelines adequate?.
Am J Gastroenterol.
1999;
94
1546-1550
-
13
Foliente R L, Kovacs B J, Aprecio R M. et al .
Efficacy of high-level disinfectants for reprocessing GI endoscopes in simulated-use testing.
Gastrointest Endosc.
2001;
53
456-462
-
14
Birnie G G, Quigley E M, Clements G B. et al .
Endoscopic transmission of hepatitis B virus.
Gut.
1983;
24
171-174
-
15
Kasugai T, Yoshii Y, Shikata J. et al .
Digestive endoscopy and HBV infection (first report) [in Japanese with English abstract].
Gastroenterol Endosc.
1984;
27
2727-2733
-
16
Bronowicki J P, Venard V, Botté C. et al .
Patient-to-patient transmission of hepatitis C virus during colonoscopy.
N Engl J Med.
1997;
337
237-240
-
17 Tytgat G N. Endoscopic transmission of Helicobacter pylori. Aliment Pharmacol Ther 1995 9 (Suppl 2): 105-110
-
18
Akamatsu T, Tabata K, Hironga M. et al .
Transmission of Helicobacter pylori infection via flexible fiberoptic endoscopy.
Am J Infect Control.
1996;
24
396-401
-
19
Bond W W, Moncada R E.
Viral hepatitis B infection risk in flexible fiberoptic endoscopy.
Gastrointest Endosc.
1978;
24
225-230
-
20
Chanzy B, Duc-Bin D L, Rousset B. et al .
Effectiveness of a manual disinfection procedure in eliminating hepatitis C virus from experimentally contaminated endoscopes.
Gastrointest Endosc.
1999;
50
147-151
-
21
Abe A, Inoue K, Tanaka T. et al .
Quantitation of hepatitis B virus genomic DNA by real-time detection PCR.
J Clin Microbiol.
1999;
37
2899-2903
-
22
Takeuchi T, Katsume A, Tanaka T. et al .
Real-time detection system for quantification of hepatitis C virus genome.
Gastroenterology.
1999;
116
636-642
-
23
Deva A K, Vickery K, Zou J. et al .
Establishment of an in-use testing method for evaluating disinfection of surgical instruments using the duck hepatitis B model.
J Hosp Infect.
1996;
33
119-130
-
24
Vesley D, Melson J, Stanley P.
Microbial bioburden in endoscope reprocessing and an in-use evaluation of the high-level disinfection capabilities of Cidex PA.
Gastroenterol Nurs.
1999;
22
63-68
-
25
Deva A K, Vickery K, Zou J. et al .
Detection of persistent vegetative bacteria and amplified viral nucleic acid from in-use testing of gastrointestinal endoscopes.
J Hosp Infect.
1998;
39
149-157
-
26
Alfa M J, Jackson M.
A new hydrogen peroxide-based medical-device detergent with germicidal properties: comparison with enzymatic cleaners.
Am J Infect Control.
2001;
29
168-177
-
27
Di Marino A JJ.
Noncompliance with FDA and society guidelines for endoscopic reprocessing: implications for patient care.
Gastrointest Endosc Clin N Am.
2000;
10
283-294
-
28
Orsi G B, Filocamo A, di Stefano L. et al .
Italian national survey of digestive endoscopy disinfection procedures.
Endoscopy.
1997;
29
732-740
-
29
Cheung R J, Ortiz D, Di Marino A JJ.
GI endoscopic reprocessing practices in the United States.
Gastrointest Endosc.
1999;
50
362-368
-
30
Cleaning and Disinfection Committee for Endoscopes, Japanese Gastroenterological Endoscopy Society .
The Cleaning and Disinfection Committee for Endoscopes of the Japanese Gastroenterological Endoscopy Society report: questionnaire survey of digestive endoscopy disinfection procedure [in Japanese].
Gastroenterol Endosc.
1999;
41
215-219
Y. Ishino, M.D.
Dept. of Gastroenterology · Jichi Medical School
Minamikawachi-machi · Kawachi Gun · Tochigi 329-0498 · Japan
Fax: + 81-285-44-8297
eMail: ishino@jichi.ac.jp