Abstract
Objective : In light of the increased use of the LMA-ProSeal® (P-LMA) it seems desirable
to consider the P-LMA for prolonged surgical cases or post-operative ventilation.
Long-term use could carry the risk of serious complications. An experimental pig
model could be used to gain basic knowledge about long-term use before conducting
clinical trails. In this randomized controlled study we tested the hypothesis
if prolonged positive pressure ventilation (PPV) with the P-LMA in the pig is
possible. Methods : The P-LMA was used for airway maintenance under general anaesthesia and PPV
in 15 German country pigs. Randomised into 3 groups, the animals were ventilated
for 90, 150 and 390 minutes, respectively. Arterial blood gas measurements were
used to determine adequacy of ventilation. Airway leak pressure (Pleak ) and maximum tidal volume (TV) were measured in the supine, lateral and prone
positions. A bronchoscopic examination was performed to detect signs of aspiration.
Results : Adequate ventilation was demonstrated both after 120 min (4,7 ± 0.4 kPa) and
360 min (4,7 ± 0.5 kPa). The corresponding PaO2 at FiO2 = 1,0 was 59 ± 5 kPa and 64 ± 10 kPa, respectively. The mean Pleak was 34 ± 7 cm H2 O. Change of position did not result in any significant change in Pleak or maximum TV kg-1 . There was no bronchoscopic evidence of aspiration in any pig. Conclusion : It was possible to ensure sufficient PPV with the P-LMA in the pig over different
periods of time and in different body positions without any complications of airway
management. The P-LMA can be used for PPV in the pig for durations of up to 390
minutes. We conclude that investigations into the long-term use of the LMA can
be carried out in a pig model.
Zusammenfassung
Ziel der Studie : In Anbetracht des zunehmenden Einsatzes der LMA-ProSeal® (P-LMA) erscheint es
erwägenswert, die P-LMA bei langdauernden operativen Eingriffen oder postoperativen
Nachbeatmungen einzusetzen. Ein Langzeiteinsatz könnte das Risiko ernsthafter
Komplikationen in sich bergen. Vor der Durchführung klinischer Studien könnte
ein tierexperimentelles Model im Schwein dazu verwendet werden, grundsätzliche
Erkenntnisse über den Langzeiteinsatz der P-LMA zu gewinnen. Mit dieser randomisierten
Studie haben wir die Hypothese untersucht, dass eine langdauernde kontrollierte
Beatmung (PPV) unter Verwendung der P-LMA im Schwein möglich ist. Methodik : Die P-LMA wurde zur Freihaltung der Atemwege unter Allgemeinanästhesie und PPV
in 15 deutschen Landschweine verwendet. Randomisiert in 3 Gruppen wurden die Tiere
für jeweils 90, 150 und 390 Minuten kontrolliert beatmet. Die Suffizienz der Beatmung
wurde anhand von arteriellen Blutgasanalysen beurteilt. Der Atemwegsverschlussdruck
(Pleak ) und das maximale Tidalvolumen (TV) wurden in Rückenlage, Seitenlage und Bauchlage
bestimmt. Mit Hilfe einer Bronchoskopie wurde nach Anzeichen einer Aspiration
gesucht. Ergebnisse : Eine adäquate Ventilation konnte in allen Schweinen sowohl nach 120 min
(4,7 ± 0,4 kPa) als auch nach 360 min (4,7 ± 0,5 kPa) gezeigt werden. Der
korrespondierende PaO2 unter einem FiO2 = 1,0 war 59 ± 5 kPa bzw. 64 ± 10 kPa. Der mittlere Pleak war 34 ± 7 cm H2 O. Eine Veränderung der Lagerung resultierte nicht in signifikante Veränderungen
des Pleak oder maximalen TV kg-1 . In keinem Schwein wurden bronchoskopisch Anzeichen einer Aspiration gefunden.
Schlussfolgerung : Es war möglich, im Schwein ohne Komplikationen über verschiedene Zeiträume und
in verschiedenen Lagerungspositionen eine suffiziente PPV unter Verwendung der
P-LMA zu gewährleisten. Die P-LMA kann für die PPV von Schweinen von bis zu
390 Minuten Dauer eingesetzt werden. Wir schlussfolgern daraus, dass Untersuchungen
zum Langzeiteinsatz der P-LMA in einem Schweinemodel durchgeführt werden können.
Key words
Laryngeal mask airway - Experimental animal model - Aspiration - Pharyngo-laryngeal
morbidity - Positive pressure ventilation
Schlüsselwörter
Kehlkopfmaske - tierexperimentelles Model - Aspiration - pharyngo-laryngeale Morbidität
- kontrollierte Beatmung
References
1 Brimacombe J R, Brain A IJ, Berry A M. The Laryngeal Mask Airway: A review and practical
guide. London; W.B. Saunders 1997
2
Verghese C, Smith T GC, Young E.
Prospective survey of the use of the laryngeal mask airway in 2359 patients.
Anaesthesia.
1993;
48
58-60
3
Lu P P, Brimacombe J, Yang C, Shyr M.
ProSeal versus the classic laryngeal mask airway for positive pressure ventilation
during laparoscopic cholecystectomy.
Br J Anaesth.
2002;
88
824-827
4
Maltby R J, Beriault M T, Watson N C, Liepert D J, Fick G H.
LMA-Classic™ and LMA-ProSeal™ are effective alternatives to endotracheal intubation
for gynecologic laparoscopy.
Can J Anesth.
2003;
50
71-77
5
Reid L M.
Needs for animal models of human diseases of the respiratory system.
Am J Pathol.
1980;
101 (Suppl 3)
89-101
6
Kurita T, Morita K, Kazama T, Sato S.
Comparison of isoflurane and propofol-fentanyl anaesthesia in a swine model
of asphyxia.
Br J Anaesth.
2003;
91
871-877
7
Takala R SK, Soukka H R, Salo M S, Kirvelä O A, Kääpä P O, Rajamäki A A, Riutta A,
Aantaa R E.
Pulmonary inflammatory mediators after sevoflurane and thiopentone anaesthesia
in pigs.
Acta Anaesthesiol Scand.
2004;
48
40-45
8
Wemyss-Holden S A, Porter K J, Baxter P, Rudkin G E, Maddern G J.
The laryngeal mask airway in experimental pig anaesthesia.
Lab Anim.
1999;
33
30-34
9
Brimacombe J, Berry A.
A proposed fiber-optic scoring system to standardize the assessment of the laryngeal
mask airway position.
Anesth Analg.
1993;
76
457
10
Keller C, Brimacombe J, Keller K, Morris R.
Comparison of four methods for assessing airway sealing pressure with the laryngeal
mask airway in adult patients.
Br J Anaesth.
1999;
82
286-287
11
Brain A IJ, Verghese C, Strube P J.
The LMA ”ProSeal” - a laryngeal mask with an oesophageal vent.
Br J Anaesth.
2002;
84
650-654
12
Keller C, Brimacombe J, Kleinsasser A, Loeckinger A.
Does the ProSeal™ laryngeal mask airway prevent aspiration of regurgitated fluid?
A cadaveric study.
Anesth Analg.
2000;
91
1017-1020
13
Mark D A.
Protection from aspiration with the LMA-ProSeal™ after vomiting: a case report.
Can J Anesth.
2003;
50
78-80
14
Evans N R, Llewellyn R L, Gardner S V, James M FM.
Aspiration prevented by the ProSeal™ laryngeal mask airway: a case report.
Can J Anesth.
2002;
49
413-416
15
Cooper R M.
The LMA. Laparoscopic surgery and the obese patient - can vs. should.
Can J Anesth.
2003;
50
5-10
16
Peters J.
Alternatives to endotracheal intubation: a matter of statistics?.
Anesthesiol Intensivmed Notfallmed Schmerzther.
2004;
39
125-126
17
Voyagis G S, Dimitriou V K, Kyriakis K P.
Comparative evaluation of the prolonged use of the cuffed oropharyngeal airway
and the laryngeal mask airway in spontaneously breathing anaesthetized patients.
Eur J Anaesthesiol.
1999;
16
371-375
18
Brimacombe J, Archdeacon J.
The LMA for unplanned prolonged procedures.
Can J Anaesth.
1995;
42
1176-1179
19
Gandini D, Brimacombe J.
Laryngeal mask airway for ventilatory support over a 4-day period in a neonate
with Pierre Robin sequence.
Paed Anaesth.
2003;
13
181
20
Keller C, Brimacombe J, Lirk P, Pühringer F.
Failed obstetric tracheal intubation and postoperative respiratory support with
the ProSeal laryngeal mask airway.
Anesth Analg.
2004;
98
1467-1470
21
Brimacombe J, Keller C.
The laryngeal mask airway in fresh cadavers versus paralysed anaesthetized patients:
ease of insertion, airway sealing pressure, intracuff pressures and anatomic
position.
Eur J Anaesthesiol.
1999;
16
699-701
22
Bateman L, Ludders J W, Gleed R D, Erb H N.
Use of the laryngeal mask airway in rabbits during isoflurane anaesthesia.
Veterinary Anaesthesia and Analgesia.
2003;
30
117
23
Brietzke S E, Mair E A.
Laryngeal mask versus endotracheal tube in a ferret model.
Ann Otol Rhinol Laryngol.
2001;
110
827-833
24
Martins R H, Braz J R, Defaveri J, Gregório E A, Abud T MV.
Effect of high laryngeal mask airway intracuff pressures on the laryngopharyngeal
mucosa of dogs.
Laryngoscope.
2000;
110
645-650
25
Brimacombe J, Keller C.
The ProSeal laryngeal mask. A randomized, crossover study with the standard
laryngeal mask airway in paralyzed, anesthetized patients.
Anesthesiology.
2000;
93
104-109
26
Brimacombe J, Keller C, Boehler M, Pühringer F.
Positive pressure ventilation with the ProSeal versus classic laryngeal mask
airway: a randomized, crossover study of healthy female patients.
Anesth Analg.
2002;
95
770-776
27
Cook T M, Nolan J P, Verghese C, Strube P J, Less M, Millar J M, Baskett P J.
Randomized crossover comparison of the ProSeal with the classic laryngeal mask
airway in unparalyzed anaesthetized patients.
Br J Anaesth.
2002;
88
527-533
Kai Goldmann
Department of Anaesthesia and Intensive Care Therapy (Chairman: Univ.-Prof. Dr.
Hinnerk FW Wulf), Philipps University Marburg
Baldingerstrasse 1 · 35033 Marburg · Germany ·
Email: Kaigoldmann1@aol.com