Endoscopy 2006; 38(8): 808-812
DOI: 10.1055/s-2006-944612
Original article
© Georg Thieme Verlag KG Stuttgart · New York

Hemodynamic efficacy of the new resolution clip device in comparison with high-volume injection therapy in spurting bleeding: a prospective experimental trial using the compactEASIE simulator

J.  Maiss1 , C.  Baumbach1 , Y.  Zopf1 , A.  Naegel1 , M.  Wehler1 , T.  Bernatik1 , E.  G.  Hahn1 , D.  Schwab2
  • 1Dept. of Medicine I, University of Erlangen-Nuremberg, Erlangen, Germany
  • 2Dept. of Medicine II, Martha-Maria Hospital, Nuremberg, Germany
Weitere Informationen

Publikationsverlauf

Submitted 20 March 2006

Accepted after revision 10 May 2006

Publikationsdatum:
28. August 2006 (online)

Abstract

Background and study aims: Peptic ulcers are the most frequent cause of gastrointestinal bleeding. The use of hemoclips has become established as an effective form of treatment in addition to injection therapy. However, hemoclips have not previously been compared with injection therapy in an experimental setting using objective parameters.
Materials and methods: In a prospective, randomized, and controlled trial, the disposable Resolution hemoclip device (Boston Scientific, n = 40) was compared with conventional injection therapy (n = 40) in an experimental setting, using the compactEASIE simulator equipped with an upper gastrointestinal organ package to simulate bleeding. Four investigators with different levels of endoscopic experience participated in the study. On a randomized basis, each investigator treated 20 bleeding sites either by applying one clip (n = 10) or by carrying out high-volume four-quadrant injection (4 × 10 ml saline) of a spurting vessel. The efficacy of the hemostasis was assessed by continuous measurement of pressure within the afferent vessel before and after clip application or injection therapy and calculating the relative reduction in the vessel’s diameter with each treatment method. The system pressure was recorded 1 min before and 1 min after treatment. The ease of application of each method was rated by the endoscopist and by the assisting nurse using a visual analogue scale (0 - 100, with 100 being best).
Results: All of the 40 hemoclipping and injection treatments were carried out successfully. Both methods led to a significant increase in peak pressure (Resolution clip 71.8 ± 66.8 mm Hg, P < 0.001; injection 71.9 ± 53.8 mm Hg, P < 0.001), representing a significant relative reduction in the vessel diameter. There were no significant differences in peak pressure between the two treatments (P = 0.995). The mean increase in pressure during the first minute after the intervention (clip 49.3 ± 67.0 mm Hg vs. injection 19.9 ± 41.6 mm Hg) was significantly greater with the hemoclipping procedure (P = 0.021). More experienced investigators achieved a greater increase in system pressure, but the difference was not significant. The assessments of the ease of application by the assistants (84 ± 13) and endoscopists (86 ± 16) did not show any significant differences (P = 0.402) for the clipping device.
Conclusions: No significant differences between the two treatment methods were detected with regard to the immediate efficacy of hemostasis. However, long-term hemostasis was better with hemoclipping. The endoscopist’s level of experience also appears to play a role, particularly when hemoclips are used.

References

  • 1 Church N I, Palmer K R. Ulcers and nonvariceal bleeding.  Endoscopy. 2003;  35 22-26
  • 2 Higham J, Kang J Y, Majeed A. Recent trends in admissions and mortality due to peptic ulcer in England: increasing frequency of haemorrhage among older subjects.  Gut. 2002;  50 460-464
  • 3 Gralnek I M, Jensen D M, Gornbein J. et al . Clinical and economic outcomes of individuals with severe peptic ulcer hemorrhage and nonbleeding visible vessel: an analysis of two prospective clinical trials.  Am J Gastroenterol. 1998;  93 2047-2056
  • 4 Barkun A, Bardou M, Marshall J K. Consensus recommendations for managing patients with nonvariceal upper gastrointestinal bleeding.  Ann Intern Med. 2003;  139 843-857
  • 5 Laine L. Endoscopic therapy for bleeding ulcers: room for improvement?.  Gastrointest Endosc. 2003;  57 557-560
  • 6 Lin H J, Hsieh Y H, Tseng G Y. et al . A prospective, randomized trial of large- versus small-volume endoscopic injection of epinephrine for peptic ulcer bleeding.  Gastrointest Endosc. 2002;  55 615-619
  • 7 Raju G S, Gajula L. Endoclips for GI endoscopy.  Gastrointest Endosc. 2004;  59 267-279
  • 8 Cipolletta L, Bianco M A, Marmo R. et al . Endoclips versus heater probe in preventing early recurrent bleeding from peptic ulcer: a prospective and randomized trial.  Gastrointest Endosc. 2001;  53 147-151
  • 9 Chung I K, Ham J S, Kim H S. et al . Comparison of the hemostatic efficacy of the endoscopic hemoclip method with hypertonic saline-epinephrine injection and a combination of the two for the management of bleeding peptic ulcers.  Gastrointest Endosc. 1999;  49 13-18
  • 10 Gevers A M, De Goede E, Simoens M. et al . A randomized trial comparing injection therapy with hemoclip and with injection combined with hemoclip for bleeding ulcers.  Gastrointest Endosc. 2002;  55 466-469
  • 11 Maiss J, Molzahn A, Ruderich F. et al . National German inquiry on endoscopic and drug therapy in peptic ulcer bleeding [abstract].  Endoscopy. 2005;  37 (Suppl I) A 219
  • 12 Maiss J, Dumser C, Oeztuerk Y. et al . Hemodynamic efficacy of two endoscopic clip devices in the treatment of bleeding vessels: an experimental testing using the compactEASIE model.  Endoscopy. 2006;  38 575-580
  • 13 Cook D J, Guyatt G H, Salena B J. et al . Endoscopic therapy for acute nonvariceal upper gastrointestinal hemorrhage: a meta-analysis.  Gastroenterology. 1992;  102 139-148
  • 14 Laine L, Peterson W L. Bleeding peptic ulcer.  N Engl J Med. 1994;  331 717-727
  • 15 Hepworth C C, Kadirkamanathan S S, Gong F. et al . A randomised controlled comparison of injection, thermal, and mechanical endoscopic methods of haemostasis on mesenteric vessels.  Gut. 1998;  42 462-469
  • 16 Randall G M, Jensen D M, Hirabayashi K. et al . Controlled study of different sclerosing agents for coagulation of canine gut arteries.  Gastroenterology. 1989;  96 1274-1281
  • 17 Lai Y C, Yang S S, Wu C H. et al . Endoscopic hemoclip treatment for bleeding peptic ulcer.  World J Gastroenterol. 2000;  6 53-56

J. Maiss, M. D.

Dept. of Medicine I

Universität Erlangen-Nürnberg · Ulmenweg 18 · 91054 Erlangen · Germany

Fax: +49-9131-8535209

eMail: juergen.maiss@med1.imed.uni-erlangen.de