Free‐Flap Monitoring with Tissue‐Oxygen Measurement
Further Information
Publication History
Accepted for publication 1996
Publication Date:
08 March 2008 (online)
ABSTRACT
Flap monitoring by oxygen measurement with a microcatheter pO2 probe was carried out in 17 free flaps. The probe represents a new development among monitoring methods, allowing the measurement of oxygen pressure in all types of flaps including buried free flaps and vascularized bone grafts. Fourteen flaps survived with no problem; Three flaps with different degrees of problem are reported, and the advantages and sources of error using this monitoring method are discussed.