Horm Metab Res 1991; 23(12): 617-618
DOI: 10.1055/s-2007-1003769
Progress in Practical Endocrinology

© Georg Thieme Verlag, Stuttgart · New York

Combination of Microdialysis and Glucosensor Permits Continuous (On Line) s. c. Glucose Monitoring in a Patient Operated Device: I In Vitro Evaluation

F. S. Keck2 , W. Kerner, C. Meyerhoff1 , H. Zier1 , E. F. Pfeiffer1
  • 1Institut für Diabetestechnologie, Medizinische Klinik und Poliklinik, Universität Ulm, Ulm, Germany
  • 2Abteilung Innere Medizin I, Medizinische Klinik und Poliklinik, Universität Ulm, Ulm, Germany
Weitere Informationen

Publikationsverlauf

1991

1991

Publikationsdatum:
14. März 2008 (online)

Summary

A device for continuous glucose monitoring in fluids was obtained by combining the microdialysis technique with a measuring flow chamber of the “Glucosensor Unitec Ulm” using the GOD method for determining amperometrically blood glucose profiles. The in vitro experiments demonstrate that the relative recovery of glucose by this device is inversely related to the flow rate of the microdialysis perfusion fluid, which, in turn, is inversely related to the response time of the device. The glucose signal increases linearly with the area of the microdialysis working membrane (r = 0.98), and with the glucose concentrations of the standard solutions (r > 0.95). The variation coefficient for repeated measurements is below 8%. The accuracy of the device as demonstrated by mean measuring deviation ranges between 1 and 3.8%.