RSS-Feed abonnieren
Bitte kopieren Sie die angezeigte URL und fügen sie dann in Ihren RSS-Reader ein.
https://www.thieme-connect.de/rss/thieme/de/10.1055-s-00000085.xml
Thorac Cardiovasc Surg 1998; 46(3): 154-156
DOI: 10.1055/s-2007-1010214
DOI: 10.1055/s-2007-1010214
Case Report
© Georg Thieme Verlag Stuttgart · New York
Successful Treatment of a Novacor LVAD Malfunction Without Repeat Sternotomy
Weitere Informationen
Publikationsverlauf
1998
Publikationsdatum:
19. März 2008 (online)

Abstract
Implantable left-ventricular assist devices have been remarkably free of mechanical failures. We describe an uncommon Novacor N100 PC© LVAS malfunction caused by an internal short circuit of the device due to urine aspiration via the vent line. Device replacement was managed via a subcostal approach without sternotomy. Patient recovery was uneventful and successful transplantation was performed one month after the device exchange.
Key words
Implantable LVAD - Malfunction - Subcostal approach - Reimplantation