Thorac Cardiovasc Surg 1981; 29(2): 108-109
DOI: 10.1055/s-2007-1023454
DOI: 10.1055/s-2007-1023454
Strut Fracture in the New Bjørk Shiley Mitral Valve Prosthesis
Further Information
Publication History
1980
Publication Date:
19 March 2008 (online)
Summary
The case of a patient with the new type Bjørk-Shiley aortic and mitral valve prosthesis is described.
Three months after implant she suffered acute heart failure and died. Post-mortem examination revealed a fractured outlet strut in the mitral valve prosthesis with dislocation of the disc. The fracture was regarded as due to excessive brittleness caused by demonstrated deposition of chromium-tungsten-carbide.
Key words
Strut fracture - Bjørk-Shiley Valve Prosthesis