Subscribe to RSS
DOI: 10.1055/s-2008-1052701
Significance of balloon pressure recording during angioplasty
An experimental studyBedeutung der Registrierung des Ballonkatheter-Füllungsdrucks bei der AngioplastiePublication History
Publication Date:
20 March 2008 (online)

Summary
During angioplasty of artificial stenoses, atherosclerotic human cadaver arteries, and normal canine arteries, pressure and volume of the dilatation balloons were continuously recorded. We found that a sudden yield of a lesion corresponded to a sudden drop in the pressure curve and an increase of the balloon volume. Volume monitoring was insensitive, but pressure recording was very precise. Continuous pressure recording, using a non-compliant inflation system, correctly demonstrated small breaks and ruptures of atherosclerotic vessels, changes not seen on fluoroscopy.
Zusammenfassung
Druck und Volumen der Dilatations-Ballonkatheter wurden während der Angioplastie künstlicher Stenosen, atherosklerotischer Arterien menschlicher Leichen und normaler Hundearterien kontinuierlich registriert. Wir stellten fest, daß bei plötzlichem Nachgeben einer Läsion ein entsprechender plötzlicher Druckabfall und ein Anstieg des Ballonkathetervolumens stattfand. Während die Anzeige des Volumens nicht empfindlich reagierte, lieferte die Anzeige der Druckschwankungen ein sehr genaues Kriterium. Die kontinuierliche Aufzeichnung der Drücke unter Anwendung eines „non-compliant inflation System” zeigte mit Genauigkeit kleinere Brüche und Rupturen atherosklerotischer Gefäße an, die mittels Fluoroskopie nicht erkennbar gewesen waren.