Evaluating the use of commercial drug-specific calibrators for determining PT and APTT reagent sensitivity to dabigatran and rivaroxaban
Publication History
Received:
16 April 2014
Accepted after major revision:
22 August 2014
Publication Date:
27 November 2017 (online)
Summary
Suitable laboratory methodologies for quantifying the non-vitamin K oral anticoagulants (NOAC) include liquid chromatography-tandem mass spectrometry (LC-MS/MS) or drug-calibrated assays such as the dilute thrombin time for dabigatran or anti-Xa measurements for rivaroxaban. In situations when these tests are unavailable, it has been suggested that using commercial drug calibrators on APTT and PT assays would theoretically provide reagent sensitivity to these drugs. The purpose of this study was to determine whether commercial drug calibrators deliver similar reagent sensitivity information as samples from patients receiving dabigatran or rivaroxaban as part of their routine care. Two laboratory sites tested commercial calibrator material for dabigatran and rivaroxaban (Hyphen Biomedical) using PT and APTT reagents and data was compared to samples collected from patients taking NOACs that were quantified by LC-MS/MS. Correlation statistics and calculating the amount of drug required to double the clotting time of normal plasma were performed. All drug calibrator material correlated more strongly (R2> 0.95) for any reagent/drug combination than patient samples (R2 ranged from 0.29–0.86). Dabigatran calibrator results and patient data were equivalent for SynthASil and PTT-A APTT reagents. The dabigatran and rivaroxaban calibrator material over-estimated drug sensitivity for all PT reagents when compared to sensitivity data calculated based on drug levels obtained by LC-MS/MS from patient samples. In conclusion, drug-specific calibrators overestimated reagent sensitivity which may underestimate in vivo drug concentration in a given patient. Further studies are required to assess whether this method of determining relative sensitivity of NOAC on routine coagulation assays should be recommended.
-
References
- 1 van Ryn J. et al. Dabigatran etexilate--a novel, reversible, oral direct thrombin inhibitor: interpretation of coagulation assays and reversal of anticoagulant activity. Thromb Haemost 2010; 103: 1116-11127.
- 2 Dager WE. et al. Dabigatran effects on the international normalized ratio, activated partial thromboplastin time, thrombin time, and fibrinogen: a multicenter, in vitro study. Ann Pharmacother 2012; 46: 1627-1636.
- 3 Mani H. et al. Ex vivo effects of low-dose rivaroxaban on specific coagulation assays and coagulation factor activities in patients under real life conditions. Thromb Haemost 2013; 109: 127-136.
- 4 Hawes EM. et al. Performance of coagulation tests in patients on therapeutic doses of dabigatran: a cross-sectional pharmacodynamic study based on peak and trough plasma levels. J Thromb Haemost 2013; 11: 1493-1502.
- 5 Francart SJ. et al. Performance of coagulation tests in patients on therapeutic doses of rivaroxaban. A cross-sectional pharmacodynamic study based on peak and trough plasma levels. Thromb Haemost 2014; 111: 1133-1140.
- 6 Gosselin RC. et al. Measuring dabigatran concentrations using a chromogenic ecarin clotting time assay. Ann Pharmacother 2013; 47: 1635-1640.
- 7 Douxfils J. et al. Comparison of calibrated chromogenic anti-Xa assay and PT tests with LC-MS/MS for the therapeutic monitoring of patients treated with rivaroxaban. Thromb Haemost 2013; 110: 723-731.
- 8 Mani H. et al. Accurate determination of rivaroxaban levels requires different calibrator sets but not addition of antithrombin. Thromb Haemost 2012; 108: 191-198.
- 9 Samama MM. et al. Rivaroxaban Anti-Factor Xa Chromogenic Assay Field Trial Laboratories. Evaluation of the anti-factor Xa chromogenic assay for the measurement of rivaroxaban plasma concentrations using calibrators and controls Thromb Haemost 2012; 107: 379-387.
- 10 Huisman MV. et al. Dabigatran etexilate for stroke prevention in patients with atrial fibrillation: resolving uncertainties in routine practice. Thromb Haemost 2012; 107: 838-847.
- 11 Turpie AG. et al. Management consensus guidance for the use of rivaroxaban-an oral, direct factor Xa inhibitor. Thromb Haemost 2012; 108: 876-886.
- 12 Baglin T. et al. Effects on routine coagulation screens and assessment of anticoagulant intensity in patients taking oral dabigatran or rivaroxaban: Guidance from the British Committee for Standards in Haematology. Br J Haematol 2012; 159: 427-429.
- 13 Baglin T. et al. Measuring Oral Direct Inhibitors (ODIs) of thrombin and factor Xa: A recommendation from the Subcommittee on Control of Anticoagulation of the Scientific and Standardisation Committee of the International Society on Thrombosis and Haemostasis. J Thromb Haemost 2013; 11: 756-760.
- 14 Douxfils J. et al. Impact of dabigatran on a large panel of routine or specific coagulation assays. Laboratory recommendations for monitoring of dabigatran etexilate. Thromb Haemost 2012; 107: 985-997.
- 15 Samama MM. et al. Assessment of laboratory assays to measure rivaroxaban-an oral, direct factor Xa inhibitor. Thromb Haemost 2010; 103: 815-825.
- 16 Gosselin RC, Dwyre DM. Determining the effect of freezing on coagulation testing: comparison of results between fresh and once frozen-thawed plasma. Blood Coag Fibrinol. 2014 Epub ahead of print.
- 17 Gosselin R. et al. Effect of telavancin (Vibativ) on routine coagulation test results. Am J Clin Pathol 2011; 136: 848-854.
- 18 van den Besselaar AM. et al. Phosphatidylglycerol and daptomycin synergisti-cally inhibit tissue factor-inducedcoagulation in the prothrombin time test. J Thromb Haemost 2010; 08: 1429-1430.