Subscribe to RSS
DOI: 10.3766/jaaa.14.1.2
Development of a 500-Hz Masking-Level Difference Protocol for Clinic Use
Publication History
Publication Date:
07 August 2020 (online)
The purpose of this series of experiments was to develop a simple, 500-Hz masking-level difference (MLD) protocol that could be implemented easily in the clinic to assess auditory perceptual abilities using an audio compact disc. Five, 300-ms tones with 250-ms intertone intervals were embedded in 3-s bursts of 200-800 Hz noise presented at 42.2-dB pressure-spectrum level with 4-5 s inter-stimulus intervals. The homophasic and antiphasic conditions were interleaved with the signal-to-noise ratios decreasing in 2-dB steps. A single-interval, "yes/no" response task was used. Three experiments were performed on 24-28 listeners with normal hearing. The mean SoNo thresholds (58.1- to 59.5-dB SPL) and the mean SπNo thresholds (45.1- to 46.0-dB SPL) produced ˜13-dB MLDs. Experiment 3 included a SoNπ condition that had a mean threshold of 48.8-dB SPL and a 10.0-dB MLD. The mean test, retest ot the SoNo and SπNo thresholds on 15 listeners was <0.5 dB. Over the three experiments, 95% of the listeners had SπNo MLDs that were ≥10 dB.