The effect of wrapping of light-cure tips on the cure of composite resin
Abstract
Aim: Dental curing lights are vulnerable to contamination during routine intra-oral use. The purpose of this study was to evaluate the effect of wrapping light-cure tips on the hardness of cured composite resin. Settings and Design: Two types of barriers were used, which are the commercially available cellophane wraps and the disposable light-cure sleeves. A new light-cure unit used with a standard light-cure tip fixed at distance of 0.5 mm from the specimens. Materials and Methods: Sixty molds of 8 mm diameter and 2 mm depth filled with same type and shade of composite resin; 20 specimens cured without wrapping the tips, and 20 used for each wrapping method with new barrier was placed for each. The light-cure intensity and the Knoop hardness value were recorded. Statistical Analysis: Kruskal-Wallis test was used to analyze the statistical differences between the groups. Pearson′s correlation analysis was used to determine the correlation between light intensity and Knoop hardness values. Results: The results of this study showed that there were significant differences in the light intensity among the groups. However, there were no significant differences in the Knoop hardness values among the 3 groups. The cellophane-wrapped around the light-cure tips is appeared to cause the least reduction of hardness from the non-wrapped tip. Conclusions: The effect of the barriers on the hardness of composite resin was small and probably clinically insignificant. The light intensity output was above the acceptable curing levels using any of the two barriers. The use of non-opaque barriers with the light-cure tips is recommended to prevent cross contamination.
Publikationsverlauf
Artikel online veröffentlicht:
01. November 2021
© 2012. European Journal of General Dentistry. This is an open access article published by Thieme under the terms of the Creative Commons Attribution-NonDerivative-NonCommercial-License, permitting copying and reproduction so long as the original work is given appropriate credit. Contents may not be used for commercial purposes, or adapted, remixed, transformed or built upon. (https://creativecommons.org/licenses/by-nc-nd/4.0/.)
Thieme Medical and Scientific Publishers Pvt. Ltd.
A-12, 2nd Floor, Sector 2, Noida-201301 UP, India
-
References
- 1 Kawaguchi M, Fukushima T, Miyazaki K. The relationship between cure depth and transmission coefficient of visible-light-activated resin composites. J Dent Res 1994;73:516-21.
- 2 Caughman WF, Rueggeberg FA, Curtis JW Jr. Clinical guidelines for photocuring restorative resins. J Am Dent Assoc 1995;126:1280-6.
- 3 Fan PL, Schumacher RM, Azzolin K, Geary R, Eichmiller FC. Curing-light intensity and depth of cure of resin-based composites tested according to international standards. J Am Dent Assoc 2002;133:429-34.
- 4 Shortall AC, Harrington E. Guidelines for the selection, use and maintenance of visible light activation units. Br Dent J 1996;181:383-7.
- 5 Autio KL, Rosen S, Reynolds NJ, Bright JS. Studies on cross-contamination in the dental clinic. J Am Dent Assoc 1980;100:358-61.
- 6 Miller CH, Palenik CJ. Infection control and management of hazardous materials. St. Louis: Mosby Pub. Co.; 1994
- 7 Nelson SK, Caughman WF, Rueggeberg FA, Lockwood PE. Effect of glutaraldehyde cold sterilants on light transmission of curing tips. Quintessence Int 1997;28:725-30.
- 8 Dugan WT and Hartleb JH. Influence of a glutaraldehyde disinfecting solution on curing light effectiveness. Gen Dent 1989;37:40-3.
- 9 Rueggeberg FA, Caughman WF, Comer RW. The effect of autoclaving on energy transmission through light-curing tips. J Am Dent Assoc 1996;127:1183-7.
- 10 Kofford KR, Wakefield CW, Nunn ME. The effect of autoclaving and polishing techniques on energy transmission of light-curing tips. Quintessence Int 1998;29:491-6.
- 11 Morrow L, Wilson NH, Setcos JC. Single-use, disposable, pre-sterilised light-activation probe: The future? Quintessence Int 1998;29:1-5.
- 12 Rueggeberg FA, Caughman WF. Factors affecting light transmission of single-use, plastic light-curing tips. Oper Dent 1998;23:179-84.
- 13 Chong SL, Lam YK, Lee FK, Ramalingam L, Yeo AC, Lim CC. Effect of various infection-control methods for light-cure units on the cure of composite resins. Oper Dent 1998;23:150-4.
- 14 Warren DP, Rice HC, Powers JM. Intensity of curing lights affected by barriers. J Dent Hyg 2000;74:20-3.
- 15 Caughman GB, Caughman WF, Napier N, Schuster GS. Disinfection of visible-light-curing devices. Oper Dent 1989;14:2-7.
- 16 Rueggeberg FA, Caughman WF, Curtis JW Jr. Effect of light intensity and exposure duration on cure of resin composite. Oper Dent 1994;19:26-32.
- 17 Soh MS, Yap AU, Siow KS. Comparative depths of cure among various curing light types and methods. Oper Dent 2004;29:9-15.
- 18 Kurachi C, Tuboy AM, Magalhães DV, Bagnato VS. Hardness evaluation of a dental composite polymerized with experimental LED-based devices. Dent Mater 2001;17:309-15.
- 19 Hodson NA, Dunne SM, Pankhurst CL. The effect of infection-control barriers on the light intensity of light-cure units and depth of cure of composite. Primary Dental Care 2005;12:61-7.
- 20 Scott BA, Felix CA, Price RB. Effect of disposable infection control barriers on light output from dental curing lights. J Can Dent Assoc 2004;70:105-10.